BJR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

British Journal of Radiology (2004) 77, 479-487
© 2004 British Institute of Radiology
doi: 10.1259/bjr/99356178

This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Onnasch, D G W
Right arrow Articles by Kramer, H-H
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Onnasch, D G W
Right arrow Articles by Kramer, H-H

Full Paper

Optimization of radiographic parameters for paediatric cardiac angiography

D G W Onnasch, PhD A Schemm, MD and H-H Kramer, MD

Paediatric Cardiology and Biomedical Engineering, University of Kiel, Schwanenweg 20, 24105 Kiel, Germany


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
In the paediatric cardiac catheterization laboratory the reduction of the radiation dose of diagnostic and interventional procedures is of high priority. Therefore, we performed an experimental study for optimizing the automatic exposure control (AEC) for cardiac angiography. With a Philips Integris BH 5000 system, six AEC programs were configured to acquire X-ray images of 8 cm to 18.5 cm thick PMMA phantoms at tube voltages between 50 kV and 90 kV, with 0.2 mm or 0.4 mm Cu filters and with or without an anti-scatter grid. At constant detector dose, entrance dose (ED) and image quality were evaluated as functions of the voltage. Changes in image quality were determined by the differential signal-to-noise ratio measured within regions of low (SNRb) and high (SNRd) attenuation. At equal voltages, ED saving was approximately 29% with the 0.4 mm Cu beam filtering as compared with 0.2 mm Cu, largely independent of object thickness. SNRb and SNRd were only dependent on the voltage. While SNRb was high at low voltages, SNRd showed a maximum at approximately 79 kV. Using a grid, ED increased with increasing object thickness by a factor of 1.9 to 3.5. At equal voltages, the grid led to significant image improvements, with SNRb and SNRd increasing by 27% and 11%, respectively. SNRb and SNRd are useful descriptors of the image quality in cardiac angiography. Highest image quality was found with tube voltages between 55 kV and 77 kV, independently of object thickness. To minimize dose, the thickness of the copper filter should be chosen to be as large as possible provided the tube's power limit allows keeping the voltage below the upper limit. In view of the substantial image improvement, the use of a grid is recommended for all patients, even for newborns.


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
Despite developments of other imaging modalities, X-ray examinations in the cardiac catheterization laboratory are still indispensable but deliver relatively high radiation doses [15]. The number of interventional procedures, which may cause long examination times, is increasing tremendously. For a specific dose–area product (DAP), the effective dose E, which is an estimate of the added risk for cancer, is highest in small patients [69].

In addition, the radiation risk increases progressively with younger age groups. Nevertheless, the available angiographic systems are supplied with X-ray exposure controls configured for adult patients. In this study we try to lower X-ray body dose and improve image quality in paediatric cardiac angiography. For a commercial angiography system, the essential factors that can be controlled are X-ray beam filtering, the tube voltage and the use of an anti-scatter grid.

Guidelines for equipping cardiac catheterization laboratories [10] recommend providing for 0.1 mm Cu filtering and a removable grid. In the literature [1114], there is an on-going controversy concerning the use of grids for moderately sized paediatric patients as well as the type of filtering [4, 13, 1517]. Technical advances in the development of high output X-ray tubes with liquid bearing systems allow examinations to be performed at higher currents and enable changes to be made in the X-ray spectrum by using heavy copper filtration [15, 17, 18]. Barkhausen et al [15] and Geijer et al [16] demonstrated that optimization of beam filtering reduces the radiation dose without compromising image quality in adults. Seifert et al [19] also observed significantly decreased radiation dose in fluoroscopic and radiographic procedures due to low-energy filters. In phantom experiments they estimated the change of image quality by analysing the change of absolute brightness values for steps of different X-ray density. However, a digital system allows the user easily to adjust the brightness scale at the display stage. Brown et al [14] reported on phantom experiments in different fluoroscopy modes and adapted the automatic exposure control (AEC) of their fluoroscope, which was initially configured with adult fluoroscopy settings, to the requirements in paediatrics. They described image quality by visually comparing images of contrast mesh objects.

Thicker filters within the X-ray beam of systems using an AEC generally increase the tube voltage. From the literature, it is not clear to what extent the reported dose reduction is linked to the modified pre-filtration and how much to the raised voltage. As a higher voltage leads to reduced image contrast, a technique using copper filters is not always adopted in clinical routine [20].

In order to estimate optimal filtration, tube voltage and grid usage for patients of different sizes, Tapiovaara et al [13] used a Monte Carlo computational model of the fluoroscopic image chain. The image quality was quantified by the observer's signal-to-noise ratio (SNR) of a low-contrast image. They came to the remarkable conclusion that the tube voltage should be adjusted as low as 50 kV in order to get the highest image quality with the lowest radiation dose. Recently Fenner et al [17] confirmed Tapiovaara's results in phantom experiments. They assessed image quality by counting the number of discs visible on standard contrast sensitivity phantoms.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
Image optimization is task dependent. In cardiac angiography the entrance dose per frame at the X-ray image intensifier (XRII) is low compared with all other radiographs. Accordingly, quantum noise is dominant. Figure 1Go gives a typical example of biplane recorded frames stored with a resolution of 512 x 512 x 8 bits. The detection of small structures, which is limited by noise, is equally important in areas of high and low radiation absorption. The utilized brightness range is approximately 130.



View larger version (89K):
[in this window]
[in a new window]
 
Figure 1. Biplane images from a cardiac angiogram (recorded at 25 frames s–1, 23 cm field of view, using a grid and 0.4 mm Cu filtration, after injection of 20 ml Imagopaque 350 (Amersham Health, Princeton)) of a 6-year-old boy (32 kg), treated as an infant because of transposition of the great arteries, showing the end-systolic state of the left ventricle in two views: (a) 10° caudal/16° right anterior oblique, source–image distance (SID) 92 cm, 66 kV, 428 mA, 4 ms; (b) 19° cranial/70° left anterior oblique, SID 107 cm, 70 kV, 497 mA, 5 ms. The mean brightness levels are indicated for 8 regions of interest.

 
Independent adjustments of tube current, peak voltage, filtering, focal spot size and detector dose cannot be accomplished with a computer controlled radiographic system. We took advantage of the AEC circuit of the system (Integris BH5000; Philips Medical Systems, The Netherlands), which keeps the dose at the XRII constant for all objects within a certain range of attenuation. This occurs as long as the test object shows a similar dynamic range to the cardiac angiograms. The variety of radiographic parameters was achieved by using different kV/mA characteristics.

With the Philips system for cardiovascular imaging the following settings can be configured or selected:

Experimental set-up
According to the acceptance certificate of the system, the AEC circuit keeps the detector dose rate constant at 0.12 µGy per frame for the 23 cm XRII field of view used in cinepulse acquisition mode. We measured values between 0.17 and 0.22 µGy. Other selectable fields of view (17 cm and 13 cm) were not used in these measurements. The acquisition frame rate was set to 12.5 frames s–1. The phantom was positioned in the isocentre of the gantry with a distance of 76.5 cm from the X-ray tube. The tube to XRII distance was 100 cm.

Six AEC programs, P1 to P6, were configured using two copper filters (P1, P3, P5: 0.4 mm Cu; and P2, P4, P6: 0.2 mm Cu) and different kV/mA curves (Figure 2Go). For the Philips system the minimum current, the maximum power and the voltages that determine the linear transition between these two domains define these curves. The pulse times were 6 ms for P1 to P4 and 4 ms for P5 and P6. The nominal focal spot size was 0.8 mm for the programs P2 and P3 and 0.5 mm for the remaining ones. All other parameters of the image chain were kept constant. During the first seven pulses of each run the tube current and peak voltage changed along the selected curve to reach the pre-programmed XRII entrance dose. For each recording the adjusted values of the X-ray peak voltage and tube current were entered into the log. It turned out that with thick phantoms the mA–kV product attained only about 80% of the configured maximum power. This may be due to some pre-programmed power reserve to avoid excessive heat at long runs.



View larger version (22K):
[in this window]
[in a new window]
 
Figure 2. Automatic exposure control curves P1 to P6 configured to arrive at a variety of tube voltages for angiographic acquisitions.

 
Figure 3Go depicts the irradiation geometry. Four rectangular PMMA objects (25 x 25 cm wide and 8, 11, 15.5 and 18.5 cm thick) were used and image sequences with and without an anti-scatter grid were recorded. Contrast structures were a piece of lead (black value, BV), a 40 mm broad copper step wedge (level thickness 0.35 mm, steps S0 to S5) and a 10 mm aluminium strip (thickness 2 mm, step S5+) (Figure 4Go). Each acquisition was taken within the control range of the AEC.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 3. The irradiation geometry. The semiconductor dosemeter is fixed 9 cm below the phantom, composed of polymethylmethacrylate plates and a contrast structure.

 


View larger version (39K):
[in this window]
[in a new window]
 
Figure 4. The test structure consisting of a copper step wedge, a strip of aluminium and a piece of lead. The radiation detector projected right, below the step wedge.

 
The entrance dose (ED) was measured by using a digital, battery operated semiconductor dosemeter (Conny, PTW Freiburg, Germany). It is calibrated in air kerma at 70 kV. According to the calibration certificate, the indication is 1% higher at 60 kV and 7% lower at 100 kV, which was not considered in our results. The dosemeter was placed in front of the scattering object below the table (Figure 3Go). The distance between focus and radiation detector was 69 cm. ED is the measured dose divided by the number of pictures (approximately 50) of the respective run.

Image evaluations
The analysis of one digital image (512 x 512 x 8 bit) taken from each run was done without any pre-processing. For a total of 48 pictures (four object sizes, with and without a grid and with six AEC programs, three with 0.2 mm and three with 0.4 mm Cu filtering) the mean grey levels and their variances were measured in eight regions of interest (ROIs) on a digital workstation. They were classified as S0 to S5, S5+ and BV as illustrated in Figure 4Go. Each ROI was 48 x 48 pixel in size and all ROIs were placed individually for each image.

The lookup table optimized by Philips for image display was not changed. A dark region corresponds to an area of high radiation attenuation. The brightness range (BR) was calculated as the difference between the brightest region S5 and the black value BV in the unprocessed image.

Two differential SNRs (or inherent contrast-to-noise ratios) were defined on the basis of the measured grey levels. They were derived for a dark (SNRd) and a bright (SNRb) region of the radiographs as the ratios of the difference of two adjacent grey levels divided by the square root of the sum of the variances {sigma}2 in those regions: Go


The difference S1–S0 was not evaluated because it was undetectable in some images.

The effect of the grid on image quality was analysed quantitatively by using the signal-to-noise improvement factor SIF [21] for the dark and bright region calculated as Go


The Bucky factor (BF) was calculated as the ratio of the ED measured with and without a grid.

The quality of the image of the thin wires in the radiation detector of the dosemeter was judged subjectively (see Figure 5Go, bottom right corner of each picture).



View larger version (81K):
[in this window]
[in a new window]
 
Figure 5. Comparison of two images of the 11 cm phantom, recorded under different radiographic parameters (see Table 1Go). For identifying the steps S0 to S5 of the step wedge see Figure 4Go for comparison.

 

    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
As an example, Figure 5Go illustrates two images of the 11 cm phantom recorded with 56 kV and 0.2 mm Cu (left image) and 76 kV and 0.4 mm Cu (right image), respectively. The quality of both exposures was acceptable, while ED of the left image was three times as high as with the right one. As presented in Table 1Go, SNRd increased from left to right, whereas SNRb decreased.


View this table:
[in this window]
[in a new window]
 
Table 1. Radiographic parameters of the images presented in Figure 5Go

 
Radiation dose
Figure 6Go presents the rapid decay of ED with rising voltage on a semilog scale. For the same object, ED varied approximately by a factor of 3, depending on the selection of beam filtering and tube voltage. Tables 2 and 3GoGo summarize all the results for the four phantoms measured with and without a grid.



View larger version (14K):
[in this window]
[in a new window]
 
Figure 6. Entrance dose versus X-ray voltage for the four objects (circles: 8 cm; squares: 11 cm; rhombi: 15.5 cm; and triangles: 18.5 cm) recorded with a grid using the automatic exposure control programs P1 to P6. Solid lines: 0.2 mm Cu filtering (programs P2, P4, P6, from left to right); dashed lines: 0.4 mm Cu filtering (programs P3, P5, P1, from left to right).

 

View this table:
[in this window]
[in a new window]
 
Table 2. Acquisition data, entrance doses per frame (ED), signal-to-noise ratios in regions of low (SNRb) and high (SNRd) attenuation as well as the brightness ranges (BR) for four objects of different thickness recorded with a grid mounted at the image intensifier. The names P1 to P6 of the programs of the automatic exposure control (AEC) refer to Figure 2Go

 

View this table:
[in this window]
[in a new window]
 
Table 3. Acquisition data, entrance doses, signal-to-noise ratios (SNRs) for images recorded without an anti-scatter grid. See Table 2Go for abbreviations

 
ED was significantly reduced, independently of the thickness of the object, when the 0.4 mm Cu filter was used. When we compared all measurements with the two filters we found the ratios 0.539 with a grid (n=24) and 0.538 without a grid (n=24). However, it is possible that an increasing tube voltage due to the more absorbent filter causes a part of the dose reduction.

Using Tables 2 and 3GoGo, several isovoltage parameter pairs (±1 kV) can be extracted for different filter settings. On average, the ratio of the doses with 0.4 to 0.2 mm Cu filtering was 0.714 (±0.136 standard deviation) with a grid (n=6) and 0.739 (±0.022) without a grid (n=8).

Image quality
BR decreased slowly with the object thickness (Tables 2 and 3GoGo). By omitting the grid, BR was reduced by 35% (±10%, n=24) on average. It can easily be re-scaled by grey level windowing. SNRd and SNRb did not change thereby.

SNRb and SNRd showed different dependences on X-ray voltage (Figure 7Go). SNRb (dashed curve) was optimal for low voltages and decreased rapidly with rising voltages. SNRd became small at low voltages, and the signal difference between the darkest Cu steps (S0, S1) vanished in the high amplitude quantum noise. In our experiments this was the case for the 0.2 mm Cu filter and small objects. SNRd (solid curve in Figure 7Go) had its maximum at approximately 79 kV.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 7. The signal-to-noise ratios SNRb (triangles) and SNRd (squares) of the images recorded with 0.2 mm (filled symbols) and 0.4 mm Cu filtering (open symbols). All data from radiographs with an anti-scatter grid.

 
When only measurements at equal voltages (±1 kV) were compared, SNRb for 0.2 mm Cu was somewhat higher than for 0.4 mm Cu filtering. In contrast, SNRd showed more favourable changes with 0.4 mm than with 0.2 mm Cu, both for the evaluations with (Table 2Go) and without a grid (Table 3Go). These differences are small and can hardly be seen in Figure 7Go.

When all the SNRb data for the two filters, 0.2 and 0.4 mm Cu, were averaged it seemed that using the thinner filter leads to a higher SNR in the area of low attenuation. However, there is a simultaneous shift of mean voltage for these two sets, which is responsible for this difference. Conversely, SNRd, measured in the high attenuation (i.e. dark) area of the image, seemed to increase by the transition from the 0.2 to the 0.4 mm Cu filter, when the mean values of all measurements from Tables 2Go or 3 were compared. In this case the associated voltage shift seems to favour the data for the thicker filter. It is therefore important to separate the effects of filter thickening and voltage change on image quality.

Effect of an anti-scatter grid
The increase of patient dose caused by the grid can be retrieved from the Tables 2 and 3GoGo. The average BF of all measurements with the same object and AEC program was 1.86 (±0.33 standard deviation). BF showed only a weak correlation with object thickness (r=0.745). However, measurements at different voltages were mixed for this analysis. For equal object thickness and filtration, the AEC selected a lower voltage when no grid was used.

In 10 measurements the voltages were equal or nearly equal (±1 kV) with and without a grid. Results are summarized in Table 4Go. The average BF was 2.58 (±0.74). Dose increase was smallest for the thin object (factor 1.9) and largest for the thickest object (about 3.5).


View this table:
[in this window]
[in a new window]
 
Table 4. Effect of a grid on dose and image quality. BF: ratio of the entrance doses measured with and without a grid at the given voltage and filtering. SIFd and SIFb: signal-to-noise improvement factors (Equation 2Go) for the regions of high and low attenuation, respectively

 
For all objects, both SNRd and SNRb showed higher values when a grid was used (Tables 2 and 3GoGo). The average SIFs were SIFd=1.12 and SIFb=1.05 (n=24). Selecting only those 10 measurements where the voltages were equal (Table 4Go), gave SIFd=1.11 and SIFb=1.27 on average. A dependency on object thickness and filter thickness was not recognized.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
Controlling the radiographic factors
To keep radiation dose as low as reasonably achievable (ALARA) a well-configured AEC of the radiographic system is essential. Modern digital systems are equipped with such control circuits to facilitate the acquisition of well-exposed radiographs. In paediatric cardiac angiography different AEC programs should be configured which are optimized for children of different ages. The described phantom experiments were designed to find the optimal radiographic settings.

The availability of high heat load X-ray tubes allows the spectrum to be changed by additional filtration, reducing the peak voltage and increasing the tube current [15]. Thus, the contribution of low-energy photons which do not penetrate the patient to form the image is reduced. While the positive effect has been demonstrated for fluoroscopy in adults [4, 15, 16, 18, 19] and children [1, 14, 17], it was not clear how the tube voltage should be chosen when large areas of the images are opacified by iodine contrast material as in cardiac angiography.

Conventional X-ray units allow only a limited change of the radiation settings [14]. Reproducible adjustments of X-ray voltage, current and pulse length are difficult to make. However, the parameters used to configure different AEC programs for objects of different sizes can be assessed on the basis of the physical and technical knowledge about generation, interaction and detection of X-rays [22]. The AEC curves (Figure 2Go) used in this study were configured on the basis of some rough assessments in order to get a broad variety of radiographic performance.

For the final application of a certain radiographic settings, phantom measurements are indispensable. We selected simple structured test phantoms, which showed similar attenuation and contrast values as clinical angiograms of children of all age groups (Figures 1 and 5GoGo). Under these "natural" conditions, ED and image quality were compared, while the XRII entrance dose was kept constant at a rather low level.

Image quality versus dose optimization is always task dependent. In contrast to chest radiography, where anatomic noise limits the detection of subtle lung nodules [23], in cardiac angiography the detector dose per frame is so small that quantum noise is limiting the recognizability of small structures.

Radiation dose quantities
In clinical routine, the patient's dose can easily be measured by recording the DAP, from which the entrance surface dose (ESD) for a given geometry can be calculated. In the selected experimental set-up (Figure 3Go), which closely meets the clinical setting, the ED is measured directly. Although there is a gap of about 9 cm between the phantom and the dosemeter, some backscatter is recorded beside the primary radiation, the more the higher the voltage. On the other hand, the semiconductor dosemeter is less sensitive to higher energies. Overall, we regard ED as a useful estimate of the patient's entrance dose for the purpose of optimization.

DAP and ESD, however, do not give good measures of relative risk of stochastic and deterministic effects. Also, these values cannot be directly compared when measured with different X-ray beam qualities, organs and projections. The effective dose E, which is a weighted sum of dose equivalents received by each organ of interest, is the appropriate radiation dose quantity. However, provided that the detected dose at the image entrance field is constant, a minimal ED value is directly related to a minimal E. Therefore, for optimizing the radiographic parameters, the measured relative ED can be used as an appropriate quantity.

In recent years there has been some effort directed to converting DAP and ESD to E [5, 6, 9, 2426]. There is still limited information on the sensitivity of the conversion factors E/DAP and E/ESD to variation of tube voltage and beam filtration. According to Wise et al [24] for chest views of adults, a 10% increase in tube voltage leads to 5% increase of E/ESD. Schmidt et al [25] found similar results measured with paediatric phantoms. In addition, they report on an increase of E/DAP of approximately 30% by adding 0.1 mm Cu. In conclusion, for a given ESD or ED, respectively, E increases with rising voltage and filter thickness. For a high ED value (low voltages, moderate filtration) there is a steep exponential decrease of dose within the phantom (or patient) to reach the required XRII dose value. Conversely, for a relatively low measured ED (high voltages, strong filtration), the dose within the patient decreases slowly towards the fixed image dose value. Consequently, the saving of E may be up to 40% smaller than expected from the ED values presented in Figure 6Go.

Descriptors of image quality changes
In high quality X-ray equipment for cardiac angiography, image noise is mostly due to the limited number of detected X-ray quanta. In this domain spatial resolution can only be enhanced by a better detector efficiency or higher entrance doses at the XRII. Improvements can be demonstrated by visualization of bar-pattern or mesh objects [9, 14, 16]. If the XRII dose is kept constant as in our experiments, image quality must be quantified differently. Contrast resolution can be measured with contrast sensitivity phantoms like the Leeds TOR and N2 phantoms [4, 17] or the CDRAD phantom [27]. However, small changes of the grey level distribution can hardly be analysed by such subjective methods.

The SNR [13, 21, 26, 28], which is proportional to the square root of the XRII dose, allows improved comparison of changes in image quality. Due to the strong dependence between attenuation and the energy of the X-ray quanta, any change of the tube voltage or beam filtration changes the grey level distribution of a structured radiograph. As image noise is signal dependent and different in areas of high and low attenuation, we measured the SNR in regions of low and high brightness, in order to analyse changes of the image characteristics. In adult and paediatric cardiology, the recognition of small brightness differences in darker regions is often as important as in the bright areas (Figure 1Go). Only for voltages of at least 55 kV the SNR is high enough to enable an observer to detect small objects within the opacified ventricle.

As in our experiments most components of the radiographic system did not vary, only relative comparisons were needed, which reflect the changes of attenuation and scattering reproducibly. For this task the differential SNR was an adequate measure, which was observed in regions of interest of both low and high attenuation. The simple copper step wedge that we used fulfilled all the requirements. For the observer the most striking fact is the change of the BR or contrast of the images acquired under different conditions. BR depends on the thickness of the object, the X-ray voltage and most of all, the use of the grid (Tables 2 and 3GoGo). Compared with the findings of Seifert et al [19], BR did not significantly decrease because of using a thicker filter. A change was only observed when tube voltages altered. Contrary to analogue imaging, a small BR is not a genuine problem as one can easily adjust the contrast at the display stage. However, this does not improve the underlying SNR. The parameters SNRd and SNRb did not depend on linear grey level windowing. This demonstrates that these ratios (Equation 1Go) are robust descriptors of the quality of image generation and detection.

We separated the effects of changes in tube voltage and in filtering on dose reduction and image quality. For equal voltages, the dose saving was about 29% by switching from 0.2 to 0.4 mm Cu. As seen from Figure 7Go, there was no significant change in the SNRs ratios for these two filters, neither for SNRd, nor for SNRb.

Tapiovaara et al [13] determined the observer's SNR for low contrast objects and concluded that a voltage as low as 50 kV is optimal for imaging with diluted iodine contrast material. But the restriction to low contrast objects does not seem to be sufficient when beam filtering or voltages are altered. Our measurements show that, for such a low voltage, SNRb is indeed maximized (Figure 7Go, dashed curve). However, at 50 kV the steps S0, S1 and S2 (Figure 4Go) were all black. In the image presented in the left pane of Figure 5Go, taken at 56 kV, the border between steps S0 and S1 was already undetectable. No information can be retrieved from darker regions in a cardiac angiogram.

The SNRs could also be measured at the intermediate steps of the phantom. The curves shown in Figure 7Go would fade into each other, yielding an almost horizontal curve for the SNR determined with steps S2 and S3.

The use of a thicker copper filter results in significant dose reduction without impairing the image quality. This holds as long the voltage is not increased simultaneously, which is usually the case when an additional filter is put into the beam. Therefore, to keep the image quality at a high level, the AEC has to be adapted to make use of the high output of advanced tubes. The AEC programs should be configured in a way that the X-ray voltage does not fall below 55 kV. On the other hand, approximately 77 kV should not be exceeded to avoid the decrease of SNRb. With adult patients the stronger filtering may necessitate larger focal spots (nominal 0.8 mm or larger instead of 0.5 mm), which leads to a reduced spatial resolution. This observation is an argument against using even thicker filters.

Use of an anti-scatter grid
As the use of a grid for moderately sized paediatric patients is judged differently [1014], we also investigated its effect on dose and image quality. ED rose by an average factor of 2.6 when the voltage was kept constant. For the smallest phantom, equivalent to a newborn, it increased by a factor of about 1.9. Regarding the image quality improvement as quantified by SIFb and SIFd, the grid should be used for all patient sizes. For equal voltages the SNR improves within the bright (less absorbing) area of the image by approximately 27% and in the dark area by 11% (Table 4Go). This result is in agreement with visual comparisons of optimal scaled images recorded with and without a grid.

The grid must be inserted or removed before the patient is put on the table. Often there is a large variation of imaging geometry even during the same cardiac examination, especially in the lateral view. Therefore, a prior decision for or against using the grid is needed. In our experiments with the smallest phantom the air-gap between the distal side of the phantom and the XRII was relatively large. As a result, the relative contribution of scattered radiation was smaller than for the thicker phantoms. This may explain the low signal-to-noise improvement of SNRb for larger voltages and small objects (Table 4Go).

In view of the variable conditions we conclude that in accordance with Tapiovaara et al [13] and Brown et al [14] the rise of ED should be accepted in view of the substantial image improvement. This recommendation stands in contrast to the national and international guidelines [1012] for paediatric radiography.

Potential limitations
The transfer of our results to clinical settings is limited by the fact that they were obtained with static phantom experiments at a fixed frame rate of 12.5 frames s–1 and a field of view of 23 cm. In addition, we used copper absorbers instead of iodine contrast material as applied in angiography. The different attenuation characteristics of iodine may modulate the curves depicted in Figure 7Go. However, other structures as ribs, heart muscle, diaphragm and interventional devices furnish further disturbing patterns. The ranges of X-ray attenuation of the radiographs were similar to cardiac angiography (Figures 1 and 5GoGo). The applied AECs have also been used routinely during the last year in our cardiac catheterization laboratory, where the AEC programs with a thicker copper filter (P3 and P5) were acceptable. Although our experiments did not prove that the results could be transferred to higher frame rates (25 and 50 frames s–1) and other fields of view, we see no reason why changes should occur.


    Conclusions
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
With a fixed input exposure at the XRII, a variety of radiographic parameters was achieved by using different AEC settings. Thus, the patient exposure and image quality could be analysed as being dependent on beam filtering and tube voltage. Compared with other measures for objective assessment of image quality, the differential SNR proved to be very effective. For cardiac angiography it should be determined not only within the area of low, but also of high X-ray attenuation.

SNRd and SNRb were functions of tube voltage and to a large degree independent of object size and beam filtration (Figure 7Go). The consequence is that the filtration selected should be as high as possible provided that the power of the X-ray tube allows the voltage to be kept in the range of high image quality. The AEC should be adapted when filtration is increased in order to keep the voltage low. Considering only low contrast structures, this means about 50 kV would be optimal. However, when structures in darker areas must also be detectable, as in cardiac angiography, the optimal voltage should be raised to increase their SNR.

In terms of the ALARA principle for cardiac angiography our results allow the following recommendations:


    Acknowledgments
 
We thank B Hoornaert and H D Nagel, Philips Medical Systems, for their support in programming the automatic exposure control of the X-ray equipment, and C Riedel for fruitful discussions.

Received for publication July 11, 2003. Revision received October 24, 2003. Accepted for publication November 25, 2003.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 

  1. Boothroyd A, McDonald E, Moores BM, Sluming V, Carty H. Radiation exposure to children during cardiac catheterization. Br J Radiol 1997;70:180–5.[Abstract]
  2. Zorzetto M, Bernardi G, Morocutti G, Fontanelli A. Radiation exposure to patients and operators during diagnostic catheterization and coronary angioplasty. Cathet Cardiovasc Diagn 1997;40:348–51.[CrossRef][Medline]
  3. Cusma JT, Bell MR, Wondrow MA, Taubel JP, Holmes DR Jr. Realtime measurements of radiation exposure to patients during diagnostic coronary angiography and percutaneous interventional procedures. J Am Coll Cardiol 1999;33:427–35.[Abstract/Free Full Text]
  4. Nicholson R, Tuffee F, Uthappa MC. Skin sparing in interventional radiology: the effect of copper filtration. Br J Radiol 2000;73:36–42.[Abstract]
  5. McFadden SL, Mooney RB, Shepherd PH. X-ray dose and associated risks from radiofrequency catheter ablation procedures. Br J Radiol 2002;75:253–65.[Abstract/Free Full Text]
  6. Axelsson B, Khalil C, Lidegran M, Schuwert P, Mortensson W. Estimating the effective dose to children undergoing heart investigations – a phantom study. Br J Radiol 1999;72:378–83.[Abstract]
  7. Rassow J, Schmaltz AA, Hentrich F, Streffer C. Effective doses to patients from paediatric cardiac catheterization. Br J Radiol 2000;73:172–83.[Abstract]
  8. Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. Am J Radiology 2001;176:289–96.
  9. Huda W. Effective doses to adult and pediatric patients. Pediatr Radiol 2002;32:272–9.[CrossRef][Medline]
  10. Hamm C, Bosenberg H, Brennecke R, Daschner F, Dziekan G, Erbel R, et al. Guidelines for equipping and managing heart catheter rooms (1st revision). Z Kardiol 2001;90:367–76. [German][CrossRef][Medline]
  11. Bundesärztekammer. Guidelines of the Bundesärztekammer for quality assurance in radiography. Deutsches Ärzteblatt 1995;92:2272–85. [German]
  12. European Commission. European guidelines on quality criteria for diagnostic radiographic images in paediatrics. Report EUR 16261, Office for Official Publications of the European Communities, L-2985 Luxembourg 1996.
  13. Tapiovaara MJ, Sandborg M, Dance DR. A search for improved technique factors in paediatric fluoroscopy. Phys Med Biol 1999;44:537–59.[CrossRef][Medline]
  14. Brown PH, Thomas RD, Silberberg PJ, Johnson LM. Optimization of a fluoroscope to reduce radiation exposure in pediatric imaging. Pediatr Radiol 2000;30:229–35.[CrossRef][Medline]
  15. Barkhausen J, Schoenfelder D, Nagel HD, Stöblen F, Müller RD. Optimization of beam filtering, kV-mA regulation curve and image intensifier entrance exposure rate to reduce radiation exposure in angiographic fluoroscopy. Fortschr Röntgenstr 1999;10:391–5. [German]
  16. Geijer H, Beckman KW, Andersson T, Persliden L. Radiation dose optimization in coronary angiography and percutaneous coronary intervention (PCI). I. Experimental studies. Eur Radiol 2002;12:2571–81.[Medline]
  17. Fenner JW, Morrison GD, Kerry J, West N. A practical demonstration of improved technique factors in paediatric fluoroscopy. Br J Radiol 2002;75:596–602.[Abstract/Free Full Text]
  18. den Boer A, de Feyter PJ, Hummel WA, Keane D, Roelandt JR. Reduction of radiation exposure while maintaining high-quality fluoroscopic images during interventional cardiology using novel x-ray tube technology with extra beam filtering. Circulation 1994;89:2710–4.[Abstract/Free Full Text]
  19. Seifert H, ElJamal A, Roth R, Urbanczyk K, Kramann B. Reduction of the radiation exposure of patients caused by selected interventional and angiographic procedures. Fortschr Röntgenstr 2000;172:1057–64. [German][CrossRef]
  20. Wraith CM, Martin CJ, Stockdale EJN, McDonald S, Farquhar B. An investigation into techniques for reducing doses from neo-natal radiographic examinations. Br J Radiol 1995;68:1074–82.[Abstract/Free Full Text]
  21. Sandborg M, Carlsson GA. Influence of x-ray spectrum, contrasting detail and detector signal-to-noise ratio (SNR) and detective quantum effciency (DQE) in projection radiography. Phys Med Biol 1992;6:1245–63.[CrossRef]
  22. Boone JM. X-ray production, interaction, and detection in diagnostic imaging. In: Beutel J, Kunkel HL, van Metter RL, editors. Handbook of medical imaging, volume 1: Physics and Physiophysics. Bellingham, USA: SPIE Press 2000;1–78.
  23. Samei E, Flynn MJ, Eyler WR. Detection of subtle lung nodules: relative influence of quantum and anatomic noise on chest radiographs. Radiology 1999;213:727–34.[Abstract/Free Full Text]
  24. Wise KN, Sandborg M, Persliden J, Alm Carlsson G. Sensitivity of coeffcients for converting entrance surface dose and kerma-area product to effective dose and energy imparted to the patient. Phys Med Biol 1999;44:1937–54.[CrossRef][Medline]
  25. Schmidt PWE, Dance DR, Skinner CL, Castellano Smith IA, McNeill JG. Conversion factors for the estimation of effective dose in paediatric cardiac angiography. Phys Med Biol 2000;45:3095–107.[CrossRef][Medline]
  26. Launders JH, Cowen AR, Bury RF, Hawkridge P. Towards image quality, beam energy and effective dose optimisation in digital thoracic radiography. Eur Radiol 2001;11:870–5.[CrossRef][Medline]
  27. Busch HP, Busch S, Decker C, Schilz C. Image quality and exposure dose in digital projection radiography. Fortschr Röntgenstr 2003;175:32–7. [German][CrossRef]
  28. Dobbins JT III. Image quality metrics for digital systems. In: Beutel J, Kunkel HL, van Metter RL, editors. Handbook of medical imaging, volume 1: Physics and Physiophysics. Bellingham, USA: SPIE Press 2000;161–222.



This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
V. Tsapaki, S. Kottou, S. Korniotis, N. Nikolaki, S. Rammos, and S. C. Apostolopoulou
RADIATION DOSES IN PAEDIATRIC INTERVENTIONAL CARDIOLOGY PROCEDURES
Radiat Prot Dosimetry, January 2, 2009; (2009) ncn313v1.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
D G W Onnasch, F K Schroder, G Fischer, and H-H Kramer
Diagnostic reference levels and effective dose in paediatric cardiac catheterization
Br. J. Radiol., March 1, 2007; 80(951): 177 - 185.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
J Partridge, G McGahan, S Causton, M Bowers, M Mason, M Dalby, and A Mitchell
Radiation dose reduction without compromise of image quality in cardiac angiography and intervention with the use of a flat panel detector without an antiscatter grid
Heart, April 1, 2006; 92(4): 507 - 510.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Onnasch, D G W
Right arrow Articles by Kramer, H-H
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Onnasch, D G W
Right arrow Articles by Kramer, H-H


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS