British Journal of Radiology (2007) 80, 926-933
© 2007 British Institute of Radiology
doi: 10.1259/bjr/51363812

This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jackson, A S N
Right arrow Articles by Dearnaley, D P
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jackson, A S N
Right arrow Articles by Dearnaley, D P

Full paper

Distortion-corrected T2 weighted MRI: a novel approach to prostate radiotherapy planning

A S N Jackson, FRCR 1 S A Reinsberg, PhD 2 S A Sohaib, FRCR 3 E M Charles-Edwards, MSc 2 S A Mangar, FRCR 1 C P South, MSc 4 M O Leach, PhD 2 and D P Dearnaley, MD, FRCR 1

1 Academic Department of Radiotherapy and Oncology, 2 CRUK Magnetic Resonance Imaging Group, 3 Department of Radiology, 4 Joint Department of Physics, Royal Marsden Hospital and Institute of Cancer Research, Downs Rd, Sutton SM2 5PT, UK

Correspondence: Dr A S N Jackson, Academic Department of Radiation Oncology, Christie Hospital NHS Trust, Wilmslow Rd, Manchester M20 4BX, UK. E-mail: Andrew.Jackson{at}manchester.ac.uk


    Abstract
 Top
 Abstract
 Introduction
 Methods and patients
 Results
 Discussion
 Conclusions
 References
 
The purpose of this study was to evaluate distortion-corrected MRI as a radiotherapy planning tool for prostate cancer and the resultant implications for dose sparing of organs at risk. 11 men who were to be treated with radical conformal radiotherapy for localized prostate cancer had an MRI scan under radiotherapy planning conditions, which was corrected for geometric distortion. Radiotherapy plans were created for planning target volumes derived from the MRI- and CT-defined prostate. Dose volume histograms were produced for the rectum, bladder and penile bulb. The mean volume of the prostate as defined on CT and MRI was 41 cm3 and 36 cm3, respectively (p = 0.009). The predicted percentage of the rectum treated to dose levels of 45–65 Gy was significantly lower for plans delineating the prostate with MRI than for those with CT. The rectal-sparing effect was confined to the lowermost 4 cm of the rectum (anal canal). There were no differences between the predicted doses to bladder or penile bulb (as defined using MRI) between plans. In conclusion, prostate radiotherapy planning based on distortion-corrected MRI is feasible and results in a smaller target volume than does CT. This leads to a lower predicted proportion of the rectum, in particular the lower rectum (anal canal), treated to a given dose than with CT.


    Introduction
 Top
 Abstract
 Introduction
 Methods and patients
 Results
 Discussion
 Conclusions
 References
 
CT represents the standard method of target localization in radical conformal radiotherapy for prostate cancer. CT is well-suited as a radiotherapy planning tool because it provides electron density data that are used for dosimetric planning. In addition, CT is free from geometric distortion, i.e. spatial information is faithfully represented to scale. CT is, however, limited by poor soft-tissue contrast such that, in the case of the prostate gland, the interface between the prostate and the pelvic floor muscles may be indistinct.

MRI offers superior soft-tissue contrast to CT, and comparisons between the two have shown that CT results in a probable overestimation in the size of the prostate and uncertainty as to the precise location of the apex of the gland, when compared with MRI [18]. This may have an impact on the radiation dose to organs at risk (OARs), in particular the rectum and penile erectile tissue [2, 810]. Definition of the prostate with MRI rather than CT could lead to sparing of these structures, with a consequent reduction in toxicity, and allow radiotherapy dose escalation.

The use of MRI as a radiotherapy planning tool has been hampered by the lack of direct electron density data available with MRI, and the presence of geometric distortion inherent in the technique. These issues must be addressed before MRI can replace CT as the imaging of choice for prostate radiotherapy planning. Bulk assignment of electron density data to MRI images is feasible, with dosimetric planning on such images resulting in only minor differences from CT [11]. Our group have described a method for correcting standard clinically available three-dimensional (3D) MRI images for patient-and system-based distortion [1214]. In this study, we evaluate the practicality of using distortion-corrected MRI for target definition in prostate cancer radiotherapy planning, and the consequences in terms of the characteristics of the target volume and the predicted radiotherapy doses to OARs.


    Methods and patients
 Top
 Abstract
 Introduction
 Methods and patients
 Results
 Discussion
 Conclusions
 References
 
The research protocol was approved by the Royal Marsden NHS Trust and Institute of Cancer Research Committee for Clinical Research and Regional Ethics Committee. Informed consent was obtained from each patient.

Eligibility
Eligible patients were those who were to undergo CT-planned radical conformal radiotherapy for carcinoma of the prostate, with no contraindications to MRI. Patients were required not to have had insertion of any metallic prostheses likely to degrade the CT and MRI images. 14 patients were recruited between December 2003 and June 2004.

CT scanning
Patients were scanned supine on a flat couch top with the arms folded on the chest. A foam headrest was used, and foam wedges were placed behind the knees. Ankle stocks were used to prevent rotation of the hips. In order to achieve a comfortably full bladder, patients were asked to empty their bladder on arrival in the department, and then drink 350 ml of water. The CT scan was performed 1 h later. No specific measures were taken to influence the degree of rectal filling. CT scanning was performed with a GE HiSpeed QX/i scanner (GE Healthcare Diagnostic Imaging, Slough, UK) using a 4x2.5 mm helical scan acquisition reformatted to 2.5 mm slice thickness.

MRI scanning
The same patient positioning was employed for MRI as for CT. A customized perspex couch insert (Figure 1Go) was used to achieve a flat scanning surface. Patients were asked to empty their bladder and then drink 350 ml of water. Scanning took place 30 min later, rather than 1 h as for CT, in order to take into account the longer MRI scanning time. Scanning was performed on a 1.5 T Philips Intera Gyroscan (Philips Medical Systems, Reigate, UK) employing a phased array wrap-around pelvic coil. The posterior part of the coil was positioned under the flat-topped couch, and the anterior part placed over the patient's pelvis. In addition to standard clinical images, 3D T2 weighted images were acquired with the following scan parameters: repetition time (TR) = 3000 ms, echo time (TE) = 120 ms, field of view (FoV) = 400x300 mm, matrix = 256x144 reconstructed to 256x192, slice thickness = 2.5 mm, number of signal averages = 1, 60–70 slices (individualized for full coverage of bladder and rectum), read bandwidth = 651 Hz per pixel. To enable subsequent correction for distortion associated with magnetic field strength inhomogeneities, two sets of data were acquired with alternating positive and negative polarity read gradients [14].


Figure 1
View larger version (101K):
[in this window]
[in a new window]

 
Figure 1. Custom-made flat Perspex couch insert for the Philips MRI scanner.

 
Distortion correction
Geometric distortions result from system- and patient-based effects. System-based distortion arises from inhomogeneities in the main magnetic field (Bo) and from non-linearities in the applied magnetic field gradients. The presence of a patient within a scanner will further perturb the homogeneity of the main magnetic field, with the magnitude and geometry of this effect being specific to the individual patient. Subject-induced distortions arise from susceptibility and chemical shift variations within the individual patient and are of particular concern at fat/non-fat interfaces (chemical shift) and tissue/air interfaces (susceptibility).

The effects of both magnetic field inhomogeneities (system- and patient-based) and gradient non-linearities can be corrected retrospectively. Patients were scanned using a standard clinical 3D spin-echo sequence modified to acquire two sets of otherwise identical images: one acquired with a forward polarity read gradient and one with a reverse polarity read gradient. Corresponding features in the two sets of images were identified and their respective intensities and positions determined [14]. The images could then be corrected for the effects of distortion arising from magnetic field inhomogeneities (system- and subject-based) as a function of these intensities and positions [13, 14]. Gradient-based distortions associated with this sequence were mapped, quantified and subsequently corrected using a 3D test object of known linearity [12, 13].

Image co-registration
CT and distortion-corrected MRI images were transferred to a radiotherapy treatment planning system (Pinnacle 3; Philips Radiation Oncology Systems, Milpitas, CA). CT and MRI images were manually co-registered using bony anatomy. In all but one case, shifts of the MRI data set with respect to CT resulted in matching of the bony anatomy with good internal consistency. In one patient, a significant angulation of the pelvis on CT made matching difficult, and data from this patient were used only for volume comparisons.

Target definition
The prostate clinical target volume (CTV) was outlined by a radiologist (SAS) and clinical oncologist (ASNJ) working by consensus. For each patient, the CT images were outlined before the MRI, with a minimum interval of 1 week between the two. Readers were blinded to the other imaging modality when outlining.

Definition of organs at risk
The bladder, including the outer wall, and the anus/rectum, including the wall and contents from the ischial tuberosities to the rectosigmoid junction, were outlined using both MRI and CT. There is evidence that the component symptoms of the syndrome of late rectal radiation toxicity may arise from treatment of different regions of the rectum/anal canal, with bleeding originating from the rectum and incontinence from the anal canal [1520]. Anatomical descriptions of the anal canal suggest a usual length of 4 cm [2123]. We therefore analysed dose distributions from the anal canal (lower 4 cm) and the remainder of the rectum separately. The penile bulb was defined using MRI as the pear-shaped structure comprising the proximal part of the corpus spongiosum. The transition between the bulb and the more distal part of the corpus spongiosum was often not obvious. This observation has also been made by others [8]. If no clear transition was apparent, the bulb was defined as not extending any more anteriorly than the anterior part of the ischial bones. The distance between the most inferior part of the prostate (on CT and MRI) and the most superior part of the penile bulb (on MRI) was recorded.

Dosimetric planning
Radiotherapy plans were generated for each patient from CT- and MRI-defined targets. MRI was used to define the dose volume histogram (DVH) for OARs. It was recognized that differences in rectal filling between MRI and CT could, in effect, move the prostate between scans, and that this might be expected to influence the predicted dose to the rectum. To minimize any influence on predicted dose to OARs from differences in rectal filling between the two scans for each patient, patients were included in the planning comparison only if there was less than a 50% difference between CT and MRI of the rectal antero-posterior (AP) diameter at the level of the mid-prostate (defined on CT). As a measure of the effect of rectal filling on the position of the prostate, the relative positions of the most anterior part of the anterior rectal wall between MRI and CT were noted.

As MRI greyscale data cannot be used directly for inhomogeneity correction, the MRI-defined CTVs and OARs were copied onto the corresponding CT. This allowed calculation of a plan based on the MRI-derived prostate contour, using the electron density data from CT. For the prostate, planning target volumes (PTVs) were generated, consisting of the prostate plus a 5 mm margin symmetrically. A margin of 5 mm was chosen to reflect the anticipated technological developments in prostate radiotherapy that may control for organ movement and allow such restricted margins. The volume of the CTV and PTV for both MRI and CT was recorded, as was the length of the PTV in the cranio-caudal direction.

For each PTV, a three-field plan was generated using an anterior and two opposing wedged lateral fields. This is the standard field arrangement used in our department for prostate radiotherapy, and has been shown to be at least equivalent to arrangements using a greater number of fields in terms of rectal dose sparing [24]. For this exercise, blocks rather than a multi-leaf collimator (MLC) were chosen for beam shaping so that results were independent of MLC leaf width. A symmetrical block margin of 6 mm was used, with an additional 2 mm margin in the superior and inferior directions to account for penumbra.

The same field arrangement, wedge angles and beam weighting was used to cover the CT- and MRI-derived PTV within each patient. For an individual patient, the only difference between treatment techniques for the MRI- and CT-defined PTV was the field shape. For each plan, the PTV was treated to a notional dose of 70 Gy, representing the standard dose employed in our department at the time of the study, which was prescribed to the centre of the CT-defined PTV for each pair of plans. An adaptive convolution dose computation algorithm was used.

DVHs were generated from plans using CT- and MRI-defined target volumes and OARs as defined by MRI. For the rectum (as a whole, and also after separation into upper and lower parts) and the bladder, DVHs were expressed as the proportion of the whole structure receiving a given dose. For the penile bulb, DVH data were collected as the dose received by a given proportion of the structure. Maximum dose (Dmax) and the mean dose were also collected. The penile bulb parameters were chosen from studies relating DVH data to erectile function [2529].

Statistical analysis
Statistical analysis was carried out using SPSS (SPSS Inc, Chicago, IL) and Minitab (Minitab Ltd, Coventry, UK). Data were tested for normality using the Anderson–Darling test. Paired Student's t-tests (two-tailed) were used for volume and DVH comparisons for normally distributed datasets; otherwise, a Wilcoxon signed rank test was used (most V70 values, and penile bulb data). A p-value of <0.05 was taken as representing statistical significance.

Patient characteristics
Patient characteristics are shown in Table 1Go. All patients received neo-adjuvant luteinising hormone-releasing hormone agonist therapy prior to radiotherapy as part of their routine clinical management. CT was performed after a median of 12 weeks of hormonal treatment (range 5–38 weeks). The MRI scan took place a mean of 3 days after the CT (range 0–8 days).


View this table:
[in this window]
[in a new window]

 
Table 1. Patient characteristics

 

    Results
 Top
 Abstract
 Introduction
 Methods and patients
 Results
 Discussion
 Conclusions
 References
 
Organ volumes
Mean organ volumes are shown in Table 2Go. This includes all 14 patients for the prostate, seminal vesicles and rectum. Three patients did not have complete inclusion of the dome of the bladder on the MRI scan, and were therefore excluded from any further analysis pertaining to the bladder. There was considerable variation in the volume of the rectum and bladder between MRI and CT for each patient, mainly representing differences in organ filling.


View this table:
[in this window]
[in a new window]

 
Table 2. Organ volumes

 
For the prostate, the MRI volume was smaller than the CT volume in 12 cases and larger in 2 (Figure 2Go). For the 13 cases in whom satisfactory co-registration between MRI and CT was achieved, the apex of the prostate was visualized as lying more inferiorly on CT than on MRI in seven cases (2.5–10 mm), 5 mm lower on MRI in one case, and the same in five cases. Overall, mean and median inferior displacement of the apex was 2.5 mm on CT compared with MRI. Bladder volume differences between CT and MRI were unrelated to prostate apex localization (r = –0.0005). There was no consistent difference between imaging techniques with regard to the AP position of the gland, which appeared to be more associated with the degree of rectal filling.


Figure 2
View larger version (26K):
[in this window]
[in a new window]

 
Figure 2. Volume of prostate(cm3) as defined by CT and MRI for all 14 patients.

 
The mean and median distances, respectively, between the most inferior part of the prostate and the penile bulb were 11.6 mm and 12.5 mm for CT (range 5–17.5 mm) and 14.1 mm and 15 mm for MRI (range 5–20 mm) (p = 0.08), in keeping with the more caudal definition of the prostate apex on CT.

Two patients were excluded from the radiotherapy planning study on the basis of a greater than 50% difference in the AP dimension of the rectum at the level of the mid-prostate. There remained differences in the dimensions of the rectum between CT and MRI, and thus the position of the prostate for the remaining 11 patients. The mean difference between the position of the most anterior part of the rectal wall between MRI and CT was such that, on average, the anterior rectal wall at the level of the mid-prostate on MRI lay 0.3 mm anterior to that on CT (range, MRI 5 mm anterior to 9 mm posterior to CT; standard deviation, 3.85 mm)

Radiotherapy plan comparisons
11 patients were used for the radiotherapy planning study. Of the original 14 patients, one was excluded, as satisfactory co-registration of CT and MRI could not be achieved, and two were excluded on the basis of large differences between the rectal diameter on MRI and CT. The characteristics of the target volumes for the 11 patients are shown in Table 3Go. It was observed that copying an MRI-defined structure to CT resulted in a small increase in the volume of that structure of approximately 2%. This effect results from different voxel sizes between MRI and CT, which the planning software accounts for by rounding the volume to the nearest CT voxel. Bladder DVH data are presented for eight patients, having excluded those three with incomplete inclusion of the bladder on MRI. For the penile bulb, DVH data are presented for 9 patients, comprising those 11 used for the rectal comparison, but excluding 2 patients in whom the most inferior extent of the penile bulb was not imaged on MRI.


View this table:
[in this window]
[in a new window]

 
Table 3. Target volume characteristics for 11 patients in the dosimetric planning study

 
The minimum requirement that 100% of the target volume is encompassed by the 95% iso-dose line for adequate PTV coverage, as defined by the International Commission on Radiation Units, was met for all plans.

DVH parameters for the whole rectum, upper rectum, anal canal and bladder from plans treating the prostate defined by CT and MRI are shown in Table 4Go.


View this table:
[in this window]
[in a new window]

 
Table 4. Dose volume histogram characteristics for the rectum, anal canal, upper rectum and bladder(defined by MRI) for plans treating the prostate defined by CT and MRI

 
A lower proportion of the rectum was irradiated to doses from 45 Gy to 65 Gy (p<0.05) using the MRI-derived plan. At these doses, 4–6% less of the rectum received a given dose with MRI when compared with CT. When the rectum was divided into upper and lower (anal canal) sections, there was a significantly reduced dose to the anal canal, but not the remaining rectum, using the MRI plan. There was no significant difference in the volume of bladder treated to any of the dose levels between CT and MRI. For the penile bulb, there was no significant difference in all but one (dose to 90% of the penile bulb) of the parameters examined.


    Discussion
 Top
 Abstract
 Introduction
 Methods and patients
 Results
 Discussion
 Conclusions
 References
 
We found that the prostate CTV defined by distortion-corrected MRI was significantly smaller than with CT. Treatment plans using a 5 mm margin from CTV to PTV showed a statistically significant reduction in the predicted proportion of the rectum treated to dose levels from 45–65 Gy using the plan derived from MRI. This effect was most apparent for the lower rectum (anal canal). We saw no difference in the dose to bladder or penile bulb between plans.

Other studies [110] have compared CT and MRI prostate definition, but few have attempted to quantify the differences between predicted doses to OARs [2, 6, 8, 9] and none has compared distortion-corrected MRI images with CT. Khoo et al [1] described how MRI provides improved definition of prostate treatment volumes compared with CT. The study examined a number of MRI sequences and scored them as to the ease of visualization of the prostate and other pelvic structures, but did not quantify or compare target volumes between imaging modalities.

Prostate volumes from published studies comparing CT and MRI are shown in Table 5Go. Roach et al [7] defined the prostate alone on CT and on MRI. The images were merged using bony landmarks. They found a reduction of 23% in the prostate volume with MRI. CT most overestimated the posterior and inferior aspects of the gland.


View this table:
[in this window]
[in a new window]

 
Table 5. Prostate volumes defined by CT and MRI from published literature

 
Kagawa et al [10] outlined the prostate with and without the seminal vesicles of 22 patients on CT and MRI. Images were co-registered using bony landmarks. The average prostate volume was 63 cm3 and 50.9 cm3 — a reduction of 19% with MRI. The prostatic apex lay 4.9 mm more cephalad and 1.4 mm dorsal on MRI than on CT, at a distance of 15.1 mm cephalad and 7.1 mm dorsal to the tip of the urethrogram cone.

In the study by Debois et al [9], 10 patients underwent CT as well as axial and coronal MRI scans. Images were mathematically aligned (the methodology was not described) and the prostate and rectum contoured using each modality by three observers. The prostate volume was smaller using MRI for 9 of the 10 patients studied. MRI gave a mean volume 28% smaller than CT. In all instances, the prostatic apex was defined more caudally, or the same, with CT when compared with MRI. The authors felt that coronal images were an essential adjunct to axial images on MRI in defining the prostatic apex. The posterior extent of the prostate gland and anterior rectal wall was better defined with MRI, and the volume of the rectum irradiated to 80% of the prescribed dose (Vrectum80%) was less using MRI than CT in 8 out of 10 cases, and greater in 2. For all patients, the overall difference in Vrectum80% between CT and MRI did not reach statistical significance.

Sannazzari et al [2] outlined the prostate and seminal vesicles of eight patients using CT and MRI. Defining the prostate using MRI resulted in a volume reduction of 19% (p<0.01). A planning study suggested that it was possible to spare 10% of the rectal, and 5% of the femoral head and bladder, volumes from the radiation field using MRI rather than CT, although the radiotherapy treatment technique was not described.

Krempien et al [6] demonstrated a 40% reduction in the volume of the prostate, including seminal vesicles, on MRI compared with CT. This resulted in a reduction of the mean percentage rectal dose by 15%, but again details of the treatment technique were not provided.

Parker et al [5] described planning studies that compared MRI and CT. Intra-prostatic fiducial markers were used for the purposes of image co-registration, thus minimizing prostate movement as a variable. There was greater intra-observer variation using CT outlines compared with MRI but, in contrast to earlier reports, they did not find a consistent prostate volume difference between CT and MRI. They speculated that this resulted from awareness of earlier studies, leading to a more conservative outlining approach on CT. Again, overestimation of the posterior margin at the base of the prostate and underestimation at the apex by CT was seen.

In the study by Rasch et al [3], the mean volume of the prostate derived from CT was 63.7 cm3 compared with 44.5 cm3, 40.7 cm3 and 37.4 cm3 using axial, coronal and sagittal MRI, respectively. Axial MRI therefore results in a 30% reduction in the contoured prostate volume compared with CT. The most marked areas of disagreement were at the apex and the base of the seminal vesicles, where differences of 6 mm caudally and 7 mm cranially, respectively, were shown. A subsequent planning study [8] using these data showed that, with 10 mm margins, MRI delineation of the prostate leads to a significantly reduced predicted dose to the rectal wall compared with CT. This could be expressed as an equivalent uniform dose to the rectal wall of 64 Gy vs 65.6 Gy for MRI- and CT-defined targets, respectively. Furthermore, MRI delineation of the prostate resulted in significantly reduced predicted dose to the penile bulb compared with CT (mean dose 7.9 Gy vs 19.5 Gy).

In summary, CT may overestimate the volume of the prostate compared with MRI, which can result in prostate volumes up to 40% smaller than with CT. Again, CT most overestimated the inferior and posterior extents of the gland. The apex of the gland as defined by MRI is reported to lie posterior and cranial to its apparent location on CT, a consistent finding in most studies but contradicting work by Milosevic et al [4], who found variation between MRI and CT in defining the apex but not consistently in any direction. This may have been as a result of the 1 cm slice thickness used for CT, which would be less able to detect sub-centimetre differences in apical position. In the current study, using a slice thickness of 2.5 mm for both imaging modalities, we have demonstrated a 13% reduction in the prostate volume using MRI when compared with CT. The reduction we saw with MRI was less than for much of the foregoing work; this may be as a result of the improved quality of CT images over recent years. Also, as described by Parker et al [5], experience with MRI may improve the interpretation of CT, modifying and reducing the volume of prostate outlined.

We found that the prostatic apex tends be located more caudally on CT than on MRI, in keeping with the majority of published studies. In contrast to the work by Steenbakkers et al [8], who employed a 10 mm margin inferiorly, we did not find a significant difference in predicted dose to the penile bulb. The dose to this structure was low in the current study for both MRI- and CT-defined target volumes, probably as a result of the 5 mm margins used, which excluded much of the penile bulb from the PTV. It is possible that a difference in dose to the penile bulb between MRI and CT plans may have become apparent with the use of larger margins inferiorly.

Like others [2, 6, 8, 9], we have found that, compared with CT, MRI delineation of the prostate results in treatment plans with a lower predicted dose to the rectum. We also found that the dose-sparing effect was confined to the lower rectum. A direct comparison with these studies is not possible, as different methods of expressing rectal dose were used.

If distortion-corrected MRI is to replace CT as the chosen modality for prostate radiotherapy planning, then account needs to be taken of the lack of electron density information from MRI. From our centre, Lee et al [11] have shown the feasibility of assigning bulk electron density data to pelvic MRI images for prostate radiotherapy planning. Using their technique, less than 2% difference in dose was found between MRI and CT in the high-dose regions of the plans. Chen et al [30] contend that no account needs to be taken of tissue heterogeneity for pelvic radiotherapy planning. In their study, in addition to planning CT, MRI images were acquired on a low-field open scanner and corrected for gradient distortion. Plans based on MRI and CT, using internal contours defined using MRI, were produced. The purpose of the study was to examine the effect of residual geometric distortion on the dosimetry of radiotherapy plans produced using MRI compared with CT, and without using tissue heterogeneity correction. Less than 2% dose difference was seen between MRI- and CT-derived plans.

A criticism of the current study is that no measures were taken to achieve a consistent rectal volume for CT and MRI scans. This led to differences in the position of the anterior rectal wall and, consequently, the position of the prostate between MRI and CT within each patient. In our current clinical radiotherapy practice, we do not attempt to control rectal filling and did not seek to do so for the purpose of this study. Some studies have employed enemas prior to scanning [8, 9] and/or have performed both scans on the same day. In the study by Parker et al [5], the use of intra-prostatic fiducial markers to match the position of the prostate between scans would have minimized the effect of organ movement. Although apparent differences in the AP position of the prostate, resulting from changes in rectal filling as visualized on CT and MRI, could affect predictions of rectal dose, we found little systematic difference between imaging modalities, and on average the rectal wall lay 0.3 mm more anterior on MRI than on CT.

In the current study, we employed the same patient positioning for MRI as for CT, including a flat-topped couch, in order to achieve comparability between imaging modalities. In the study by Steenbakkers et al [8], a consistent difference in the position of the anterior rectal wall and posterior prostate was noted such that the MRI-defined anterior rectal wall at the level of the mid-prostate lay further ventrally (anteriorly) than on CT. This was attributed to differences in the shape of the scanning tabletop between imaging modalities and the use of supports behind the knees for CT but not for MRI.

A number of other points should be considered when interpreting the current study. Firstly, this is a planning study, and doses to OARs are predictions, although evidence exists to support the predictive value of DVH data for the rectum in terms of radiation sequelae [3135]. Secondly, the clinical significance of the ~10% absolute reduction in the volume of the lower rectum/anal canal treated to doses of 45–65 Gy is uncertain.

Three patients in this study were excluded from analysis because of changes in the rectal diameter, and consequently prostate position between scans. This serves as a reminder of the phenomenon of inter-fraction movement. Dietary modification or laxatives are used in an attempt to achieve consistent rectal filling, but there is at present a lack of published evidence to support such measures routinely. Other measures include intra-prostatic fiducial markers [3638], intrarectal devices [39] and image guidance using cone-beam CT [40]. If the potential benefits of improved target definition are to be realized, then consideration should be given to employing MRI with such techniques.


    Conclusions
 Top
 Abstract
 Introduction
 Methods and patients
 Results
 Discussion
 Conclusions
 References
 
We have shown that it is feasible to use distortion-corrected T2 weighted MRI images from a standard clinical 1.5 T MRI scanner for prostate radiotherapy planning. These geometrically faithful MRI images could be used directly for radiotherapy planning after accounting for electron density. This is the first study to compare distortion-corrected T2 weighted MRI with CT in this setting. The smaller prostate volume defined with MRI compared with CT is in keeping with previous studies using uncorrected MRI. We have demonstrated that this results in a reduction of the predicted radiotherapy dose to the rectum, in particular the anal canal. The clinical significance of this remains uncertain.


    Acknowledgments
 
This work was undertaken in The Royal Marsden NHS Trust which received a proportion of its funding from the NHS Executive; the views expressed in this publication are those of the authors and not necessarily those of the NHS Executive. This work was supported by The Department of Health New and Emerging Applications of Technology (NEAT) program, grant number B132, the Institute of Cancer Research, the Bob Champion Cancer Trust and Cancer Research UK Section of Radiotherapy (CUK) grant number C46/A2131.

Received for publication March 9, 2007. Revision received April 26, 2007. Accepted for publication June 5, 2007.


    References
 Top
 Abstract
 Introduction
 Methods and patients
 Results
 Discussion
 Conclusions
 References
 

  1. Khoo VS, Padhani AR, Tanner SF, Finnigan DJ, Leach MO, Dearnaley DP. Comparison of MRI with CT for the radiotherapy planning of prostate cancer: a feasibility study. Br J Radiol 1999;72:590–7.[Abstract]
  2. Sannazzari GL, Ragona R, Ruo Redda MG, Giglioli FR, Isolato G, Guarneri A. CT–MRI image fusion for delineation of volumes in three-dimensional conformal radiation therapy in the treatment of localized prostate cancer. Br J Radiol 2002;75:603–7.[Abstract/Free Full Text]
  3. Rasch C, Barillot I, Remeijer P, Touw A, van Herk M, Lebesque JV. Definition of the prostate in CT and MRI: a multi-observer study. Int J Radiat Oncol Biol Phys 1999;43:57–66.[CrossRef][Medline]
  4. Milosevic M, Voruganti S, Blend R, Alasti H, Warde P, McLean M, et al. Magnetic resonance imaging (MRI) for localization of the prostatic apex: comparison to computed tomography (CT) and urethrography. Radiother Oncol 1998;47:277–84.[CrossRef][Medline]
  5. Parker CC, Damyanovich A, Haycocks T, Haider M, Bayley A, Catton CN. Magnetic resonance imaging in the radiation treatment planning of localized prostate cancer using intra-prostatic fiducial markers for computed tomography co-registration. Radiother Oncol 2003;66:217–24.[CrossRef][Medline]
  6. Krempien RC, Schubert K, Zierhut D, Steckner MC, Treiber M, Harms W, et al. Open low-field magnetic resonance imaging in radiation therapy treatment planning. Int J Radiat Oncol Biol Phys 2002;53:1350–60.[CrossRef][Medline]
  7. Roach M, 3rd, Faillace-Akazawa P, Malfatti C, Holland J, Hricak H. Prostate volumes defined by magnetic resonance imaging and computerized tomographic scans for three-dimensional conformal radiotherapy. Int J Radiat Oncol Biol Phys 1996;35:1011–8.[CrossRef][Medline]
  8. Steenbakkers RJ, Deurloo KE, Nowak PJ, Lebesque JV, van Herk M, Rasch CR. Reduction of dose delivered to the rectum and bulb of the penis using MRI delineation for radiotherapy of the prostate. Int J Radiat Oncol Biol Phys 2003;57:1269–79.[CrossRef][Medline]
  9. Debois M, Oyen R, Maes F, Verswijvel G, Gatti G, Bosmans H, et al. The contribution of magnetic resonance imaging to the three-dimensional treatment planning of localized prostate cancer. Int J Radiat Oncol Biol Phys 1999;45:857–65.[CrossRef][Medline]
  10. Kagawa K, Lee WR, Schultheiss TE, Hunt MA, Shaer AH, Hanks GE. Initial clinical assessment of CT-MRI image fusion software in localization of the prostate for 3D conformal radiation therapy. Int J Radiat Oncol Biol Phys 1997;38:319–25.[CrossRef][Medline]
  11. Lee YK, Bollet M, Charles-Edwards G, Flower MA, Leach MO, McNair H, et al. Radiotherapy treatment planning of prostate cancer using magnetic resonance imaging alone. Radiother Oncol 2003;66:203–16.[CrossRef][Medline]
  12. Tanner SF, Finnigan DJ, Khoo VS, Mayles P, Dearnaley DP, Leach MO. Radiotherapy planning of the pelvis using distortion corrected MR images: the removal of system distortions. Phys Med Biol 2000;45:2117–32.[CrossRef][Medline]
  13. Doran SJ, Charles-Edwards L, Reinsberg SA, Leach MO. A complete distortion correction for MR images. I. Gradient warp correction. Phys Med Biol 2005;50:1343–61.[CrossRef][Medline]
  14. Reinsberg SA, Doran SJ, Charles-Edwards EM, Leach MO. A complete distortion correction for MR images. II. Rectification of static-field inhomogeneities by similarity-based profile mapping. Phys Med Biol 2005;50:2651–61.[CrossRef][Medline]
  15. Fokdal L, Honore H, Hoyer M, von der Maase H. Dose–volume histograms associated to long-term colorectal functions in patients receiving pelvic radiotherapy. Radiother Oncol 2005;74:203–10.[CrossRef][Medline]
  16. Heemsbergen WD, Hoogeman MS, Hart GA, Lebesque JV, Koper PC. Gastrointestinal toxicity and its relation to dose distributions in the anorectal region of prostate cancer patients treated with radiotherapy. Int J Radiat Oncol Biol Phys 2005;61:1011–8.[CrossRef][Medline]
  17. Vordermark D, Schwab M, Ness-Dourdoumas R, Sailer M, Flentje M, Koelbl O. Association of anorectal dose–volume histograms and impaired fecal continence after 3D conformal radiotherapy for carcinoma of the prostate. Radiother Oncol 2003;69:209–14.[CrossRef][Medline]
  18. Berndtsson I, Lennernas B, Hulten L. Anorectal function after modern conformal radiation therapy for prostate cancer: a pilot study. Tech Coloproctol 2002;6:101–4.[CrossRef][Medline]
  19. Koper PC, Heemsbergen WD, Hoogeman MS, Jansen PP, Hart GA, Wijnmaalen AJ, et al. Impact of volume and location of irradiated rectum wall on rectal blood loss after radiotherapy of prostate cancer. Int J Radiat Oncol Biol Phys 2004;58:1072–82.[Medline]
  20. Mavroidis P, Al-Abany M, Helgason AR, Agren Cronqvist AK, Wersall P, Lind H, et al. Dose–response relations for anal sphincter regarding fecal leakage and blood or phlegm in stools after radiotherapy for prostate cancer radiobiological study of 65 consecutive patients. Strahlenther Onkol 2005;181:293–306.[CrossRef][Medline]
  21. Sinnatamby CS. Last's anatomy, 10th edn. London, UK: Churchill Livingstone; 1998
  22. Gray's anatomy, 38th edn. Edinburgh, UK: Churchill Livingstone; 1995
  23. Snell RS. Clinical anatomy for medical students, 3rd edn. Boston: Little, Brown and Company; 1986
  24. Khoo VS, Bedford JL, Webb S, Dearnaley DP. Class solutions for conformal external beam prostate radiotherapy. Int J Radiat Oncol Biol Phys 2003;55:1109–20.[CrossRef][Medline]
  25. Merrick GS, Butler WM, Wallner KE, Lief JH, Anderson RL, Smeiles BJ, et al. The importance of radiation doses to the penile bulb vs. crura in the development of postbrachytherapy erectile dysfunction. Int J Radiat Oncol Biol Phys 2002;54:1055–62.[CrossRef][Medline]
  26. Wernicke AG, Valicenti R, Dieva K, Houser C, Pequignot E. Radiation dose delivered to the proximal penis as a predictor of the risk of erectile dysfunction after three-dimensional conformal radiotherapy for localized prostate cancer. Int J Radiat Oncol Biol Phys 2004;60:1357–63.[CrossRef][Medline]
  27. Fisch BM, Pickett B, Weinberg V, Roach M. Dose of radiation received by the bulb of the penis correlates with risk of impotence after three-dimensional conformal radiotherapy for prostate cancer. Urology 2001;57:955–9.[CrossRef][Medline]
  28. Roach M, Winter K, Michalski JM, Cox JD, Purdy JA, Bosch W, et al. Penile bulb dose and impotence after three-dimensional conformal radiotherapy for prostate cancer on RTOG 9406: findings from a prospective, multi-institutional, phase I/II dose-escalation study. Int J Radiat Oncol Biol Phys 2004;60:1351–6.[CrossRef][Medline]
  29. Mangar SA, Sydes MR, Tucker HL, Coffey J, Sohaib SA, Gianolini S, et al. Evaluating the relationship between erectile dysfunction and dose received by the penile bulb: using data from a randomised controlled trial of conformal radiotherapy in prostate cancer (MRC RT01, ISRCTN47772397). Radiother Oncol 2006;80:355–62.[CrossRef][Medline]
  30. Chen L, Price RA Jr, Wang L, Li J, Qin L, McNeeley S, et al. MRI-based treatment planning for radiotherapy: dosimetric verification for prostate IMRT. Int J Radiat Oncol Biol Phys 2004;60:636–47.[CrossRef][Medline]
  31. Fiorino C, Cozzarini C, Vavassori V, Sanguineti G, Bianchi C, Cattaneo GM, et al. Relationships between DVHs and late rectal bleeding after radiotherapy for prostate cancer: analysis of a large group of patients pooled from three institutions. Radiother Oncol 2002;64:1–12.[CrossRef][Medline]
  32. Boersma LJ, van den Brink M, Bruce AM, Shouman T, Gras L, te Velde A, et al. Estimation of the incidence of late bladder and rectum complications after high-dose (70–78 GY) conformal radiotherapy for prostate cancer, using dose–volume histograms. Int J Radiat Oncol Biol Phys 1998;41:83–92.[CrossRef][Medline]
  33. Jackson A, Skwarchuk MW, Zelefsky MJ, Cowen DM, Venkatraman ES, Levegrun S, et al. Late rectal bleeding after conformal radiotherapy of prostate cancer. II. Volume effects and dose–volume histograms. Int J Radiat Oncol Biol Phys 2001;49:685–98.[CrossRef][Medline]
  34. Storey MR, Pollack A, Zagars G, Smith L, Antolak J, Rosen I. Complications from radiotherapy dose escalation in prostate cancer: preliminary results of a randomized trial. Int J Radiat Oncol Biol Phys 2000;48:635–42.[CrossRef][Medline]
  35. Wachter S, Gerstner N, Goldner G, Potzi R, Wambersie A, Potter R. Rectal sequelae after conformal radiotherapy of prostate cancer: dose–volume histograms as predictive factors. Radiother Oncol 2001;59:65–70.[CrossRef][Medline]
  36. D'Amico AV, Manola J, Loffredo M, Lopes L, Nissen K, O'Farrell DA, et al. A practical method to achieve prostate gland immobilization and target verification for daily treatment. Int J Radiat Oncol Biol Phys 2001;51:1431–6.[CrossRef][Medline]
  37. Dehnad H, Nederveen AJ, van der Heide UA, van Moorselaar RJ, Hofman P, Lagendijk JJ. Clinical feasibility study for the use of implanted gold seeds in the prostate as reliable positioning markers during megavoltage irradiation. Radiother Oncol 2003;67:295–302.[CrossRef][Medline]
  38. Pouliot J, Aubin M, Langen KM, Liu YM, Pickett B, Shinohara K, et al. (Non)-migration of radiopaque markers used for on-line localization of the prostate with an electronic portal imaging device. Int J Radiat Oncol Biol Phys 2003;56:862–6.[CrossRef][Medline]
  39. Wachter S, Gerstner N, Dorner D, Goldner G, Colotto A, Wambersie A, et al. The influence of a rectal balloon tube as internal immobilization device on variations of volumes and dose–volume histograms during treatment course of conformal radiotherapy for prostate cancer. Int J Radiat Oncol Biol Phys 2002;52:91–100.[CrossRef][Medline]
  40. Letourneau D, Martinez AA, Lockman D, Yan D, Vargas C, Ivaldi G, et al. Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patients. Int J Radiat Oncol Biol Phys 2005;62:1239–46.[CrossRef][Medline]




This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jackson, A S N
Right arrow Articles by Dearnaley, D P
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jackson, A S N
Right arrow Articles by Dearnaley, D P


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS