| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Full Paper |
1 Medical Physics Directorate, University Hospital of North Staffordshire NHS Trust, Princes Road, Hartshill, Stoke on Trent ST4 7LN and 2 X-ray Department, Queens Hospital, Belvedere Road, Burton-on-Trent ST16 3SA, UK
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The Ionising Radiation (Medical Exposure) Regulations (IR(ME)R) 2000 require that image quality and radiation dose are optimized for every medical exposure. The relative importance of CT to patient dose in diagnostic radiology is recognised by the further requirement that "special attention" be paid to dose optimization for this practice [4]. However, dose optimization in CT is not straightforward. Variations in equipment design between manufacturers and models, such as the level of filtration, different focusisocentre distances, and variations in collimator and detector efficiency can have a significant impact on dose [5]. Lack of any "automatic exposure control" on most existing CT scanners, combined with the wide dynamic range of digital detectors may allow wide differences in user-selectable factors between different CT units. Factors affecting patient dose include the choice of slice width, pitch, kV, mAs per rotation and the number of slices per examination.
European guidelines on CT imaging have been available for several years [6]. These guidelines detail the required image quality and suggest diagnostic reference levels (DRLs) for nine examinations. Of these reference levels only four have been adopted in the UK [7], and much of the data was collected from a UK survey undertaken in the early 1990s [1], prior to the widespread introduction of spiral scanning.
Previous studies of patient dose in CT have typically included a variety of examinations on a range of different scanners [810], and variations in patient dose in such surveys will be affected by both inherent equipment design and user-selectable factors. However, practical dose optimization for each exposure relates principally to the user-selectable exposure factors.
The purpose of this study was first to identify the variation in user-selectable exposure factors employed by different hospitals for each of a range of standard examination protocols carried out on the same model of multislice CT scanner; second to determine the effect of these variations on patient dose; and third to compare the range of patient doses with the relevant DRL values. By restricting this study to a single model of CT scanner, variations in patient dose owing to equipment design are removed. In addition each unit must meet the manufacturer's specifications for image quality, and this is verified independently during commissioning and routine surveys. Hence each CT scanner can achieve comparable image quality. This means that the variations in patient dose will be a direct result of the variation in user-selectable exposure factors.
| Method |
|---|
|
|
|---|
Dose measurements were made on one Toshiba Asteion Multislice CT using a 10 cm long CT pencil ionization chamber (Victoreen, USA) positioned free in air along the central axis of the scanner. Measurements were made for a range of different slice widths and exposure factors. The CT dose index, CTDI10cm,air [11] for each of these measurements was then calculated:
|
|
This CT chamber was also used to make dose measurements within a 16 cm head and 32 cm diameter body polymethyl methacrylate standard CT dosimetry phantom as specified by the Food and Drug Administration [12]. The weighted CT dose index, CTDIw [6] was then calculated for each phantom using typical exposure factors:
|
|
Using the results from the questionnaires and by applying scaling factors (determined from the CTDI10cm,air measurements) to the measurements made above, CTDIw values were calculated for all of the CT scans involved in the study.
It is anticipated that there will be variations in dose between the different CT scanners, resulting from differences in X-ray beam collimation, generator factors such as kVp, ripple and mAs calibration, and X-ray tube target angle. The manufacturers specify a ±20% margin for variations in CTDIw between scanners (personal communication: Toshiba, 2002), but it is understood that this includes an allowance for the positional errors resulting from variation in measurement technique. A more accurate method to compare tube outputs is the measurement of CTDIair, as this is measured at the isocentre and is less prone to systematic positioning errors and the effects of over-scan for example. Data from a national survey undertaken by ImPACT (St Georges Hospital, London) in 1996 give a variation in CTDIair of 7% between identical models, however this also includes "operator" dependent variables such as differences in equipment and chamber calibrations. Given that our results are based on CTDIair measurements the variation in CTDIair between multiple scanners of the same model is likely to be of the order of 5% (personal communication: ImPACT, 2002).
| Results |
|---|
|
|
|---|
|
|
|
For chest high resolution CT (HRCT) scan there was a factor of more than two between the highest and lowest CTDIw values (Figure 1c
). The choice of slice width varied from 0.5 mm to 1.0 mm between hospitals. For standard chest scans (non HRCT) (Figure 1d
), there was a smaller variation between the highest and lowest CTDIw values.
An even larger variation was observed for FESS scans (Figure 1e
), where there was a factor of seven between the highest and lowest CTDIw values.
For the neck soft-tissue scans (Figure 1f
) the CTDIw values varied little between six of the hospitals. At one hospital (G), however, the calculated CTDIw value was around a quarter of the mean value for all the hospitals.
Overall, hospital (E) had the lowest CTDIw values for four of the six examinations. The CTDIw values were 24% less than the average value for abdominal scans; 9% less overall for routine head scans; 47% less than the average for HRCT chest scans; 7% less than the average for non-HRCT chest scans; 64% less than the average for FESS and 3% less for neck soft-tissue scans.
| Discussion |
|---|
|
|
|---|
In all seven hospitals, the reference level for abdominal scans was easily achieved. This may imply that either the scanning parameters have been optimized, or that the DRL is now too high for modern scanners because it is based on CT data collected in the early 1990s when spiral scanning was in its infancy.
More interesting however is that only two hospitals adjust their exposure parameters according to patient size. Larger patients will introduce more photon scatter and attenuation, and if the scanning parameters are not altered the noise present in the images for large patients will be greater than that for average patients. Dose reductions of up to 45% can be achieved by adjusting exposure factors with patient size [14], which indicates that although the abdominal scans meet the DRL there still exists further room for optimization.
For chest scans (non HRCT) the CTDIw values followed a similar pattern to abdominal scans. All hospitals were lower than the 30 mGy DRL, but by comparing these values with the results obtained at centre E, there still exists scope in the other hospitals for dose reduction.
The CTDIw values for routine head scans were close to, or exceeded the DRL. For the cerebral region five hospitals achieved a CTDIw lower than the reference value, however the average value for all seven hospitals exceeded the DRL. For the posterior fossa region all hospitals exceeded the DRL. This is not unusual as other studies indicate that this is a difficult target to achieve [9]. Scans through this region of the brain utilize narrow slices (typically 2 mm). These high dose values on multislice CT scanners may well be due to engineering design and there may be little scope for further reduction. Such scanners have a lower geometric efficiency compared with single slice CT scanners. This is particularly apparent for narrow slice widths where the geometric efficiency can fall to below 50% [15]. Hence half of the radiation dose received by the patient does not contribute to the image.
Both the HRCT and FESS scans showed a large variation in CTDIw values between hospitals. It is interesting to note the similarities between these examinations. Both contain areas of the body that generate high contrast images, i.e. they contain air, tissue and bone, making good quality images relatively easy to achieve. In addition both scans involve the identification of relatively small structures within the body. This requires good spatial resolution, achieved through the selection of a narrow slice width (typically 12 mm).
The dose-limiting factor with such narrow slice selections is the noise present in the image. For narrow slice widths the number of X-ray photons received at the detector is reduced, compared with a wider slice width under similar conditions of exposure. In many cases the mAs per scan is increased to maintain the same signal to noise ratio in the image. However, where high contrast details are being visualized, the presence of increased noise may not be deleterious to the diagnosis. This is an area that requires further investigation to ensure that images are of sufficient quality to answer the diagnostic question but patient doses are kept as low as reasonably practicable.
The variation in CT dose found here is consistent with findings from other studies [810]. It is a requirement of IR(ME)R that "special attention" be paid to the optimization of high dose procedures and this requires a multidisciplinary approach. This paper has highlighted the significance of the choice of user-selectable exposure parameters, and also the dose savings that can be achieved through close collaboration between medical physicists, radiographers and radiologists.
| Conclusion |
|---|
|
|
|---|
For abdominal scans the DRL was achieved and the variation between hospitals was small. However, there still remains scope for further reduction of patient doses, particularly when due consideration is given to patient size while selecting exposure factors.
For HRCT and FESS examinations there was a considerable variation between hospitals, implying that there is substantial scope for dose optimization.
At one hospital, where dose optimization has been implemented, dose savings between 3% and 64% have been achieved whilst maintaining diagnostic image quality. This has only been achieved through close collaboration between radiographers, physicists and radiologists.
| Acknowledgments |
|---|
Received for publication October 22, 2002. Revision received May 30, 2003. Accepted for publication June 10, 2003.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. A. Jaffe, T. T. Yoshizumi, G. Toncheva, C. Anderson-Evans, C. Lowry, C. M. Miller, R. C. Nelson, and C. E. Ravin Radiation Dose for Body CT Protocols: Variability of Scanners at One Institution Am. J. Roentgenol., October 1, 2009; 193(4): 1141 - 1147. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.B. Nauer, A. Eichenberger, P. Dubach, J. Gralla, and M. Caversaccio CT Radiation Dose for Computer-Assisted Endoscopic Sinus Surgery: Dose Survey and Determination of Dose-Reduction Limits AJNR Am. J. Neuroradiol., March 1, 2009; 30(3): 617 - 622. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kubo, P.-J. P. Lin, W. Stiller, M. Takahashi, H.-U. Kauczor, Y. Ohno, and H. Hatabu Radiation Dose Reduction in Chest CT: A Review Am. J. Roentgenol., February 1, 2008; 190(2): 335 - 343. [Abstract] [Full Text] [PDF] |
||||
![]() |
A J Van Der Molen, W J H Veldkamp, and J Geleijns 16-slice CT: achievable effective doses of common protocols in comparison with recent CT dose surveys Br. J. Radiol., April 1, 2007; 80(952): 248 - 255. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Tsapaki, J. E. Aldrich, R. Sharma, M. A. Staniszewska, A. Krisanachinda, M. Rehani, A. Hufton, C. Triantopoulou, P. N. Maniatis, J. Papailiou, et al. Dose Reduction in CT while Maintaining Diagnostic Confidence: Diagnostic Reference Levels at Routine Head, Chest, and Abdominal CT--IAEA-coordinated Research Project Radiology, September 1, 2006; 240(3): 828 - 834. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |