| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Full Paper |
1 Medical Physics Department, Torbay Hospital, Newton Road, Torquay, Devon TQ2 7AA and 2 Medical Physics Department, Royal Marsden NHS Trust, Downs Road, Sutton, Surrey SM2 5PT, UK
Correspondence: Dr R Wilks, Burraton Bungalow, Broadclyst, Exeter, Devon EX5 3DB, UK
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
If no direct correction is made for the presence of the lung, some allowance for the increased dose may be made by adjusting the field weightings and wedge angles in order to bias the dose distribution towards the breast apex. Preferably, however, an inhomogeneity correction to the plan is employed to give the actual dose distribution. A knowledge of the size, shape and density of the lung is required in order to perform the lung correction accurately, and this is usually performed using CT data, either from a CT scanner or a simulator-CT system.
The use of a CT scanner for the imaging of breast patients presents difficulties, however, because of the problem of maintaining the same treatment position within the limited CT aperture. The abducted arm (if used), and the effect of the slope of the inclined breast board both tend to move the patient towards or beyond the confines of the CT aperture, so that compromises on patient position may be required in practice [36]. Also many centres have limited access to a general-purpose CT scanner, for the large number of breast patients presenting for treatment. A simulator-CT is able to accommodate the treatment position satisfactorily, although the time to acquire the data is a consideration, e.g. delays due to the cooling of the X-ray tube.
The Osiris system (Qados Ltd, Sandhurst, UK) [7, 8] attempts to circumvent the use of a CT image entirely, by measuring the patient contours optically (either on a simulator or the treatment unit itself), and fitting the lung shape mathematically with the assistance of a portal image obtained using one of the tangential fields. The process is relatively quick and is easily repeatable on either the simulator or the treatment unit.
In our two centres (Torbay and Royal Marsden), the complete process of radiotherapy breast planning is normally carried out with the Osiris system. This includes the measurement of patient contours, geometrical field placement calculation, isodose production and monitor unit (MU) calculation. An inherent part of this is the estimation of lung position from the mathematical fitting algorithm.
The purpose of this paper is to report on the comparison of the fit of the calculated lung shapes with those measured for the same patients on a CT scanner, together with the estimated and measured dosimetric consequences of the geometrical differences found. Data from the independent trials of each centre were combined for this study.
| Materials and methods |
|---|
|
|
|---|
An initial lung model, which had been in use on the system for a number of years, was to fit the lung with an ellipse. From observation of a large number of CT scans, it was clear that this was a reasonable approach, provided the external shape of the patient was also used to determine the shape of the ellipse. Moreover, it was only necessary to provide a fit in the area within the tangential fields. However, as a result of the collaboration between our two centres, the results obtained when analysing the joint data were used to modify the methodology of the lung fitting. This somewhat simpler model required the fitting of an arc of a circle to the anterior portion of the lung.
Figure 1
shows the position of the circle relative to the mediallateral back edge (MN) of the tangential fields. When a portal image is taken, the position and angle of the back edge is known. The image is used to establish the amount of lung, p, in the field by means of a line which is a tangent to the shape of the lung, i.e. it touches the edge of the lung without intersecting with it (line FG in Figure 1
). The beam divergence of line FG relative to MN is small and is ignored. For the fitted lung shape, the radius, a, of the circle was calculated directly from the measured value of p and the value of lung path length, L, by the relationship: a=p/2+L2/8p. The value of L was determined from a linear fit to the CT data, as described below (Figure 2
). In practice, the values of the lung, p, for each longitudinal position of transverse outline may be read automatically from a digitized portal image, to reduce the operator workload when multiple transverse surface contours are taken.
|
|
After the size and position of the fitted lung was calculated, a mathematical check was made on the minimum distance between the fitted lung and the surface contour. This was calculated automatically for each patient and when the lung was closer to the surface than the pre-defined minimum distance, the radius of the circle was decremented in 1 mm steps until the minimum distance was satisfactory or until the total radial decrement had reached 20 mm.
The tissue thickness, t, was given a nominal minimum value of 15 mm which was chosen from observation of the CT images. If there was a small amount of lung in the fields (i.e. a small value of p) the calculated value of t was greater than the nominal minimum value. As the value of p increases, so t decreases, and is normally limited to 15 mm. However, some "chest wall" patients have a large amount of lung and a thin chest wall. A slightly improved fit in the medial region was obtained for these patients by decreasing the value of t to a minimum of 10 mm. This calculation was performed automatically from the comparison of the surface contour and the fitted lung shape.
Verification of lung fitting
Entry of patient geometry
At one centre (Royal Marsden), a series of 20 CT patients were entered into the Osiris system via tracing of suitable hardcopies (central slice only), with the depth of lung on the central axis being entered via the computer keyboard. A hardcopy was produced of the external contour and lung for comparison with the CT hardcopy. Three measurements were taken on the hardcopies of the CT and Osiris outlines: the depth of lung, the maximum width of lung intersecting the baseline of the tangential fields and the shift along the back edge of the centre of the Osiris-fitted lung shape with respect to the CT lung.
At the other centre involved in this study (Torbay), CT scans of chest patients were read into the Osiris system electronically and planned as if for breast tangential treatment. Several slices of each patient were used, leading to a total of 46 images for 17 patients.
Dosimetry
Two planning computers were used to assess the effect on the dosimetry of any inaccuracies of the lung fitting algorithm:
1. ADAC Pinnacle, Milpitas, CA, USA (Royal Marsden)
2. GE Target-2, Prism Microsystems Ltd, Bickington, UK (Torbay)
The CT outlines were planned and compared with those produced using the lung fitting method described above. For both planning computers, the breast tissue was assigned a bulk density of 1.0 and a lung tissue value of 0.3. The Pinnacle system used a collapsed cone algorithm, which corrects for the lack of lateral scatter at the anterior of the breast outline. The Target-2 computer planner makes no such correction, but the lack of scatter was partially compensated for by increasing the calculated monitor units by 2.0%, as determined experimentally. The lung corrections made were of the simple equivalent path length method [9].
For both sets of plans from each centre, representative points within the breast on either side of the lung were taken in order to assess the dosimetric effects of the presence of lung. These points were chosen to be at a distance of 7 mm from the mediallateral line so as not to be within the penumbra of the tangential fields, and were mid-way between the patient surface and the lung contour. A nominal beam energy of 6 MV was used for all the planning calculations.
| Results |
|---|
|
|
|---|
From analysis of the lung positions in the CT image data set, the fitted lung positions were modelled by the simple expression: v=0.4 x (MN+0.15) x (LLmin), where v was the distance of the centre of the lung along MN as measured from M, and Lmin was the extrapolated value of the fitted line of Figure 2
on the vertical axis (nominally 40 mm). The maximum value of v calculated by this formula was limited to 0.5 x MN, so that the position of the centre of the lung was always between 40% (small lungs) and 50% (large lungs) of the length MN, as measured from the medial mark. The position of the centre of the fitted lung compared with that of the CT shape showed a mean difference of 0.2 mm away from the medial mark, and a standard deviation of 10.0 mm (52 data points).
A simplified beam model using differential calculus (not discussed here) was used to investigate the relationship between the path length error due to an inexact fit and its associated dosimetric error for a range of patient geometries. Figure 3
summarizes the results found for a calculated dosimetric error of 2.0% in the breast either side of the lung for a given lung density of 0.30. Figure 3
shows the error in lung path length, dL, which is required to produce a 2.0% error for a given lung path length, L, for a family of mediallateral marks separations, s. The variation of dL with lung density is relatively insensitive (e.g. a lung density of 0.25 reduces the point for s=200 mm, L=100 mm from 23.8 mm to 22.3 mm, a slightly tighter restriction). Hence this analysis shows that, to a reasonable approximation, there is an increase in the dose to the breast either side of the lung of approximately 1% for each additional 1 cm of lung. This was confirmed by the results of measurements on the computer plans produced at each centre.
|
Representative examples of the fit obtained in practice are given in Figures 4 and 5![]()
, where Figure 4
shows a good fit, and Figure 5
is the worst fit of lung shape obtained in the combined series of 64 CT slices. Figure 4
uses CT image data directly, whereas the outlines have been traced in from hardcopies of CT scans for Figure 5
, as explained in the section on Entry of patient geometry.
|
|
As a result of this check using a realistic breast and lung phantom, it was concluded that the dosimetric calculations on the combined sets of patient data would be sufficiently accurate for all patients when comparing the effects of lung correction, both for fitted and for CT-derived lung shapes.
Patient planning data
Each patient was planned using the standard breast planning technique of each centre: isocentric and with the back edges of the tangential fields aligned. The reference points described in the section on Dosimetry above were used to assess the dosimetric effects of the fitted lung shapes compared with the lung shapes defined by the CT images.
Figure 6
shows a histogram of dose errors to the reference points both with and without lung correction. When no lung correction was made, the errors were all negative (planned doses were less than actual doses) and range from about 4%, for very little lung in the fields, to 12% when a large amount of lung was present. When lung correction was performed using the fitted lung, the errors were shown to be within about ±2%, as expected from Figure 2
. The means and standard deviations of the two populations were 8.4% ±1.8% and 0.2%±1.0%, respectively. It is encouraging that the corrected histogram was relatively symmetrical around zero, indicating that there was no significant systematic error in the correction for the population as a whole.
|
| Discussion |
|---|
|
|
|---|
a) The lung shape may deviate from an arc
CT scans have shown relatively small deviations for the majority of patients. An exception is shown in Figure 5
, where the lung shape is flattened to produce a longer lung path length than that calculated by the model. However, even in this case, the understimate of breast dose of 2.0% may be compared with an underdosage of approximately 8% if no lung correction had been made.
b) The fitted lung may be at a different position along the mediallateral edge
The dose to both the lung and the breast from summation of the two opposing tangential fields is not a sensitive function of the lung positionthe path length through the lung is of more relevance.
c) Patient rotation between outlining and taking the check film may over- or under-estimate the amount of lung in the tangential fields
Comparisons of the portal images for lung distances, p, and air gaps outside the breast may be made with the original outlines to correct for this source of error. For example, a significantly larger air gap than expected from the surface contour implies the possibility of patient movement in a direction which would underestimate the amount of lung. A simple correction using the measured and expected air gap is used routinely at Torbay to reduce this problem.
d) No allowance is made for the presence of other structures (e.g. cardiac apex)
Caution on applying any correction in radiotherapy also applies in this case, and inspection of the portal image will indicate whether a particular transverse plane will be over- or under-corrected if the lung is taken into account for dose calculation. A larger value of lung density (e.g. 0.5) may be selected in these cases, if desired. Work is in progress to include a fitted shape for the cardiac apex inside the tangential fields.
For treatments using radiation fields other than tangential fields with the back edges aligned (e.g. directly opposing fields) a suitable correction to the amount of measured lung in the portal image may be applied. The resulting lung and breast doses calculated using the fitted lung will then be expected to be of the same accuracy as described in this paper. However, for fields using different gantry angles, for example smaller "top-up" fields, suitable portal images may be used to establish the amount of lung in these fields, if any. The typical displacements found in this study of the fitted lung along the mediallateral line, when compared with the CT image, will have a negligible effect in practice on the doses to the lung and breast for the tangential fields, but may result in an incorrect estimate of the amount of lung within such "top-up" fields. Hence, this lung fitting method should be applied with caution to field arrangements other than simple tangential fields.
In addition, no matter how accurate the lung shape, either measured by CT, ultrasound or modelled in some other way, the patient will breathe during treatment, so that a variable amount of lung will be treated throughout each exposure. If the lung moves through a vertical distance b, the change in lung path length is approximately 2b (obtained form the linear fit of Figure 2
, and assuming a typical gantry angle of 45° from the horizontal). Measured breathing movements may reach 10 mm in a nervous patient, although 5 mm is more typical. Hence breathing may account for about 10 mm in lung path length error, or 1% error in the breast dose. Errors due to patient movements and setup errors will increase the uncertainty, so that a working figure of an expected 2% dose error seems reasonable, and this is of the same order as the maximum dose error due to the lung fitting algorithm described. It thus may be argued that there seems little incentive in attempting to find increased lung-fitting accuracy beyond this simple model.
| Conclusions |
|---|
|
|
|---|
Received for publication January 7, 2003. Revision received October 9, 2003. Accepted for publication November 12, 2003.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E M DONOVAN, J R YARNOLD, E J ADAMS, A MORGAN, A P J WARRINGTON, and P M EVANS An investigation into methods of IMRT planning applied to breast radiotherapy Br. J. Radiol., April 1, 2008; 81(964): 311 - 322. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |