Figure 3


Figure 3. Plot of dose width product(DWP) measurements using an Unfors detector (techniques A1 – with full width half maximum (FWHM) derived from thermoluminescent dosemeter (TLD) profile and A2 – FWHM measured with a ruler and film) and a pencil ionization chamber (technique B), against the DWP derived from summation of doses across the beam from TLDs (technique C1). The line of identity is a 45° trendline.