| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Full paper |
1 Medical Physics Unit and 2 Radiology Department, General Hospital 'Konstantopoulio-Agia Olga', Athens, 3 Greek Atomic Energy Commission, Agia Paraskevi, Athens, and 4 Department of Medical Physics, Medical School, University of Athens, Athens, Greece
Correspondence: Dr Virginia Tsapaki, Medical Physics Unit, General Hospital 'Konstantopoulio-Agia Olga', 3-5 Agias Olgas Str., Nea Ionia 14233, Athens, Greece. E-mail: virginia{at}otenet.gr
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
During recent years, patient dose has become a major issue and because of the increasing awareness and greater realisation of the effects of ionizing radiation, X-ray users are now more demanding of dose information and dose reduction [6]. The "European Guidelines on Quality Criteria for Diagnostic Radiographic Images" document defines the diagnostic requirements for normal, basic radiographs, specifying anatomical image criteria and important image details; it indicates criteria for the radiation dose to the patient and gives examples for good radiographic technique by which the diagnostic requirements and dose criteria can be achieved [7]. Two types of dosemeter are commonly used for estimating ESD to patients during radiographic examinations, namely thermoluminescent dosemeters (TLDs) and ionization chambers [8]. TLDs have the advantage of being physically small, enabling them to be stuck directly and unobtrusively on the patient's skin with very little interference in patient mobility or comfort. They will fully measure the radiation backscattered from the patient, an essential component of the entrance surface dose (ESD), and are unlikely to obscure useful diagnostic information. Ionization chambers, being more bulky and requiring connecting cables, are usually difficult to attach in sufficiently close contact to the patient's skin to ensure complete measurement of the backscattered radiation. They also severely restrict patient mobility and cast interfering shadows on radiographs. They are consequently not recommended for direct measurement of ESD. Ionization chambers can, however, be used to make measurements of the absorbed dose to air, in free air, on the axis of the X-ray beam without a patient or phantom present. Such measurements can be corrected using appropriate backscatter factors and the inverse square law to estimate the ESD. TLDs are recommended for direct measurement of ESD and are available in a variety of physical forms and in different materials. The National Radiological Protection Board (NRPB) recommends individual chips or pellets of lithium fluoride or lithium borate [8].
Certain studies are found in the literature that use indirect methods to estimate ESD based on free-in-air measurements of tube output and calculations using patient exposure factors [911]. Two of these studies [10, 11] compared the indirect method with the direct method of TLD and found good correlation of the results. However, the limitation of these studies was that output measurements were performed on either one tube potential [10] or specific tube potentials [9, 11]. In the case where the output was not measured in the tube potential utilized in the specific patient, it was derived from the value measured at the nearest possible potential, assuming that it was proportional to kVp2 [11].
The objectives of the current study were to derive a mathematical method for calculating the ESD from exposure factors for all tube potentials used in clinical practice and compare the calculated values with the ESDs determined using TLDs kindly provided and measured by the Greek Atomic Energy Commission (GAEC).
| Methods and materials |
|---|
|
|
|---|
Quality assurance (QA) measurements were carried out with a PMX-III R/CT digital multimeter (RTI Electronics, Mölndal, Sweden) comprising a R25 solid state detector with a calibration traceable to a primary standard. According to the manufacturer specifications this instrument is accurate to within 3% for output measurements and has an energy dependence between 0.99 to 1.01 for tube potentials in the range of 50 kV to 140 kV. Tube and generator QA measurements were carried out in accordance with the IPEMB 32 (Part I) protocol [12] and AEC QA checks according to the IPEMB 32 (Part IV) protocol [13].
The results of the QA measurements showed that:
ESD calculation method
A number of equations are found in the literature for describing the output variation with tube potential. Simpkin and Dixon [14] have used a third order polynomial for the output of a radiography tube and a second order polynomial for a Mo/Mo mammography tube. They noted, however, that a simple power relationship of the form O/P = a(kVp)b could also describe the output from the specific radiographic tube to within ±5%. The latter equation is consistent with the linear relationship between the logarithms of output and tube potential mentioned in the IPEMB report [12]. George et al [11] have used the same power low relationship setting however, with the b factor equal to two, following the commonly adopted concept of output variation with the square of tube potential.
In this study, the output values measured across the range 50140 kVp were fitted by all four relationships mentioned above and the results are given in Table 1
. It is obvious that the third and second order polynomials provide the best fit with the smallest mean and maximum errors compared with the other two equations.
|
Taking all of the above into account, it was decided that the output of the X-ray tube will be described by the second order polynomial given by the following equation, with fitting parameters a = 0.0099, b = 1.1594 and c = -44.75.
|
|
Having established the output variation with tube potential, the ESD can then be calculated using the following equation:
|
|
ESD measurement procedure
Validation of the calculation method previously described was undertaken using TLDs (Harshaw TLD-100 rods: Harshaw-Bicron, Solon, OH [LiF:Mg,Ti, 3x3x9 mm]) provided by the GAEC. TLD calibrations were carried out in the GAEC Secondary Standard Dosimetry Laboratory using ISO Narrow 4037 reference quality beams. The TLDs were read using a Harshaw 4500 TLD reader and annealed in a PTW oven (PTW Company, Freiburg, Germany). The TLD energy response was ±18% across the range 50150 kVp, the uncertainty of measurement was estimated to be less than ±10% and the minimum detectable dose 30 µGy.
The five most frequently performed diagnostic radiographic examinations were included in the study; posteroanterior (PA), chest and lateral (LAT) chest, anteroposterior (AP) abdomen (erect and supine) and AP urinary tract. 43 patients took part in the study; 13 patients had both PA and LAT chest radiographs, 10 patients had AP abdomen erect radiographs, 10 patients had AP abdomen supine radiographs and 10 patients had AP urinary tract radiographs. The minimum number of patients for each examination was 10, based on the requirements given by the European Commission guidelines (EG) [7]. For each radiographic projection, the mean patient weight was within the range of 70 kg ±5. For each radiograph, the tube potential, milliampere settings, FSD, FFD, cassette size, patient weight and age were recorded. The image quality of all radiographic examinations included in the sample was satisfactory according to the radiologists of the department and fulfilled all image criteria set by the European guidelines [7].
| Results |
|---|
|
|
|---|
|
|
Table 3
presents the comparison of technical parameters (tube potential and FFD) and ESDC with the corresponding recommendations of the EG [7]. The only examination for which no recommendations were found in these guidelines was the abdomen radiograph. All remaining procedures fall within the EG suggestions, with the mean ESDC in all radiographic examinations being substantially lower than European Guidelines dose reference levels (EG DRL). More specifically, mean ESDC for PA chest is 1.7 times lower than the EG DRL, for LAT chest mean ESDC is 3.0 times lower and for urinary tract 7.2 times lower than the EG DRL. Mean ESDC values found in this work were also lower than the DRLs recently proposed by NRPB [20] given in the same table. It is evident that the low dose levels in Konstantopoulio-Agia Olga General Hospital can probably be attributed to the technical parameters used (for example, the high tube potential in chest examinations) and the use of a high speed class screenfilm combination.
|
| Discussion |
|---|
|
|
|---|
Thus, the slight differences found between ESDC and ESDTLD, for all examination types except PA chest, can be explained by the uncertainties of the measurements using either method. For PA chest, further investigation was required to test whether the measurement or the calculation method provided the most reliable ESD values. For this reason, the following experiment was executed: PA chest clinical conditions were simulated using a 1 mm Cu plate placed on the centre of the X-ray field, so as to cover the central chamber of AEC and produce similar mAs values with PA chest. In order to measure ESD the dosemeter was placed beside the phantom, so as not to block the AEC chamber. A series of exposures was performed with the central AEC chamber activated and the corresponding milliampere, milliseconds settings and ESD were recorded along with the optical density (OD) of the film. In all these exposures, the exposure times selected by the AEC system were lower than 4 ms; that is, comparable with the generator "switching time". In such short exposure times the stability of the generator is questionable and the mAs value indicated may not be accurate, while it is also probable that the pre-set tube potential value cannot be obtained. Indeed, it was found that the mean ESD was smaller than the theoretically predicted value by approximately 40% while the optical density of the film was 1.4 OD.
The experiment was repeated by manually setting the tube current and the exposure time (values selected were always higher than 10 ms) so that similar values of mAs were applied, as before. This time, the ESD measured by the dosemeter was in agreement with the theoretical value calculated using Equation (1) and the optical density of the film was 2.7. The almost-double OD of the film eliminated the possibility that the reduced output measured with the dosemeter was due to false dosemeter indication in such short exposure times.
With these additional measurements, it was evident that below a certain exposure time (for this specific unit it seemed to be approximately 10 ms), the output was no longer consistent with the corresponding output values of longer exposure times even if the milliampere settings were equal. Thus the loss of linearity observed was not a matter of low milliampere settings, but of very short exposure times a fact that had not been initially accounted for in the routine quality control. While in the QA tests a very good output linearity has been observed for a large range of milliampere settings, it was now evident that this linearity was not maintained at the very short exposure times observed in the PA chest examination. Since reproducibility problems of both the generator and the AEC measuring chamber may also appear in such short exposure times, it would be preferable to slightly lower the tube potential selection so that the exposure time in clinical conditions will be higher than the value below which the stability of the generator is not assured. In some generators there is also the alternative option of pre-selecting a low tube current value, of around 100 mA, so as to ensure that the exposure times under AEC operation will be longer than 10 ms.
Taking into account the output reduction observed for short exposure times, it may be considered that the ESDTLD values are more reliable than was initially thought. Still, since the output reduction increases with decreasing exposure times, it was not possible for us to verify the TLD accuracy. This, however, could be achieved using Plexiglas® plates to reproduce the clinically observed milliampere settings along with a calibrated dosemeter capable of fully recording the backscattered radiation.
| Conclusion |
|---|
|
|
|---|
More detailed investigation should be carried out in very small ESD values (around 100 µSv) for the following reasons: (1) this dose region is close to the minimum detectable dose level of commonly used TLD; (2) this dose level is observed in the PA chest examination, which is the most frequent X-ray examination in all diagnostic radiological departments; and (3) it is highly probable that X-ray systems are not reliable for very short exposure times (smaller than 10 ms) when a high tube potential technique is used (as recommended in the EG).
Received for publication November 30, 2005. Revision received June 8, 2006. Accepted for publication June 27, 2006.
| References |
|---|
|
|
|---|
ko K. First results on patient dose measurements from conventional diagnostic radiology procedures in Serbia and Montenegro. Radiat Prot Dosim 2005;113:3305.This article has been cited by other articles:
![]() |
W. E. Muhogora, N. A. Ahmed, A. Almosabihi, J. S. Alsuwaidi, A. Beganovic, O. Ciraj-Bjelac, F. K. Kabuya, A. Krisanachinda, M. Milakovic, G. Mukwada, et al. Patient Doses in Radiographic Examinations in 12 Countries in Asia, Africa, and Eastern Europe: Initial Results from IAEA Projects Am. J. Roentgenol., June 1, 2008; 190(6): 1453 - 1461. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |