British Journal of Radiology (2006) 79, 123-129
© 2006 British Institute of Radiology
doi: 10.1259/bjr/84646476
Image quality and breast dose of 24 screenfilm combinations for mammography
A D Dimakopoulou, MSc
1
I A Tsalafoutas, PhD
2
E K Georgiou, MD, PhD
1 and
E N Yakoumakis, PhD
1
1 Medical Physics Department, Medical School, University of Athens, 75 Mikras Asias, 115 27, Athens , 2 Medical Physics Unit, Konstantopoulio-Agia Olga Hospital, 3-5 Agias Olgas, Nea Ionia, 142 33, Athens, Greece
 |
Abstract
|
|---|
In this study the effect of different mammographic screenfilm combinations on image quality and breast dose, and the correlation between the various image quality parameters, breast dose and the sensitometric parameters of a film were investigated. Three Agfa (MR5-II, HDR, HT), two Kodak (Min-R M, Min-R 2000), one Fuji (AD-M), one Konica (CM-H) and one Ferrania (HM plus) single emulsion mammographic films were combined with three intensifying screens (Agfa HDS, Kodak Min-R 2190 and Fuji AD-MA). The film characteristics were determined by sensitometry, while the image quality and the dose to the breast of the resulting 24 screenfilm combinations were assessed using a mammography quality control phantom. For each combination, three images of the phantom were acquired with optical density within three different ranges. Two observers assessed the quality of the 72 phantom images obtained, while the breast dose was calculated from the exposure data required for each image. Large differences among screenfilm combinations in terms of image quality and breast dose were identified however, that, could not be correlated with the film's sensitometric characteristics. All films presented the best resolution when combined with the HDS screen at the expense of speed, and the largest speed when combined with the AD-MA screen, without degradation of the overall image quality. However, an ideal screenfilm combination presenting the best image quality with the least dose was not identified. It is also worth mentioning that the best performance for a film was not necessarily obtained when this was combined with the screen provided by the same manufacturer. The results of this study clearly demonstrate that comparison of films based on their sensitometric characteristics are of limited value for clinical practice, as their performance is strongly affected by the screens with which they are combined.
 |
Introduction
|
|---|
The main concern in mammography screening is the detection of features characteristic of breast disease. These features often have sizes of the order of 1 mm and differ from the normal tissue only slightly in composition, thus setting high requirements for the resolution and contrast that an imaging system must offer in order to be appropriate for mammography [1]. On the other hand, given the high radiosensitivity of the breast and the large number of women examined many times during their life, it is evident that the doses during mammography should be kept as low as possible.
While digital mammography may look promising, the vast majority of mammography examinations are still carried out with screenfilm systems. In recent years, most film manufacturers have presented new films and intensifying screens for mammography that reduce the dose to the breast and produce the image quality required to maintain the diagnostic sensitivity and specificity of mammography at high levels. However, while the design is the major factor in determining the performance of a film, this may be affected by the processing conditions, such as the chemicals used, their temperature and the processing time [2]. Inappropriate chemicals or a developing temperature lower than recommended may result in unacceptable mammograms and this is why some films have been modified to be less dependent on processing conditions [3, 4].
Film characteristics can be determined and monitored for changes due to processing by sensitometry. However, film performance will be dependent on the screen with which it is combined and thus for clinical practice the characteristics of the screenfilm combination rather than those of film or screen separately are of interest [5]. The screenfilm characteristics can be determined and monitored using an appropriate quality control (QC) phantom, with which changes in image quality due to processing or other reasons can be identified.
One of the parameters routinely monitored with the QC phantom is the background or reference optical density (OD) of the mammographic images. Apart from the personal preferences of radiologists, it has been shown that for a given screenfilm combination, subtle details and small contrast differences are best accentuated when the film OD is within a certain range [68]. For this reason it has been recommended that each institution should determine the optimum OD for the screenfilm used and the processing conditions specific to it [7].
In this study, eight films were combined with three intensifying screens and the resulting 24 screenfilm combinations were compared in terms of image quality and breast dose. Film characteristics were determined by sensitometry, whereas the image quality and speed of the screenfilm combinations were assessed using a QC phantom to obtain images within three different OD ranges. Our main objective was to investigate the effect of different screens on a certain film and search for any correlation between the image quality, breast dose and the sensitometric parameters of a film.
 |
Materials and methods
|
|---|
The eight single emulsion mammographic films tested in this study were the MR5-II, HDR, HT (Agfa-Gevaert N.V., Mortsel, Belgium), Min-R M, Min-R 2000 (Eastman Kodak Company, New York, NY), AD-M (Fuji Photo Film Co. Ltd, Tokyo, Japan), CM-H (Konica Corporation, Tokyo, Japan) and HM plus (Ferrania Sp A, Ferrania (SV), Italy). One box from each film type was used to avoid little differences that may exist among different film batches or films that may have been stored for different times and under different storage conditions [9]. The three intensifying screens used in this study were the HDS (Agfa), the Min-R 2190 (Kodak) and the AD-MA (Fuji).
For each film a 21-step sensitometric strip was produced, using an X-Rite 334 sensitometer (X-Rite, Grandville, MI) operated in the green spectrum. All films were processed in a daylight processor (Curix Capacity, Agfa) with nominal processing time 90 s (22 s developing time) and with the developer temperature set to 36°C. The developer type was the Eos Dev (Agfa) and the fixer type was the G334i (Agfa). All films were processed sequentially, immediately after exposure and on the same day, to avoid day-to-day variations in processing conditions caused by the ageing of the chemicals that may have variable effects on the characteristics of each film [9, 10].
The OD of the 21 steps of the sensitometric strips was measured using a calibrated optical densitometer (RMI 331, X-Rite). For each film the Hurter-Driffield (H&D) curve was plotted and the following sensitometric parameters were derived: OD of base plus fog (ODb+f), maximum OD (ODmax), average gradient (AG), film gamma (
) and film speed. The AG and
are the slopes of the H&D curve for ODs from 0.25+ODb+f to 2.0+ODb+f and from 1.0+ODb+f to 2.0+ODb+f, respectively. AG and
are both used as indices of film contrast, however, only
can be used to reproduce the linear part of the H&D curve. The film speed was defined as the reciprocal of the relative light exposure required to obtain an OD of 1+ODb+f. Using this definition, the higher the film speed the less exposure is needed for a given OD. In order to illustrate the expected increase in breast dose according to sensitometry when a film other than the fastest one is used, the sensitometric relative dose index (SRDI) was defined as the reciprocal of the relative speed value. The SDRIs were expressed as percentages of the smallest SRDI value (highest speed) that was taken as 100%. It must be noted that from preliminary sensitometric tests it has been confirmed that for different sheets of the same film type processed within the same day, variations of less than ±0.01 in ODb+f, ±10% in speed and ±0.1 in ODmax, AG and
should be expected.
To evaluate the characteristics of screenfilm combinations, a mammography QC phantom was employed (breast phantom, Model 18-222; Nuclear Associates, Division of Victoreen Inc., NY). This phantom is realistically shaped and equivalent to an average firm breast of 4.5 cm compressed thickness, consisting of 50% adipose and 50% glandular tissue. It includes 12 groups of calcium carbonate specks (simulating microcalcifications), 7 hemispheric masses composed of 75% glandular and 25% adipose equivalent tissue (simulating tumours) and a wax insert with 5 embedded nylon fibres (simulating glandular tissue fibrils). The phantom also contains a five-step stepwedge, simulating breast areas with compositions 100% adipose, 70% adipose30% glandular, 50% adipose50% glandular, 30% adipose70% glandular and 100% glandular tissue. Finally, two line-pair test targets (520 lp mm1 each), one parallel and one perpendicular to the anodecathode axis and a central area where the background OD is measured, are included. A similar phantom (without the nylon fibres and with only one line-pair test target) was used by Nassivera and Nardin [11].
The phantom was exposed using a Senographe 500T mammography unit (CGR, Buc, France). All exposures were made with the Mo/Mo target filter combination, constant tube potential (28 kVp), large focal spot (0.3 mm nominal size) and without the breast compression paddle. Using manual mAs selection technique, images of the phantom were acquired until for each screenfilm combination three films with OD as close as possible to the central OD of three different optical density ranges (0.701.10, 1.111.50, 1.512.00) were produced to account for the wide range of ODs that can be encountered within actual mammographs. Phantom images were processed in the same processor and on the same day as the sensitometric strips, so both film characteristics and screenfilm performance were determined under the same processing conditions. It must be noted that as the Kodak and Fuji cassettes were not compatible with the Agfa daylight processor, the films exposed with these cassettes had to be transferred manually to the Agfa cassette in order to be fed into the processor.
For each one of the resulting 72 phantom images, the OD of the central area was measured with the densitometer, as well as the OD of the areas simulating 100% adipose and 100% glandular tissue. The OD difference of these two areas can be used as an index of screenfilm contrast (CI) and, according to the phantom manufacturer, it should be
0.28.
All phantom images were examined using a viewing box especially designed for mammography, featuring adjustable brightness, masking shutters and a magnifying glass. The shutters were closed down to the phantom image size and the brightness was adjusted as necessary to obtain the best possible conditions for viewing each type of simulated lesion, while for speck groups the magnifying glass was also used. A magnifying glass supplied with the QC phantom was used to inspect the line pair object. The above details are mentioned, as viewing conditions are very important for interpreting mammograms or scoring phantom images [12].
Two observers scored the images independently and any disagreements were resolved by consensus. Five scores were recorded for each film: one for the speck-groups, one for masses, one for fibres and two for the two line-pair test targets. For ambiguous decisions concerning not clearly visualized structures, a 0.5 mark was assigned. In order to have a single index characterizing the screenfilm performance, a total score (TS) was calculated using the following weighting coefficients: 0.4 for specks, 0.35 for masses and 0.25 for fibres. These coefficients were selected after consulting with five radiologists about the clinical importance of each simulated structure for diagnosis. Since the two scores for the line-pair test targets have no straightforward clinical relevance, their mean value was calculated for reference only (resolution score).
For screenfilm combination comparisons in terms of breast dose, the entrance surface air-kerma (ESAK) at the phantom surface was calculated from the mAs selected for each exposure. For the range of mAs selections utilized in this study, the output at 28 kVp defined at the phantom entrance surface was 98±2 µGy mAs1. Furthermore, the ESAK required to achieve a net OD of 1 was calculated by interpolation from the ODs and the ESAKs of the three films acquired for each combination. The resulting ESAKs for a net OD = 1 were used to derive the relative dose index (RDI), expressed as a percentage of the smallest observed value, that was considered as 100%. Using this definition, the larger the RDI, the larger the dose to the breast and the smaller the speed of a given screenfilm combination.
To investigate the correlation between the various image quality parameters, sensitometric parameters, OD and dose indices linear regression analysis was used. A correlation coefficient (r) larger than 0.7 was taken as an indication of good correlation. Specifically, the correlations of all the image quality scores (TS, specks, masses, fibres and resolution) with AG,
, SRDI, CI, OD and ESAK were investigated. Furthermore, the correlation of TS with resolution, the correlations of CI with AG,
, SDRI, OD and ESAK and the correlations of SDRI with RDI and ESAK were also investigated.
 |
Results
|
|---|
The H&D curves for all the films studied are plotted in Figure 1
, while their sensitometric parameters are given in Table 1
. In Figure 1
, the large differences among the H&D curve shapes, the high speed of Min-R 2000, the low speed of AD-M and the non-typical but similar H&D curve shapes of HDR and HM plus should be noted. From Table 1
, it can be seen that the Min-R 2000 presents the highest speed and
, while the CM-H has the highest AG.

View larger version (27K):
[in this window]
[in a new window]
|
Figure 1. The Hurter-Driffield(H&D) curves of the eight films included in this study are given. The optical densities (ODs) of the 21 steps of the sensitometric strips correspond to log relative exposure values (LogE) that range from 0 to 3, in steps of 0.15 each. In these figures, only the ODs for LogE ge;0.9 (steps 7 to 21) are presented in order to enhance the visibility of the differences in the linear part and the shoulder of the H&D curves.
|
|
View this table:
[in this window]
[in a new window]
|
Table 1. The results of the sensitometric evaluation of the eight mammographic films included in this study. The lowest speed was arbitrarily defined as 100%. The sensitometric relative dose index (SRDI) was defined as the reciprocal of the relative speed, considering the lowest value (highest speed) as 100%
|
|
Concerning the screenfilm comparisons, for the 72 phantom images evaluated in this study, speck scores ranged from 6 to 11, mass scores from 2.5 to 6, fibre scores from 2 to 4, total scores from 4.3 to 7.3, resolution scores from 10 to 14.5 lp mm1 and CI from 0.26 to 0.61. For each of the three OD ranges the respective ESAK ranges were: 2.04.9 mGy, 2.46.2 mGy and 3.18.8 mGy. The results for the 24 combinations studied in terms of the ESAK at the phantom surface, the background OD, the TS, the CI and the resolution score are presented in Tables 2a, 2b and 2c

for the three different OD ranges, respectively. In the two last columns of each table, the screenfilm RDI and the mean value of TS in the three OD ranges (TSm) are also given. From these tables it can be seen that the screen affected the image quality of a given film as well as the dose required for obtaining a phantom image of a certain OD. Furthermore, it can be seen that for a certain screenfilm combination the TS, CI and resolution were dependent on the background OD. The screenfilm combinations that exhibited the largest dependency with OD were the AD-MA/HM plus for the TS, the Min-R 2190/Min-R 2000 for the CI and the HDS/MR5-II for the resolution score.
View this table:
[in this window]
[in a new window]
|
Table 2a. The results of the evaluation of the 24 filmscreen combinations, from the quality control (QC) phantom images with optical density (OD) in the range 0.71.1. The screenfilm combination relative dose index (RDI) has been expressed in terms of the entrance surface air-kerma (ESAK) required to produce a net OD of 1, considering the lowest value observed (2.19 mGy) as 100%
|
|
View this table:
[in this window]
[in a new window]
|
Table 2b. The results of the evaluation of the 24 filmscreen combinations, from the quality control (QC) phantom images with optical density (OD) in the range 1.111.5. The screenfilm combination relative dose index (RDI) has been expressed in terms of the entrance surface air-kerma (ESAK) required to produce a net OD of 1, considering the lowest value observed (2.19 mGy) as 100%
|
|
View this table:
[in this window]
[in a new window]
|
Table 2c. The results of the evaluation of the 24 filmscreen combinations, from the quality control (QC) phantom images with optical density (OD) in the range 1.512.0. The screenfilm combination relative dose index (RDI) has been expressed in terms of the entrance surface air-kerma (ESAK) required to produce a net OD of 1, considering the lowest value observed (2.19 mGy) as 100%
|
|
In order to highlight some of the major results of this study, the best screen for a given film and the best film for a given screen in terms of TS, CI, resolution and RDI are given in Tables 3 and 4
, respectively, where the OD range for which the best score is obtained is also noted. It is evident that, while for TS and CI there was variety in the screen with which a film was best combined, for resolution and RDI the best screen was common for all films. Indeed, all films presented the best resolution when combined with the HDS screen and the smallest RDI (largest speed) when combined with the AD-MA screen. The largest resolution offered by the HDS screen was at the expense of speed, while the largest speed offered by the AD-MA screen was at the expense of resolution but not at the expense of TS.
In summary, Tables 3 and 4
clearly demonstrate two points that deserve special attention. First, the highest TS, CI, resolution and the lowest RDI were observed for different combinations and thus an "ideal" combination was not identified. Second, the best performance of a given film or screen was not always obtained when combined with the screen or film, respectively, from the same manufacturer.
Finally, concerning the correlation between image quality parameters, sensitometric parameters and dose, no correlation coefficient larger than 0.7 was calculated in any of the correlations tested. The largest correlation coefficient calculated was that between CI and OD, which was 0.66 but increased to 0.87 when the MR5-II combinations were excluded, demonstrating that for modern high contrast films the screenfilm contrast increases with OD. It must be clarified, however, that this correlation has been assessed for ODs up to 2.0 and it is expected that for higher ODs the CI will start to decrease again as the films become saturated.
The poor correlations of image quality parameters (TS, specks, masses, fibres, resolution and CI) with film contrast (
, AG) and the poor correlation of RDI with SRDI, confirmed that film performance is strongly affected by the screen. Concerning the lack of correlation of TS with CI, resolution and dose, the following remarks should be made. While combinations with high TS generally had high CI, there were many cases with high CI and low TS. There were also many combinations with high TS that, however, exhibited low resolution score and vice versa. Finally, slow combinations did not always give high TS, as expected according to the general principle that the higher the dose the lower the quantum mottle.
 |
Discussion
|
|---|
The major conclusion of this study was that film characteristics are modified by intensifying screens in such a significant and variable way, that comparisons among films based on the manufacturer's specifications or sensitometry are of limited value. Indeed, a film with given technical specifications or sensitometric characteristics, when combined with different screens may exhibit improved or degraded performance.
Instructive of the variable effect that a screen may have on a film, it can be seen that while Min-R 2000 was the fastest film and remained the fastest when combined with the Fuji screen, the RDI ratio of CM-H and Min-R 2000 combinations with Fuji screen was 1.04 while for film only the respective ratio of SDRI was 1.32. That means that the Fuji screen spectral emission better matched the spectral sensitivity of CM-H compared with Min-R 2000. The slowest film, according to sensitometry, was the AD-M, which remained the slowest when combined with all screens. The SDRI for AD-M was 167 while its smallest RDI was 191 and it was observed when combined with the Fuji screen.
Examples of the largest variations observed in score and dose when a film was combined with different screens are: the HM plus where the TS was 36% larger with the Fuji than with the Agfa screen (Table 2c
), the Min-R 2000 where the CI was 46% and 22% larger with the Agfa than with the Kodak screen (Tables 2a and 2c
, respectively), the Min-R 2000 where the resolution was 26% larger with the Agfa than with the Fuji screen (Table 2b
) and the HT (Agfa) where 63% more dose is required with the Agfa screen than with the Fuji screen.
As previously mentioned, from Tables 24

some conclusions may be drawn concerning the superiority of certain combinations over others in terms of image quality or speed. However, the absolute values of scores and other screenfilm characteristics may be quite different on other mammographic facilities, given the strong dependence of film characteristics on processing conditions [13, 14]. This must be emphasised, as the objective of this study was not to recommend or condemn certain films or screens but to investigate the effect of screens on the performance of films. Although most screenfilm comparisons in the literature have been carried out using the same processing conditions for all films, it must be noted that the general notion is that a film would perform optimally when it is processed according to the recommendations of the manufacturer. Even so, this does not annul the fact that the breast dose and image quality for a film optimally processed will again vary, depending on the screen with which it is combined, and that some films will be affected by the screen more than others.
Even if it were assumed that the processing conditions were optimal for all films, it would again be difficult to select the best screenfilm combination from those studied, as there are no established criteria about what increase in breast dose is justified by a superior image quality. For example the AD-MA/AD-M presented a TS of 7.3 (11 specks, 6 masses, 3 fibres) and an ESAK of 6.2 mGy while the AD-MA/Min-R 2000 a TS of 6.4 (9 specks, 5 masses, 4 fibres) with an ESAK of 2.4 mGy. To conclude which is the best combination, one has to decide if the 14% increase in TS could justify the 158% increase in breast dose. The same question still holds when considering that certain combinations (as the AD-MA/AD-M) exhibited slightly larger TS for larger ODs but with disproportional increase in breast dose.
An important remark should also be made concerning the OD of the films studied. It is obvious that the films included in Table 2a
are of too low OD and few of the films included in Table 2c
are of too high OD, compared with the target OD range of 1.3 to 1.8 proposed for mammography [1]. Nonetheless, certain combinations exhibited better scores in Table 2a
than in Tables 2b
and 2c
, while most of the films of Table 2c
with ODs larger than 1.8 exhibited scores similar to those of Table 2b
. In clinical practice, however, given that the wide OD variations within a mammogram are not uncommon, some areas may present similar ODs with those of Table 2a
or larger than 1.8 and therefore the performance of a screenfilm combination within all OD ranges is of interest. In this context, comparisons based on the TSm may be considered more relevant to the clinical situation than comparisons based on the TS within only one OD range. The variability of TS with OD should always be considered when selecting the central OD setting of the automatic exposure control (AEC) system based on the results of phantom scores.
Some final comments should be made concerning the method used to assess the image quality of screenfilm combinations. Phantom scoring does not always represent clinical practice, as in actual mammograms the performance of a given combination will be also dependent on the breast type [15]. Furthermore, phantom scoring may be somewhat biased, as it relies on the detection of structures known to be present at specific positions [16]. Nevertheless, phantoms are considered as the best way for the objective evaluation of image quality and various models with fixed or randomly positioned details are extensively used. Caldwell et al [17] agreed on the usefulness of such phantoms for the objective evaluation of image quality and also reported that a subjective assessment of image quality is better accomplished with an anthropomorphic breast phantom than with actual mammograms, where the variability among radiologists was higher. However, they noted that no significant correlation was found between the various methods used to evaluate image quality and concluded that more work is required to obtain an index of true image quality correlated with the probability of correct diagnosis.
In conclusion, image quality and dose in mammography are more strongly dependent on screenfilm combination than on film or screen separately. While sensitometry remains an important tool for determining and monitoring the film characteristics [18], it is of little value when the image quality and breast dose in clinical mammograms are of concern. Therefore, any change of film or screen type in a mammographic facility should be carefully investigated with a phantom, for determining the performance of the selected screenfilm combination and for adjusting the AEC system to the optimum OD range for this combination.
Received for publication February 26, 2005.
Revision received May 23, 2005.
Accepted for publication June 7, 2005.
 |
References
|
|---|
- van Woudenberg S, Thijssen M, Young K. European protocol for the quality control of the physical and technical aspects of mammography screening. In: Perry N, Broeders M, de Wolf C, Kirkpatrick A, Tornberg S, editors. European guidelines for quality assurance in mammography screening (3rd edn). Luxembourg: Office for Official Publications of the European Communities, 2001
- Brink C, De Villiers JFK, Lötter MG, Van Zyl M. The influence of film processing temperature and time on mammography image quality. Br J Radiol 1993;66:68590.[Abstract/Free Full Text]
- Tabar L and Haus AG. Processing of mammographic films: technical and clinical considerations. Radiology 1989;173:659.[Abstract/Free Full Text]
- Kimme-Smith C, Bassett LW, Gold RH, Zheutlin J, Gornbein JA. New mammography screen/film combinations: imaging characteristics and radiation dose. AJR Am J Roentgenol 1990;154:7139.[Abstract/Free Full Text]
- Kirkpatrick AE, Law J. A comparative study of films and screens for mammography. Br J Radiol 1987;60:738.[Abstract/Free Full Text]
- Robson KJ, Kotre CJ, Faulkner K. The use of a contrast-detail test object in the optimization of optical density in mammography. Br J Radiol 1995;68:27782.[Abstract/Free Full Text]
- McParland BG, Boyd MM, Yousef KAL. Optimizing optical density of a Kodak mammography film-screen combination with standard-cycle processing. Br J Radiol 1998;71:9503.[Abstract]
- McParland BJ. A comparison of two mammography film-screen combinations designed for standard-cycle processing. Br J Radiol 1999;72:735.[Abstract]
- Kimme-Smith C, Bassett LW, Gold RH, Chow S. Increased radiation dose at mammography due to prolonged exposure, delayed processing and increased film darkening. Radiology 1991;178:38791.[Abstract/Free Full Text]
- Fernandez JM, Guibelalde E. Technical note: Physical evaluation of recent Kodak films for mammography. Br J Radiol 1993;66:82832.[Abstract/Free Full Text]
- Nassivera E, Nardin L. Daily quality control programme in mammography. Br J Radiol 1996;69:14852.[Abstract/Free Full Text]
- Pisano ED, Britt GG, Lin Y, Schell MJ, Burns CB, Brown ME. Factors affecting phantom scores at annual mammography facility inspections by the U.S. Food and Drug Administration. Acad Radiology 2001;8:86470.
- Tsalafoutas IA, Dimakopoulou AD, Koulentianos ED, Serefoglou AN, Yakoumakis EN. The variation of the sensitometric characteristics of seven mammographic films with processing conditions. Br J Radiol 2004;77:66671.[Abstract/Free Full Text]
- Kimme-Smith C, Rotschild PA, Bassett LW, Gold RH, Moler C. Mammographic Film-Processor Temperature, Development Time and Chemistry. AJR Am J Roentgenol 1989;152:3540.[Abstract/Free Full Text]
- Meeson S, Young KC, Rust A, Wallis MG, Cooke J, Ramsdale ML. Implications of using high contrast mammography X-ray film-screen combinations. Br J Radiol 2001;74:82535.[Abstract/Free Full Text]
- Jackson VP, Harrill CD, White SJ, Gillespie KR, Mail JT, Katz BP. Evaluation of a dual-screen, dual-emulsion mammography system. AJR Am J Roentgenol 1989;152:4836.[Abstract/Free Full Text]
- Caldwell CB, Fishell EK, Jong RA, Weiser WJ, Yaffe MJ. Evaluation of mammographic image quality: pilot study comparing five methods. AJR Am J Roentgenol 1992;159:295301.[Abstract/Free Full Text]
- West MS and Spelic DC. Using light sensitometry to evaluate mammography film performance. Med Phys 2000;27:85460.[CrossRef][Medline]