| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Full Paper |
1 SCK-CEN, Belgian Nuclear Research Centre, Boeretang 200, 2400 Mol, 2 Department of Radiology, University Hospitals of the KU Leuven, Herestraat 49, 3000 Leuven and 3 Free University of Brussels, Pleinlaan 2, 1050 Brussels, Belgium
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The effective dose E, a measure for stochastic risk, cannot be measured directly. A practical approach starts from entrance surface dose (ESD) or dosearea product (DAP) measurements and uses dedicated conversion coefficients.
Different organizations determined conversion coefficients that link ESD or DAP measurements with E. Previously, DAP values for successive projections of angiographic procedures of the lower limbs were multiplied by published conversion coefficients for simple radiographs of corresponding anatomical regions [2, 3]. The following limitations were observed:
We concluded that the use of these conversion tables might not provide sufficient guidance for establishing optimization strategies for angiographic procedures. As the quantity effective dose (E) is likely to be a useful parameter for the optimization of a technique, more dedicated conversion coefficients seemed necessary.
This paper deals with the calculation of the conversion coefficients and the influence of different parameters on these coefficients.
| Material and methods |
|---|
|
|
|---|
In six different hospitals, the following parameters were registered for this type of examination: radiation quality, anatomical area irradiated, beam projection and the use of extra absorption filters.
Two hospitals (C and D) have an overcouch configuration, i.e. the X-ray tube is above the patient support table. The four other hospitals (A, B, E and F) have an undercouch configuration.
Questionnaires regarding working procedure were completed by all the radiologists in all hospitals. For every exposure, the projections of the beam and the exposed area of the body were clearly documented (Table 1
). In Figure 1
the field positions for exposure of the abdomen and pelvis are shown for hospitals AF, respectively. However, it is very probable that the exact field position will vary not only between the centres, but also between the patients from the same centre. In hospital C, the AP projection of the pelvis (Figure 1c
, bold line) does not adjoin the AP projection of the abdomen. It starts much lower, at the bladder height. The adjoining pelvis projections are the RPO and LPO projections. Therefore, in this hospital the AP projection only partly overlaps the LPO and RPO projections of the pelvis. Consequently, the exposure of the upper legs, adjoining the pelvis AP projection, also starts lower compared with the other centres. In hospital D, there is no exposure of the abdomen.
|
|
The radiation spectra were characterized by the total filtration and the tube potential value. With a PMX-3 solid state dosemeter (RTI Electronics, Sweden) the half value layer (HVL) was determined for every tube in the different hospitals, by measuring dose rate for different thicknesses of aluminium. From these measurements, the total filtration is determined using the conversion tables in the application note of RTI electronics [4]. These filtrations, in terms of equivalent aluminium, comprising X-ray tube and patient support table for the undercouch configurations, are shown in Table 2
, together with possible additional copper filtration and the average peak voltage used for every projection. With the "Catalogue of Diagnostic X-ray Spectra and other Data" of IPEM [5] the different spectra in the hospitals were determined for every projection separately, based on the total filtration of the beam and the tube potential.
|
The phantom's sex and age has to be chosen, as well as the organs to be included in the model. For both sexes, the adult phantom is chosen (21 years for male and 15 years for female model).
For one specific spectrum, projection and exposed body area, MCNP calculates doses in every organ of the phantom. With these organ doses and the tissue weighting factors specified by ICRP [2], effective doses were calculated. The theoretical DAP values associated with a given number of simulated X-ray photons were calculated from the average fluence through a surface at 20 cm from the source. Dividing the calculated effective dose by the theoretical DAP value gives the conversion coefficient for one specific exposure.
Calculation and comparative study of the conversion coefficients
Conversion coefficients were calculated for every specific projection in each hospital and for male and female patients separately. For every hospital the radiation fields are chosen as illustrated in Figure 1
and radiation quality chosen as in Table 2
. The difference between the coefficients will be studied in more detail. The mean difference between conversion coefficients for men and women is calculated for each anatomical region. The factorial experiment is used to explain the differences in conversion coefficients between the centres, caused by various parameters.
Factorial experiments
Several parameters, such as field size, field position, patient thickness and radiation spectrum (kVp and filtration) are known to affect the conversion coefficients for estimating the effective dose. Ideally, the influence of each parameter would be independent, so that an estimate of a conversion coefficient is the product of several parameters [9]:
|
|
is a representative value of the coefficient and
i,
j...
n are the factors which take into account the effect of the variations of the n parameters. It is important not only to determine if the parameters have an influence on the conversion coefficients, but also if there is a significant interaction between the parameters. This means that Equation (1)|
|
ij...
in...
jn... are correction factors which include the additional effect of interaction between the parameters. In general, higher-order interactions do not contribute significantly to the observed variations. This model converts into a linear model if natural logarithms are taken of Equation (2)In our experiment, the effect of the following parameters on the calculation of conversion coefficients for the effective dose were considered:
The low and high level values chosen for the parameters are determined from our detailed observations for several interventional procedures in different hospitals. We believe that these ranges are representative values for the parameters considered in interventional radiology for the exposure of abdomen and pelvis. For tube voltage (6585 kVp) and for filtration (4.1 mm Al6.5 mm Al+0.1 mm Cu). In Figure 2
, the low and high levels for the field position are illustrated.
|
Statistical analysis is performed for each of the projections on the logarithms of (E/DAP) conversion coefficients and the main factorial effects and all pair-wise interactions were investigated. These analyses were done using the generalized linear model of the computer package STATISTICA.
The results of the factorial experiment are applied to the differences between the conversion coefficients, calculated for angiography of the lower limbs for the six specific hospitals.
| Results |
|---|
|
|
|---|
|
Effective doses can be calculated for the full procedure by multiplying the conversion coefficients in Table 3
by the measured DAP values per projection in all of the six hospitals:
with i the different projections.
In Table 4
, the average DAP data from each hospital is given for each projection. This results in average effective doses of 8.0 mSv for hospital A (15 patients), 15.5 mSv for hospital B (16 patients), 8.0 mSv for hospital C (20 patients), 3.9 mSv for hospital D (13 patients), 16.8 mSv for hospital E (18 patients) and 11.3 mSv for hospital F (12 patients).
|
The results of the factorial analysis are given in Table 5
. The first value is a representative value for the conversion coefficient and corresponds to the geometrical mean of the eight possible combinations. The main and interaction effects are expressed as multipliers, illustrating how big the effects of the different parameters and their pair-wise interactions are on the mean value of the conversion coefficients. In Table 5
, the multipliers are given for the low level of every parameter and their interactions. According to the constraints in the statistical analysis [9], the multiplier for the high level is the reciprocal of the multiplier for the low level. For multipliers <1, the conversion coefficient increases and for multipliers >1, the coefficient decreases when the parameter changes from the low to the high level. By analogy with West et al [9], variations between the low and high value of less than 2% are described as "insignificant", of 25% as "small", of 515% as "moderate" and of over 15% as "large".
|
15%) for the abdomen AP and PA projection, but a very large effect (3151%) for the pelvis AP and PA projection. For every projection, the pair-wise interactions between the three parameters can be categorised as insignificant, except for the interaction between tube potential and filtration, which should be categorised as small. Moreover, for the pelvis AP projection variations of 7.5% are observed, where the interactions with the parameter "field position" is involved. This tells us that the effect on the conversion coefficients of changing the tube potential or the filtration will depend on the field position for the pelvis AP projection. The effect is moderate.
Conversion coefficients for every combination of the three parameters can be calculated per projection, following Equation (2)
.
For example, for an abdomen PA exposure at 65 kVp and 6.5 mm Al+0.1 mm Cu with a high field position, the conversion coefficient can be calculated from Table 5
as:
CF=0.170 x 0.842 x (1/0.834) x (1/0.914) x (1/0.986) x (1/0.995) x 0.998=0.191
Since for the abdomen PA projection most interactions are insignificant, they can be omitted. For other levels of a specific parameter, the multiplier can be determined from an interpolation between the low and the high multiplier of that parameter.
| Discussion |
|---|
|
|
|---|
Overcouch versus undercouch configurations
The position of the different radiation sensitive organs in the human body determines the difference in conversion coefficients between the anterior and posterior projections. For the anterior projection of the abdomen, radiation sensitive organs like the liver, stomach and lower large intestine (LLI) have larger doses compared with the posterior projection. For the latter projection, the doses to the mineral bone surface and the red bone marrow are larger compared with the anterior projection. However, they contribute less to the effective dose, because their doses are lower compared with those of the liver, stomach and LLI.
Abdomen projection
From the statistical analysis we can see that the maximum variation in conversion coefficients for the change in field position for the AP and PA exposure of the abdomen is around 15%. The difference between the conversion coefficients in Table 3
, for this projection, is therefore mainly due to differences in tube configuration and radiation spectrum. The variation for these projections due to tube potential and filtration is large. This explains why hospitals A and F have larger conversion coefficients for the abdomen exposure. From Table 2
, we observed the largest filtration for hospital A and the highest tube potential value for hospital F. The conversion coefficient in hospital C is larger due to the anterior projection, as explained in the previous paragraph.
Pelvis projection
The influence of the field position is much larger for the pelvis exposure: 31% for the AP-projection and 51% for the PA-projection. As mentioned before, this is caused by the difference in exposure of the gonads. This explains the high conversion coefficient for male patients in hospital C in Table 3
: the inclusion of the gonads in the radiation field and the direct exposure due to the anterior projection. For the female patients, the conversion coefficient for the pelvis exposure in hospital C is much smaller compared with the other hospitals. The ovaries are far away from the radiation field in this hospital (Figure 1
). Again for the pelvis exposure, we observe from the statistical analysis that tube potential and filtration have a large influence for the PA and AP projections, respectively. This explains the difference between the conversion coefficients in the hospitals where the difference in field position with respect to the gonads is small.
Upper leg projection
For the upper leg projection, the conversion coefficient for the male patients in hospital D is much larger than in other hospitals. Again, this can be explained by the inclusion of the gonads in the radiation field and by the anterior projection. In hospital C, the conversion coefficient for the upper leg for men is lower than in any other centre. Because of the lower position of the radiation field for this projection (Figure 1c
), the gonads are more distant from the field. We can also observe a difference in conversion coefficients for the hospitals with tube-under-table configuration: for hospital E (0.032 mSv Gy1cm2) where the gonads are not in the radiation field, for hospital B (0.046 mSv Gy1cm2) with gonads partly exposed and for hospital F (0.064 mSv Gy1cm2) with gonads completely exposed. The higher conversion coefficient in hospital A (0.095 mSv Gy1cm2) is mainly caused by the larger filtration.
Patient thickness
The calculation of conversion coefficients for the six hospitals is made for standard-sized patients. In practice, this will not be the case. For the same tube potential and filtration, thicker patients will have smaller conversion coefficients because the organ doses are smaller compared with the standard patient. In practice, however, thicker patients could have larger tube potential values, implying a higher conversion coefficient. However, this will not necessarily mean a larger effective dose to the patient. A higher tube potential might imply a lower mAs value, depending on the tube potential curve and the dose level set in the system. Consequently, the DAP value could decrease. The steeper the tube potential curve, the smaller will be the mAs and the DAP.
Optimization guidelines
When minimizing effective doses for patients, the doses to the gonads play an important role. For procedures such as angiography of the lower limbs, exposure of the gonads is inevitable. In all hospitals the gonads for both genders are often exposed completely or partly in the pelvis projection. However, due to the possible additional oblique views, the pelvis is often more exposed than the other body regions. Therefore, it would be preferable for male patients to expose the gonads with the upper legs projection. Moreover, the beam path through the upper legs is smaller than through the pelvis. This results in lower tube potential and/or mAs values and consequently in lower DAP values. The higher conversion coefficient for the upper legs projection (due to the presence of the gonads) will contribute less to the effective dose, because of the lower associated DAP value.
For the female patients, the ovaries could be exposed with the abdomen projection to avoid the extra exposure from the oblique projections of the pelvis. The ovaries, however, are situated at that point where the abdominal artery splits into the two arteries leading to the legs. As it is mainly for this position of the arterial system that the oblique projections are performed for angiography of the lower limbs, it is not possible to include the ovaries in the abdominal exposure from a medical point of view.
Kicken [11] reported (E/DAP)-conversion coefficients for an abdomen PA projection. They were 0.102 mSv Gy1cm2 for the male patients for 80 kVp and 0.171 mSv Gy1cm2 for the female patients for 75 kVp, both with a total filtration of 6.0 mm Aleq. They are comparable with the conversion coefficients in our study.
The conversion coefficients in NRPB-R262 [2, 12] are interpolated for 75 kVp and 5 mm Aleq as 0.240, 0.149, 0.268 and 0.112 for the abdomen and pelvis AP and PA projections, respectively. These values, averaged over male and female patients, are comparable with the average values given in Table 5
.
The conversion coefficients in this study are for the procedure of angiography of the lower limbs in several hospitals. However, they can also be used for similar projections in other procedures. For example, the conversion coefficients for the pelvis projections can be used in an iliac stenting procedure or the conversion coefficient for the abdomen irradiation can be used in a stent procedure of the kidneys or for a TIPS (transjugular portosystem shunt) procedure, provided that filtration and tube potential are known and that DAP values for each of these projections in the specific procedure are available.
One of the limitations of the study is the small number of participating hospitals. Also, an accurate calculation of effective dose requires a separate DAP value for every projection. This is not routinely available yet.
| Conclusions |
|---|
|
|
|---|
30%). It is recommended to establish a set of conversion coefficients for the entire range of tube potential and filtration in angiography and interventional radiology; In order of priority, we propose to treat separately undercouch and overcouch set ups, to account then for male versus female patients, to adjust conversion coefficients in terms of radiation quality and finally, if necessary, to fine tune for differences in radiation field. If the position of the radiation field is close to or involves the gonads, greater attention should be given to the calculation.
Received for publication December 22, 2003. Revision received September 2, 2004. Accepted for publication September 29, 2004.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H GEIJER, E NORRMAN, and J PERSLIDEN Optimizing the tube potential for lumbar spine radiography with a flat-panel digital detector Br. J. Radiol., January 1, 2009; 82(973): 62 - 68. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Smans, L. Struelens, M.-T. Hoornaert, F. Bleeser, N. Buls, D. Berus, P. Clerinx, F. Malchair, F. Vanhavere, and H. Bosmans A study of the correlation between dose area product and effective dose in vascular radiology Radiat Prot Dosimetry, July 1, 2008; 130(3): 300 - 308. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. Martin RADIATION DOSIMETRY FOR DIAGNOSTIC MEDICAL EXPOSURES Radiat Prot Dosimetry, March 28, 2008; (2008) ncm495v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. T. Bahreyni Toossi, H. Zare, S. Bayani, and S. Esmaili Organ and effective doses of patients arising from coronary angiography and percutaneous transluminal coronary angioplasty at two hospitals in Mashhad-Iran Radiat Prot Dosimetry, February 1, 2008; 128(3): 363 - 366. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Struelens, F. Vanhavere, H. Bosmans, R. Van Loon, and M. Geukens Data analysis from a multi-centre, comparative study of angiographic examinations leading to practical guidelines for the optimisation of patient doses Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 87 - 92. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |