| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Full paper |
Medical Radiation Science, School of Health Sciences, Faculty of Health, University of Newcastle, University Drive, Callaghan, NSW 2308, Australia
Correspondence: Dr Helen M Warren-Forward, Medical Radiation Science, Faculty of Health, University of Newcastle, Box 16, Hunter Building, University Drive, Callaghan, Newcastle, NSW 2308, Australia. E-mail: helen.warren-forward{at}newcastle.edu.au
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Since the discovery of X-rays by Roentgen in 1895, film has usually been the universal detector in diagnostic radiography. Technological advances in image acquisition techniques have been evolving over the past two decades with the introduction of the first Fuji 101 (Fujifilm; Minato-Ku, Tokyo, Japan) computed radiography (CR) system in 1983. By 1992, other manufacturers (Kodak (Australasia) Pty. Ltd, Coburg, Victoria, 3058 Australia and Agfa, Agfa Healthcare, Burwood, Victoria, 3125) entered the CR market place [3].
The main advantages of CR include: wider dynamic range (latitude), ability for post processing (manipulation of radiographic contrast and brightness), multiple viewing options, electronic transfer and storage options [4]. The wide exposure latitude of CR means that a greater range of radiation exposures (mainly currenttime product (mAs)) can be used to produce a diagnostic image [4]. Underexposure and overexposure with filmscreen radiography (FSR) are associated with poor image quality and the film may need to be repeated; whereas underexposure in CR results in a grainy image (high noise) that could be tolerated with some post processing, and higher than appropriate exposures in CR results in an increased image quality due to an increase in signalnoise ratio [58]. Gross artefacts are not evident on CR until exposure levels are 100 times lower or 500 times greater than those used in FSR [8]. Thus, while there is a potential for reduced repeat examinations and hence dose reduction in CR, it has been reported that radiographers develop a tendency to use higher than necessary exposure factors to improve image quality and avoid repeat radiographs. This tendency to increase exposure factors is referred to as exposure factor creep [6].
While FSR allows the radiographer to visually assess whether exposures are optimum, this direct visual assessment is lacking with CR systems. To allow for exposure parameter control and to safeguard from overexposure in CR, manufacturers have introduced numerical parameters that are a direct estimate of the incident exposure on the imaging plate [8]. Each manufacturer provides its own "exposure indicator", with an algorithm for its calculation; for example, this is called the exposure index (EI) by Kodak [9] and sensitivity (S) value by Fuji [10].
The EI used by Kodak represents the average raw data pixel value within a set anatomical region; a higher than average exposure is represented by an elevated EI with an increase in the EI by 300 representing a doubling of the plate exposure [8]. Kodak recommends that for all examinations, the EI should fall between 1700 and 1900 [11]. Unlike the Kodak convention, the Fuji S value decreases if the exposure increases [11], therefore the radiographer needs to have direct knowledge and training of the CR system they are using.
Even though these indices are provided by manufacturers to assist in providing a suitable image, it has been proposed that these can also be used as an indicator of patient dose levels [7, 8]. If the amount of radiation reaching the imaging plate is higher than recommended by the manufacturer, this could imply that the patient is receiving too much radiation or that the index recommended is too low (too high depending on system in use). The literature suggests that the recommended indices are not optimal and set too high, leading to increased patient dose [8].
The current research aims to investigate (i) whether radiographers are producing images with exposure indices within the manufacturers recommended range (MRR), (ii) the phenomenon of exposure creep, and (iii) the relationship between EIs and radiation dose.
| Equipment and methods |
|---|
|
|
|---|
Retrospective study
The Kodak Computed Radiography Systems 800 and 850 were used throughout this study. The difference between the systems is that the 800 reads fewer image plates per hour than the 850 (11). Two hospitals were selected for this study; Centre A installed CR in late 2000 and is a large teaching hospital, with a high patient throughput and a large number of radiographers. Centre B installed CR in mid 2004 and is smaller both in terms of patient throughput and radiographer numbers.
The PA chest and lateral (LAT) lumbar spine examinations were chosen as the PA chest is the most frequent examination performed (approximately 30%) in Australian radiology departments [12] and the LAT lumbar spine examination represents a high entrance surface dose examination [13]. These projections also examine different regions of the body and require different positioning and exposure factors.
Data were collected from January 2004 to June 2005 at centre A and between July 2004 and June 2005 for centre B. For the PA chest examination, 100 patients (first 10 patients (sorted alphabetically) from each day for the first 10 days) each month was sampled, while for the LAT lumbar spine (lower frequency examination) all available data was collected over the time period. Data collected included information on the examination performed, exposure index, and date and time of examination (allowing for "in or out" of hours analysis). Only data from patients over the age of 16 years were analysed.
Phantom study
The phantom study was conducted in hospital B. The phantom used was a Lung/Chest phantom (model CNR/R5330; Oxford Scientific, Silverwater, NSW 2128, Australia). The phantom was positioned as for a PA Chest examination and a series of exposures were taken at 125 kVp, 180 cm focusfilm distance (FFD) under automatic exposure control, as is standard in this department. The imaging plate was read using standard chest processing algorithm and the average of the 3 EIs (1550) was used to control all other exposure factors.
The phantom was then exposed to a range of tube potentials (70125 kVp) ensuring that the EI on the imaging plate remained fairly constant at 1550 (Table 1
). Three high sensitivity LiF:Mg,Cu,P thermoluminescent dosemeters (TLDs) (GR-200A; Solid Dosimetric Detector and Method Laboratory, Beijing, China) were positioned on the entrance surface of the phantom for each exposure, with the mean being calculated and used in the analysis.
|
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
As mentioned previously, there is a temptation for radiographers to set higher exposures than necessary without the risk of overexposure [8]. Experience has shown that when many radiology departments changed from FSR to digital, patient doses noticeably increased. This is due to many reasons; however, most commonly it is due to radiographer knowledge that an overexposed image can be resolved with post-processing techniques [4].
Regardless of the exposure indicators set by manufacturers, the values are sensitive to segmentation algorithms, tube potential, tube filtration, delay between exposure and readout, patient positioning and FFD [7]. Thus there are expected inherent differences between each radiology department, depending on the chosen tube potential, FFD and tube filtration in use. The use of a varying tube potential affects the amount of radiation absorbed by the patient and that reaching the detector. Therefore, it cannot be assumed that there is a good correlation between exposure index and patient doses.
Hospital A has a larger range of EIs for both examinations; reasons for this trend could be due to hospital A having a larger number of patients and greater number of staff, thus allowing a greater chance for variation in radiographic technique, directly impacting on EIs. Other explanations include radiographer complacency with regard to hospital A having had CR implemented for a longer period of time, or perhaps radiographers becoming aware of exposure factors that can be used to guarantee good image quality which may be outside the MRR.
The above reasons for EI variation are based on a hospital level; however, there are a number of factors which can affect EI variation on a patient level. These include: whether or not automatic exposure control (AEC) is used and, if so, was the patient centred correctly over it; differences in patient body habitus, which would affect the exposure factors required; and for the chest exam, whether or not filters are used to reduce breast shadowing.
For the LAT lumbar spine, 30% of examinations conducted by hospital B and 38% of PA chest examinations by hospital A were produced with EIs lower than MRR. These were all judged to be clinically acceptable by the reporting radiologists. This implies that the MRR may be set too high and can be reduced to values less than 1700. In the phantom study, the EIs were assessed to be 1550; again, all images were acceptable. Since a reduction in EI by 300 represents a halving of dose, the overall reduction from the median values (1770 to 1930) to 1550 would result in significant dose savings.
The first part of the phantom study shows the relationship between EI and ESD for a fixed tube potential. For a single X-ray unit, if the same tube potential is used and no other factors are changed, EI could be used as an indicator of ESD, as suggested by Peters and Brennan [8], that "it is safe to assume that a strong correlation exists between higher exposure indices and higher patient dose", and the opposite exists for low exposure indices.
The second part of the phantom study shows the relationship between a varied tube potential, EI and dose. As tube potential increased, a decreasing mAs was used to produce the same EI (1550) for each examination. The ESD as expected also decreased for each exposure, even though the EI remained constant. Other studies [8, 15] have suggested that if EIs are above the MRR, an unnecessary level of radiation reaches the imaging plate, resulting in an increase in patient dose. The results from this current study strongly suggest that EI should not be used as an indication of ESD when comparing examinations conducted between different radiographers and institutions using different tube potentials.
In the specific case of chest radiography, there are factors which may cause an increase in patient dose with changes to CR. The use of high tube potentials has been one of the main ways that patient dose has been reduced in the past. However, the nature of the phosphors used in CR is such that they are not suited to high tube potential techniques [16, 17]. CR systems absorb low energy photons more readily than those of higher energy levels, leading to excessively noisy images [18, 19]. A change to lower tube potential technique with corresponding increase in mAs and patient absorbed dose has been reported to be necessary [20, 21] in order to maintain image quality. However, Chotas et al found that, keeping the effective dose to the patient constant, the signal to noise ratio on CR chest images was almost constant between 60 kVp and 120 kVp [22]. In this present study both departments routinely use high applied potentials (above 100 kVp), demonstrating that diagnostic images are able to be produced at higher applied potentials in CR.
As stated in a study by Willis, exposure creep is a recognized phenomenon in CR [6]. A reason suggested for this is that there is a preference by radiographers and radiologists for overexposed images, rather than the grainy, noisy appearance of underexposed images. The most pronounced change in this study was the gradual increase (7.1%) in EI (Figure 3
) over an 18 month timeframe. As well as supporting reasons stated by Willis [6], another explanation for this exposure factor creep may be due to radiographers using higher than needed exposures so a repeat radiograph is not needed.
Comparing EIs during working hours and out of hours showed that there was little difference (for example in hospital A, 0.71% and 0.27% difference for the PA chest and LAT lumbar spine, respectively) between median exposure indices for both examinations and departments. These results do not support those reported by Peters and Brennan, who showed that there were "significantly higher exposure indices for images produced out of hours than in the normal working day" [8]. Mobile chest examinations (which were the focus of the Peters and Brennan study) have inherent exposure inconsistencies; for example, the absence of automatic exposure chambers, which would be available if performed in the radiographic department. The mobile chest examination away from the radiographic department could also be expected to place more pressure on radiographers to produce a good image, reducing the risk of repeat examinations and a return to the ward, than departmental examinations.
As radiation dose optimization is a significant concern in medical imaging, this research is aimed to increase radiographer awareness of exposure indices set by manufacturers and exposures used in the clinical setting. It is important for radiographers to understand CR concepts and not become complacent, in order to assure that dose is minimized with this technology. If radiation protection issues in digital radiography are not given careful attention, radiation exposure to patients will increase without concurrent benefits [23].
Due to the significant findings of this study and importance of reducing patient dose, it is recommended that further radiographer education and training should be implemented. Initial training should include general education about CR, explanation of EIs and how they are related to patient dose. In addition, regular quality assurance programs should be set-up to record and monitor changes in EIs over time in order to control any exposure creep. The results of this program should be fed-back to the radiographers so that they can become aware of their values and how they compare to others, allowing them to take appropriate action if needed. Manufacturers should also review the acceptable range of EIs recommended as this study has shown that phantom images produced with EIs as low as 1550 and for a range of tube potentials (70125 kVp) (phantom study) are clinically acceptable.
| Conclusion |
|---|
|
|
|---|
The phenomenon of exposure creep was also recognized. Over an 18 month period, a 7.1% increase in EIs was noted for the LAT lumbar spine at hospital A, highlighting the importance of continual radiographer training and quality assurance programs in order to keep patient doses as low as reasonably achievable.
The phantom study demonstrated that a small increase in EI produces a large increase in entrancesurface dose. This study also demonstrated that EI cannot always be used as an indicator of changes to patient dose, but that a number of other technical factors (predominately tube potential) need to be considered. The results showed that an EI of 2000 produced at 125 kVp can deliver the same patient dose as an EI of 1700 produced at 70 kVp, where the EI difference of 300 represents a doubling of dose to the detector.
The results of the retrospective study and phantom study together recognize the possibility that the Kodak MRR of 17001900 can be reduced for examinations of the PA chest and LAT lumbar spine, and the authors suggest that manufactures consider reassessing their current recommended levels. A reduction of as little as 10% of the MRR (15301700) can result in a dose saving of nearly 50% and this study has shown that EIs in this range produce diagnostically acceptable images.
Received for publication September 21, 2005. Revision received April 17, 2006. Accepted for publication May 30, 2006.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
N. KOWALCZYK and E. COMER Exposure Indicator Degradation From CR Plate Processing Delays Radiol. Technol., May 1, 2009; 80(5): 401 - 409. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. L FAUBER Exposure Variability and Image Quality in Computed Radiography Radiol. Technol., January 1, 2009; 80(3): 209 - 215. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. A. Tsalafoutas, G. A. Blastaris, A. S. Moutsatsos, P. S. Chios, and E. P. Efstathopoulos Correlation of image quality with exposure index and processing protocol in a computed radiography system Radiat Prot Dosimetry, June 1, 2008; 130(2): 162 - 171. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |