BJR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

British Journal of Radiology (2003) 76, 464-472
© 2003 British Institute of Radiology
doi: 10.1259/bjr/82965103

This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gallacher, D J
Right arrow Articles by Saunders, J E
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gallacher, D J
Right arrow Articles by Saunders, J E

Full Paper

Use of a quality index in threshold contrast detail detection measurements in television fluoroscopy

D J Gallacher, MPhil1, A Mackenzie, MSc2, S Batchelor, MSc1, J Lynch, MSc3 and J E Saunders, BSc4

1 Medical Physics Department, Guy's and St Thomas' Hospital, London SE1 7EH, 2 King's Centre for Assessment of Radiological Equipment (KCARE), King's Hospital, London, 3 Medical Physics Department, Princess Margaret Hospital, Swindon and 4 Formerly Medical Physics Department, Guy's and St Thomas' Hospital Trust, London SE1 7EH, UK


    Abstract
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 
The use of a single index to assist in quality control procedures of X-ray television fluoroscopy systems was investigated. A single quality index was devised incorporating a measure of threshold contrast detail detectability (TCDD) performance and taking into account image intensifier input kerma rate, field size, differences in radiation beam quality, and pulsed fluoroscopy. This was applied to a number of clinical systems to investigate changes in image quality index quantified over time. Accepted measurement protocols were used to obtain these measurements. The results show system performance for different systems and can establish the decline in performance parameters over time or assess non-optimal image quality with clinical systems in field measurements. The systems studied were assessed with a variety of performance parameters including TCDD results, low contrast sensitivity, limiting resolution, and image intensifier input kerma rate under clinical modes of operation. The TCDD quality index, and dose normalized quality index, were found to be useful image quality assessment parameters for serial testing of systems, which augment the use of graphical methods for the display of TCDD curves.


    Introduction
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 
The use of contrast detail test objects in the assessment of imaging performance and quality assurance of television (TV) image intensifier fluoroscopy systems has been well established for many years [1]. By visually assessing a series of low contrast disks decreasing in contrast for a range of detail sizes, the contrast threshold versus detail diameter diagram can be plotted [2] for a given set of test conditions. For a given air kerma rate at the image intensifier (II) input plane, one might define a better performing system to have a comparatively lower contrast threshold of detection. Values of the visual threshold signal-to-noise ratio (SNR) at which a detail is perceptible have been reported in the literature: Chesters and Hay [3] estimated a value of between 2 and 3 in dynamic imaging, and Workman and Cowen [4] reported a value of 4 for computed radiography.

A series of such test objects were produced by the FAXIL group at the University of Leeds [5] and we will consider the use of the TO10 test object primarily designed for image intensifier TV fluoroscopy systems. An accepted protocol for the use of this test object in serial testing has been in widespread use over a long period [69] and is recommended in IPEM 77 [10].

Two contrast threshold curves obtained with the TO10 test object were published in report STB/7/82 [6]. One curve was obtained from measurements on new systems (we shall term STB1) and the other for systems of average age 5 years (we shall term STB2). Both curves were for fields of view (FOV) in the region of 23–25 cm and II input plane air kerma rates of 260 nGy·s-1. These two curves have been used for many years as standard reference data with which to compare image intensifier threshold contrast detail detectability (TCDD) performance. The measured II system data were normally visually compared with the STB/7/82 data on a log-log graph plot of contrast threshold versus detail diameter.

FAXIL published further data in the 1992 Test Object Users Manual [9] in which a curve for the 25 cm FOV was produced (we shall term F25). Also, curves relating to image intensifier FOVs in the ranges 30/36 cm (we shall term F33) and 15/17 cm (we shall term F16) were produced in the users manual [9]. The contrast threshold data in the 1992 Users Manual was converted to a new quantity however using the concept of a threshold detection index HT(A), which takes into account the detail area and is given in Equation 1Go. Furthermore, the threshold detection index is plotted against square root of detail area as opposed to detail diameter. Go


where A is detail area and CT is threshold contrast.

From Equation 1Go it can be seen that the smaller the contrast threshold CT, the greater the value of the threshold detection index, corresponding to a higher image quality score. For all three data curves (F16, F25 and F33) the II input kerma rate was not specified.

Launders et al [11] updated the viewing protocol for FAXIL TCDD test objects to include a variable distance protocol (VDP) measurement. This enables the observer to vary their viewing distance from the display monitor to optimize the image viewing conditions for both large and small test object details and hence tends to result in slightly higher image scores [11].

From the measurement data, TCDD curves are drawn and are useful in quantifying the image quality of the system. The TCDD curves are not intuitively simple and the production of a single quality image factor to describe image quality is useful. McRobbie et al [12] devised a single figure of merit F for image intensifier systems. Go


where R=measured limiting high contrast spatial resolution (lp·mm-1);

n=noise (using a Leeds N3 test object at one fixed detail size);

Kair=II input kerma rate (nGy·s-1); and

k=1.48 mm·nGy1/2·s-1/2.

The constant k was chosen to normalize F to unity for values n=3%, Kair=0.35 µGys-1, and R=1.2 line pairs per mm (lp·mm-1). These values were chosen to reflect a typical level of performance. The F index has been shown to be a useful measure, but may suffer from the following potential limitations. First, the figure of merit F is proportional to the limiting high contrast spatial resolution, which was measured without attenuation and with the system operating under automatic exposure factors. However, II input kerma rate and threshold contrast are measured under different conditions with a copper attenuator. Second, the limiting high contrast resolution is primarily designed to ensure that the system is optimally focused [9]. Defocusing is readily adjustable and the high contrast spatial resolution does not deteriorate significantly over time and so does not contribute to image quality factors. Clinically low contrast small details are more significant than high contrast small objects. The McRobbie quality factor only uses one detail size (1.1 cm diameter) and so cannot indicate a true image quality factor.

We believe that TCDD curves measured with the TO10 test object are a more useful basis for evaluating overall image quality. The height of the curve is closely related to the overall system SNR and provides a wide range of detail: the authors suggest our measure may be more sensitive to system performance over a range of detail sizes.

There are confounding factors in attempting to make a comparative index of performance. Problems with the use of subjective visual assessment test phantoms have been noted [13], and other performance measurements of a more objective nature such as the measurement of detective quantum efficiency (DQE) will remain the benchmarks for verifiable assessment of imaging systems. However, these types of measurement may not be practical on a routine basis. The availability of a field measurement protocol which aids serial testing and comparative analysis of systems would be a useful resource for the physicist performing both routine quality control (QC) and commissioning field tests.

As detailed in the update on the recommended viewing protocol by Launders et al [11], care does need to be exercised in making comparative measurements between systems because of differences in beam quality, II input kerma rate, II size and FOV selected for the measurement. Any image processing differences, e.g. noise reduction, will also need to be considered so as not to confound the validity of any such comparison.

In this paper we demonstrate a simple method for making a comparison between systems using the threshold contrast detail Leeds Test object TO10 using the parallel property of TCDD curves. Corrections of II input kerma rate, beam quality, field size and pulsed fluoroscopy can be incorporated.


    Theory
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 
Parallel property of curves
By parallel property we mean the curves can be scaled to show a good fit in terms of their overall shape, i.e. the relative heights at any given apparent detail size retain the same relative proportions.

F25, STB1 and STB2 curves were based upon measured data from several systems [6, 9] and therefore their overall shape is generally representative. All the curves follow a characteristic peaked shape (Figure 1Go). The use of scaling (normalization) factors was investigated to test if a single characteristic curve, with appropriate normalization factors could be used as a standard reference curve. Accordingly, STB1 and STB2 were scaled and best fitted to F25 by using the mean ratio of threshold detection index for all respective detail sizes to F25 to estimate the scaling factors. Figure 2Go shows the very close correlation between F25, STB1 and STB2, which is not surprising as all relate to a 25 cm image intensifier. This justifies the use of a single quality index using the parallel property.



View larger version (6K):
[in this window]
[in a new window]
 
Figure 1. Comparison of the standard threshold detection index plots relating to 25 cm field of view versus detail diameter. {triangleup}=F25, FAXIL Data (F25, 1992); x =STB1, STB/7/82 New Systems Data (DOH, 1982); +=STB2, STB/7/82 Systems of age 5 years (DOH, 1982).

 


View larger version (6K):
[in this window]
[in a new window]
 
Figure 2. Three 25 cm field of view curves best fit normalized to the FAXIL F25 (upper curve) versus detail diameter. {triangleup}=F25, FAXIL Data (F25, 1992); x =STB1, STB/7/82 New Systems Data (DOH, 1982); +=STB2, STB/7/82 Systems of age 5 years (DOH, 1982).

 
Definition of TCDD single quality index
The quality index, Q, is defined in Equation 3Go. The ratio of the measured detection index HT(Ai) for a given detail area Ai against the corresponding value for the F25 curve HTref(Ai) is taken (Table 1Go). The mean value of these ratios, summed over all available detectable data points, will be the quality index. Go


As the overall system performance is averaged over a range of detail areas the use of a single quality index can be seen to be a potentially useful indicator of system performance for the TCDD test. By definition a system with Q value of one would indicate the same level of performance as indicated by F25. A small error resulting from truncating the number of scoring details may result, but it should be noted that Q reflects the mean scoring over several details and this error will in general be less than 3%.


View this table:
[in this window]
[in a new window]
 
Table 1. Reference threshold contrast detail detectability values for the FAXIL Curves F16, F25 and F33

 
Reference curve
The reference curve (HTref(Ai)) could be any typical TCDD curve: we have chosen F25 [9]. This is the most recent data published by Leeds University and has been shown to be comparable with older curves. For the 1982 data [6] Q=0.78 for STB1 (new systems) and Q=0.61 for STB2 (average age 5 years).

From the literature [6, 9, 11] all these curves were based on a range of systems, with different image intensifiers, operating at different efficiencies and with different TV camera lag characteristics. All reference curve data were at 70 kVp and 1 mm copper beam attenuation.

Input kerma rate correction
The image quality of a system will depend on the II input kerma rate measured at the II entrance plane. As the dose rate is increased the quantum noise will be reduced, allowing easier visualization of low contrast details. In order to compare systems, or even the same system, operating at different II input kerma rates, the quality index Q needs to be modified by a dose normalizing factor. The dose normalization is based on the quantum noise which follows Poisson statistics (FAXIL [11], McRobbie [12]). The noise will therefore be inversely proportional to the square root of the number of detected photons, or their input rate [14]. The threshold detection index is itself proportional to the SNR of the system and hence modifying factors applied to this index must reflect the effect of quantum statistics on noise level [11]. Hence, the modifying factor applied to the threshold detection index values will be the square root of the ratio of the actual measured input kerma rate against a standard reference value of input kerma rate. Other sources of noise in the imaging chain will be present such as electronic noise or intensifier structural noise, however relative to quantum noise these are negligible in fluoroscopy [15].

The quality index will be scaled to that which would be obtained when running the system at a fixed reference value (Table 2Go). Dose normalized quality index Qn is first defined in Equation 4Go. Go


Kref is the reference input kerma rate (429 nGy·s-1) for the F25 curve. This value was obtained using Equation 4Go applied to the known Q value of the STB1 curve (0.78) at the stated input kerma rate of 261 nGy·s-1 for this data [6]. Km is the measured input kerma rate for a given system under consideration.


View this table:
[in this window]
[in a new window]
 
Table 2. Reference input air kerma values for use for different field of view (FOV) sizes

 
The method of choosing the reference exposure essentially normalizes the range of systems used in the F25 data. This is effectively the way McRobbie et al [12] adopted a k value of 1.48 in their figure of merit F to represent a typical level of system performance.

Further corrections are required owing to variations in inherent subject contrast of the test object with different tube potential (kVp) and beam attenuation (in mm of copper), and variation of energy absorption in the II input phosphor with different beam qualities must also be taken into account.

Correction of contrast differences in the test object with different beam qualities
In older protocols [1, 2, 5, 6] the specified test conditions were 70 kVp with 1 mm copper beam filtration for the TO10 test object. However, in the 4th edition of the Leeds test objects manual [9] threshold contrast values were published between 65 kVp and 80 kVp at 5 kVp intervals at 1 mm, 1.5 mm and 2 mm levels of copper beam filtration for all the rows and details sizes in the TO10 test object. Measurements may then be made on systems for which standard factors of 70 kVp with 1 mm copper cannot be obtained. The true threshold contrasts can be measured taking into account the differences in radiation beam quality. These corrected threshold contrasts are used to calculate normalized quality index Qn.

Correction of image intensifier input phosphor absorption efficiency with beam quality
The correction factor for input phosphor absorption is necessary because the ratio of mass energy absorption coefficients of the caesium iodide (CsI) II input phosphor to air is not constant over the range of beam qualities used in the Leeds TO10 test object measurements. Hence the energy absorption in the image intensifier input phosphor, for a given input air kerma at the intensifier face, is a function of radiation beam quality. It follows that this also affects the conversion efficiency of the image intensifier as a whole and so a correction should be applied to allow comparison with other systems and various dose level modes on a system.

Field measurements of systems must generally be carried out in automatic mode where the tester has no direct control over system tube potential and current. The fluoroscopy automatic brightness control (ABC) system will control operating factors to ensure a constant luminance from the output phosphor of the image intensifier by following a characteristic kVp/mA curve. Hence, for any given operating mode, the tester must add sufficient copper beam attenuation to attain a certain tube potential range (65–80 kVp) generally attempting to achieve close to 70 kVp. We therefore require a correction to be applied to convert test conditions (kVp and copper thickness) back to standard values of 70 kVp and 1 mm Cu used for the reference curves.

The output luminance of the II will be proportional to the rate at which energy is deposited in the input phosphor. In maintaining a constant luminance by feedback control the ABC effectively regulates a constant energy deposition rate in the input phosphor [16]. In practice only the II input kerma rate is easily measurable. The ratio of energy absorbed in the CsI input phosphor to measured input air kerma will vary depending upon the mean energy of the fluence spectrum [17]. Values of the ratio of mass energy absorption coefficients for CsI and air were calculated from published data [18] at different photon energies in the relevant range and are presented in Table 3Go. By using the Institute of Physics and Engineering in Medicine (IPEM) Report 78 X-ray spectrum processor code [19, 20] radiation beam qualities for the full range of test conditions were investigated. Mean spectral energy for each kVp and copper filter combination was calculated using the spectrum processor and values are presented in Table 4Go. Values of the ratio of mass energy coefficients from CsI to air, denoted R(V,X) where V is the tube potential and X the total thickness of copper added to the beam, were estimated from a best fit cubic from values in Table 3Go. In order to apply these results in a ready form a ratio Rn(V,X)=R(V,X)/R(70,1.0) has been calculated, normalized to the standard conditions (Table 4Go). The modified form of Equation [4] is shown later in Equation 5Go together with a further correction factor still to be discussed.


View this table:
[in this window]
[in a new window]
 
Table 3. Ratio of mass energy absorption coefficients for caesium iodide to air

 

View this table:
[in this window]
[in a new window]
 
Table 4. Ratio of mass energy R(V,X) absorption coefficients for caesium iodide to air for different beam qualities, and normalized beam quality factor Rn(V,X) to 70 kVp, 1 mm Cu

 
Reilly and Sutton in their ABC model [17] used only the II input kerma rate to characterize their fluoroscopy systems on the basis that when operating under conditions of beam hardening with 1 mm copper or more the mean energy fluence ratio of energy absorption coefficients for CsI/air was substantially constant. This was reported in other work [21]. We find that the correction required to be applied to the dose normalized quality index Qn is in most cases less than 5%, reaching 6.6% for the most extreme beam conditions of 80 kVp and 2 mm copper. It should be noted that the recommended calibrated measurement conditions of between 1 mm and 2 mm copper additional beam filtration will in general override the inherent filtration level in the tube assembly [9]. However, dose reduction filters in the system should be taken into account [22].

Quality index for different FOV sizes
The normalized quality index Qn can be extended to cover all FOVs. Launders [11] details a method to extend the use of the TO10 test object to different FOVs. In this approach the relative sizes of the FOVs are used to correct the apparent detail size between different FOVs and the TCDD curves are plotted on the same square root of detail area axis to enable comparison between performance on different FOV sizes. Appropriate kerma rate normalization factors are employed to modify the overall magnitude of the TCDD curves to account for different input kerma rates that relate to the different FOVs.

The FAXIL data for the 30/36 cm FOV (F33) and 15/17 cm FOV (F16) were used [9] (Table 1Go). Apparent detail size correction factors were calculated from the ratio of these sizes (33 cm and 16 cm, respectively, for the mid point of each range) to F25. These were then used to correct the square root of area (horizontal axis) ordinate points plotted for each respective curve F33 or F16 so that the apparent detail size will match F25. Curves F16, F25 and F33 are plotted in Figure 3Go, where the curves all follow the same general shape when corrected by the ratio in apparent detail size. The reference exposure for F16 and F33 can be estimated using the parallel property of these curves to F25. A minimum root mean square (RMS) difference best fit was used to calculate the normalization factors for best fitting the curves. The multiplying factors applied to each curve to obtain the best fit were 1.30 for F33 and 0.78 for F16. These are presented in Figure 4Go, where the curves show good correlation, justifying that all quality index data can relate back to a single reference curve F25. From these normalization values, the values of Kref relating to each respective FOV were calculated from Equation 4Go and are shown in Table 2Go.



View larger version (10K):
[in this window]
[in a new window]
 
Figure 3. FAXIL data curves for 30/36 cm, 25 cm, 15/17 cm field of view versus apparent detail size. {square}=F33, FAXIL Data (F33, 1992); {triangleup}=F25, FAXIL Data (F25, 1992); {circ}=F16, FAXIL Data (F16, 1992).

 


View larger version (8K):
[in this window]
[in a new window]
 
Figure 4. Best root mean square difference fit normalized curves for 30/36 cm, 25 cm and 15/17 cm versus apparent detail size. {square}=Normalized F33 to F25; {triangleup}=F25, FAXIL Data (F25, 1992); {circ}=Normalized F16 to F25.

 
To calculate Qn for a given FOV and measured input kerma rate Km choose the appropriate reference curve closest in field size to the system under measurement: Qn is calculated with the value of Kref appropriate for that reference curve. In this way different FOVs and input kerma rates can be compared via a single index Qn.

Extending the quality index to cover pulsed fluoroscopy systems
When pulsed fluoroscopy is used the temporal averaging effect by the observer is reduced. To compensate for this some manufacturers may increase the dose per pulse to reduce the noise level. The overall effect may nevertheless be dose reduction. Aufrichtig et al [23] investigated the effect of pulse rate and dose on perceptual quality of fluoroscopic images. They found that dose reductions were possible for a matched image score (equivalent perception dose) in threshold detection measurements. Mean dose savings of 22%, 38% and 49% for fluoroscopy at 15, 10 and 7.5 pulses per second (pps), respectively, were reported. A pulse rate of 30 pps was treated as continuous by Aufrichtig on the basis that perceptual differences at higher pulse rates than this were small owing to limitation in the human visual system [23]. It is proposed to make allowance for pulsed fluoroscopy modes by extending the definition of the quality index. The approach taken here will be to introduce an equivalent perception dose factor {Lambda}(f) calculated from Aufrichtig's results, where f is the pulse frequency in pps (Table 5Go). Under this definition a system operating at a pulsed fluoroscopy frequency f would need to be operating at a fraction {Lambda}(f) of the input kerma rate in order to show an equal normalized quality factor Qn compared with the continuous fluoroscopy mode of operation. This is justified as the equivalent perception dose at a given pulsed sampling frequency in fluoroscopy should give rise to the same value for normalized quality index Qn when comparing different pulsed frequencies on the same system.


View this table:
[in this window]
[in a new window]
 
Table 5. Modifying factors for pulsed fluoroscopy ({Lambda}(f))

 
Normalized quality index
The quality factor defined in Equation 3Go is useful for serial testing, i.e. repeat testing on the same equipment under specified conditions. The normalized quality index Qn defined in Equation 5Go must be used to compare systems operating on different FOVs, under different II input kerma rates, at different pulsed fluoroscopy rates and under different beam test conditions. Go


where Km=measured input kerma rate of the system in nGy·s-1;

Kref=reference input kerma rate of the system in nGy·s-1;

Rn(V,X)=radiation beam quality normalized factor and is a function of V (tube potential) and X (beam attenuation in mm copper); and

{Lambda}(f)=pulsed fluoroscopy correction factor.


    Methods
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 
The quality index has been used at Guy's and St Thomas' Hospital for 8 years as part of the annual QC program. Measurements were carried out independently under single field test conditions by two experienced observers, one of whom had 10 years experience (Observer 1) and the other 3 years (Observer 2), in using the TO10 test object.

The results were reviewed for four hospital fluoroscopy systems A, B, C and D during use of the quality index as part of a QC programme.

The Leeds TO10 contrast detail test object was placed on the cover plate of the image intensifier. Sufficient copper was placed in the beam until a kVp in the range 65 kV to 80 kV was reached. The air kerma rate was measured using an MDH 9010 ion chamber and electrometer system. In accordance with the recommended protocols in the instruction book [9] the data were obtained using a fixed distance viewing protocol (FDP) of approximately four times the diameter of the blanking circle from the system display monitor. Previous values were only examined after the tests had been completed so as not to influence their impression of the image quality of the system. Where possible, both observers were used. The value of Observer 1 was involved in all the measurements.

It is recognized that the presence of measurement errors due to within observer, between observer and between sample variance will be present in these measurements [24, 25]. No attempt has been made to quantify these in this study, hence reflecting field practice measurements and not repeated tests under laboratory conditions. However, regular blind testing comparisons were made to verify overall image score consistency between the observers in this study. It is recognized that the variations in threshold contrasts will be of the order of 16% for single observer and 11% for two observer measurements [25].

Some measurements were carried out using the VDP to investigate this effect on the image quality of systems C and D.


    Results
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 
Table 6Go shows results from system A, a general purpose over couch tube fluoroscopy system with a 27 cm FOV II, over 5 years (the system was installed in 1989) with results from Observer 1. It can be seen that the Q index falls slightly over time as the system ages. II input kerma rates were increased over time to compensate for conversion loss, increased camera noise and monitor ageing to maintain image quality. The Qn values show a more significant decline, which reflects the increasing input kerma rate at the intensifier with this system. No major TV chain components, e.g. image intensifier, TV camera or display monitor, were changed during the sample period.


View this table:
[in this window]
[in a new window]
 
Table 6. Measured quality index values on system A (27 cm field of view (FOV) image intensifier (II))

 
On this system the Qn index has been a useful indicator of declining performance in serial testing. It should be noted that this system started from a relatively low input kerma rate requirement at the start of the monitoring period. The final input kerma rates were within accepted tolerance ranges from the manufacturer's specification [7] even by the end of the monitoring period. Replacement of TV camera and monitor were recommended on this system as a result of the QC programme. Changes in TCDD test performance may have been more difficult to register using a solely graphical approach to TCDD values, and the application of this technique may have positive implications in the management of system upgrades or replacement programme within a radiology department.

Table 7Go shows results for system B, a new large format II system with pulsed fluoroscopy modes. Even with the large differences in input kerma rate in the different modes of operation the Qn values reflect satisfactory imaging performance according to these indicators. This is also true with the pulsed fluoroscopy modes. It should be noted that these pulsed modes of operation result in a loss of temporally changing information, because of the lower rates of sampling, and this is not reflected in a static test object test. Also, the presence of noise reduction in the digital image processing algorithms will be expected to have an effect on image quality. The presence of online digital image processing and pulsed fluoroscopic acquisition of images may be relevant factors for consideration on data from systems of recent manufacture.


View this table:
[in this window]
[in a new window]
 
Table 7. Measured quality index values on system B (40 cm field of view (FOV) image intensifier (II))

 
Variable distance protocol
The use of VDP and FDP were investigated on system C, a large format C-arm vascular II system operating under automatic exposure control mode in different FOVs. The measurement data for FDP was used to calculate Q and Qn for all FOVs tested and the results are presented in Table 8Go. For comparison, VDP data have been used to calculate quality index values although these are based on FDP curves.


View this table:
[in this window]
[in a new window]
 
Table 8. Measured quality index values on system C (38 cm field of view (FOV) image intensifier (II))

 
From Table 8Go it can be seen that the system has shown consistently satisfactory performance over the 3 sample survey years. It should be noted that consistent values for Qn were obtained across all three FOVs tested. The VDP gave a greater value of Qn due to enhanced detail scores compared with FDP. This is because at shorter viewing distances visualization of small details is improved, and at larger viewing distances the large detail score tends to be enhanced.

Further measurements of Qn using the FDP were carried out on system D, a general purpose C-arm fluoroscopy system, and are presented in Table 9Go. These data again show satisfactory performance over the sample years, and a general trend for the VDP measurements to score greater Q index values than the FDP counterparts.


View this table:
[in this window]
[in a new window]
 
Table 9. Measured quality index values on system D (38 cm field of view (FOV) image intensifier (II))

 

    Discussion
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 
We have presented a method for creating a single quality index Qn (Equation 5Go), where a value of 1 is equivalent to the image quality of reference curve F25. This value takes into account II input kerma rate, beam quality, FOV size and pulsed fluoroscopy. The quality index has been used successfully in Guy's and St.Thomas' Hospital for many years for serial testing of equipment and comparing systems during equipment evaluation for purchase.

From many years of experience, image assessment criteria are proposed as an aid to the classification of equipment performance levels in Table 10Go. These may subsequently be reviewed on the basis of a larger sample data set, perhaps with appropriate modifications for pulsed fluoroscopy and operation of digital noise reduction image processing.


View this table:
[in this window]
[in a new window]
 
Table 10. Proposed fluoroscopy threshold contrast detail detectability image criteria levels

 
It should be noted that the Qn index should not be tested in isolation from other system parameters. In particular the limiting resolution measured with the Huttner Type 18 test object must be verified to ensure that system focusing is satisfactory. Tests with other test objects including TO.M1 and TO.N3 were also carried out to verify other aspects of system performance.

While it was necessary to estimate air kerma rates for the F16, F25 and F33 curves, this does not affect the relationship between the curves. Future work may nevertheless produce a new curve and associated II input kerma rate reference level with which to compare fluoroscopy systems.

As recommended in the update on viewing protocol [11] the same approach could be applied to data obtained with VDP after establishing reference curves based on a VDP approach as the shape of the VDP curves may be expected to differ from FDP. It is important to specify when plotting the data obtained and calculating the quality index whether these relate to the FDP or VDP. The production of curves using VDP will be the subject of further work.

The standardization of observer scoring levels between centres and between individual observers remains a potentially problematic area, but this could perhaps be assisted by the introduction of standard TCDD hard copy images with which individuals working within the field could assess their own scores against a standard reference level. At a local level consistency should be verified between observers.

The presence of different digital image processing algorithms, such as noise reduction, should also be taken into account when making any comparative assessment of systems. Systems with high lag, or with digital frame averaging noise reduction techniques will score a higher Qn value. However, these images could suffer from movement artefacts.

Dynamic imaging characteristics may be subject to other tests but it is not intended at this stage to take these into account within the dose normalized quality index Qn.


    Conclusions
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 
The use of the dose normalized quality index Qn for serial assessment of image intensifiers in conjunction with other tests has been shown to be a useful indicator of image quality in fluoroscopic modes in an equipment quality control programme, using standard fixed viewing distance protocols.

Making comparisons between systems is inherently difficult as there are many factors that would need to be taken into account and not all of them are known. The quality index Qn is a useful comparative indicator which assesses the relative efficiency taking into account differences in II input kerma rate, beam quality and fluoroscopic pulse sampling rate. However, this number cannot be used in isolation and factors such as lag, distortion and high contrast resolution must also be considered.

The use of the quality index approach will facilitate more effective operation of quality control programmes and assessment of criteria for replacement or upgrade of systems on a more objective basis than the comparison of TCDD curve data alone. The system will enable clearer communication with the radiology department. The proposal that most modern systems in good adjustment would be expected to return normalized quality index values Qn of greater than 1.0 has been supported by results obtained in this study.

Received for publication February 22, 2001. Revision received March 18, 2003. Accepted for publication April 10, 2003.


    References
 Top
 Abstract
 Introduction
 Theory
 Methods
 Results
 Discussion
 Conclusions
 References
 

  1. Hay GA, Clarke OF, Coleman NJ, Cowen AR. A set of X-ray test objects for quality control in television fluoroscopy. Br J Radiol 1985;58:335–4.[Abstract]
  2. HPA. 1981, Measurements of the performance characteristics of diagnostic x-ray systems used in medicine, Part II, X-ray Image Intensifier Systems, TGR32, Part 2 (Hospital Physicists Association, London).
  3. Chesters MS, Hay GA. Quantitative relation between detectability and noise power. Phys Med Biol 1983;28:1113–25.[CrossRef]
  4. Workman A, Cowen AR. Signal, noise and SNR transfer properties of computed radiography. Phys Med Biol 1993;38:1789–808.[CrossRef]
  5. Hay GA, Clarke OF, Coleman NJ, Cowen AR, Craven DM. Instructions for the use of the Leeds test objects (supplied with each set of test objects). 1979.
  6. DHSS. The Testing of X-Ray Image Intensifier Systems (Working Group Report STB/7/82), London: Department of Health and Social Security, 1982.
  7. Medical Devices Agency. The Testing of X-Ray Image Intensifier Television Systems (Working Group Report), Evaluation Report MDA/95/07. London: Medical Devices Directorate, Department of Health and Social Security, 1995.
  8. Medical Devices Agency. The Testing of X-Ray Image Intensifier Television Systems, Evaluation Report MDA/98/68. London: Medical Devices Agency, Department of Health, 1998.
  9. Cowen AR, Clarke OF, Coleman NJ, et al. Leeds X-ray Test Objects Instruction Manual (4th edn), Leeds: FAXIL, The General Infirmary, 1992.
  10. IPEM/CoR/NRPB. Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Imaging Systems, IPEM Report 77. York: Institute of Physics and Engineering in Medicine, 1997.
  11. Launders JH, McArdle S, Workman A, Cowen AR. Update on the recommended viewing protocol for FAXIL threshold contract detail detectability test objects used in television fluoroscopy. Br J Radiol 1995;68:70–7.[Abstract]
  12. McRobbie DW, Hancock AP, Castellano IA. A figure of merit for the assessment of image intensifier systems. Br J Radiol 1992;65:878–84.[Abstract]
  13. International Commission on Radiation Units and Measurements (ICRU). Report 54, Medical Imaging - The Assessment of Image Quality, Bethesda, Maryland, USA ICRU, 1995.
  14. Kotre CJ, Marshall NW, Faulkner K. An alternative approach to contrast-detail testing of X-ray image intensifier systems. Br J Radiol 1992;65:686–90.[Abstract]
  15. Marshall NW, Kotre CJ, Robson KJ, Lecomber AR. Receptor dose in digital fluorography: a comparison between theory and practice. Phys Med Biol 2001;46:1283–96.[CrossRef][Medline]
  16. Dowsett DJ, Kenny PA, Johnston RE. The physics of diagnostic imaging. London: Chapman & Hall Medical, 1998:228–53.
  17. Reilly AJ, Sutton DG. A computer model of an image intensifier system working under automatic brightness control. Br J Radiol 2001;74:938–48.[Abstract/Free Full Text]
  18. Hubbell JH, Seltzer SM. Tables of X-ray mass attenuation coefficients and mass-energy absorption coefficients 1 keV to 20 MeV for elements Z=1 to 92 and 48 additional substances of dosimetric interest. NISTIR 5632. Gaithersburg, MD: National Institute of Standards and Technology, 1995.
  19. Sutton DG, Reilly AJ. Report 78 Spectrum Processor. York: IPEM, 1997.
  20. Cranley K, Gilmore BJ, Fogarty GWA, Desponds L, Sutton DG. Catalogue of diagnostic X-ray spectra & other data, IPEM Report 78 (CD-ROM). York: IPEM, 1997.
  21. Boone JM, Pfeiffer DE, Strauss KJ, Rossi RP, Lin PP, Shepard JS, et al. A survey of fluoroscopic exposure rates: AAPM task group No.11 Report. Med Phys 1993;20:789–94.[CrossRef][Medline]
  22. Nicholson RA, Thornton A, Akpan M. Radiation dose reduction in paediatric fluoroscopy using added filtration. Br J Radiol 1995;68:296–300.[Abstract]
  23. Aufrichtig R, Xue P, Thomas CW, Gilmore GC, Wilson DL. Perceptual comparison of pulsed and continuous fluoroscopy. Med Phys 1994;21:245–56.[CrossRef][Medline]
  24. Swets JA, Pickett RM. Evaluation of Diagnostic Systems: Methods from Signal Detection Theory. New York: Academic Press 1982;80–93.
  25. Marshall NW, Faulkner K, Kotre CJ, Robson K. Analysis of variations in contrast-detail measurements performed on image intensifier systems. Phys Med Biol 1992;37:2297–302.[CrossRef]



This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
C. Walsh, D. Gorman, P. Byrne, A. Larkin, A. Dowling, and J. F. Malone
QUALITY ASSURANCE OF COMPUTED AND DIGITAL RADIOGRAPHY SYSTEMS
Radiat Prot Dosimetry, March 4, 2008; (2008) ncn047v1.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
K. Kepler and A. Vladimirov
Optimisation strategies introduced for CR at health care centres in Estonia
Radiat Prot Dosimetry, March 1, 2008; 129(1-3): 127 - 131.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
I D Honey, A Mackenzie, and D S Evans
Investigation of optimum energies for chest imaging using film-screen and computed radiography
Br. J. Radiol., May 1, 2005; 78(929): 422 - 427.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
D S Evans, A Mackenzie, C P Lawinski, and D Smith
Threshold contrast detail detectability curves for fluoroscopy and digital acquisition using modern image intensifier systems
Br. J. Radiol., September 1, 2004; 77(921): 751 - 758.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gallacher, D J
Right arrow Articles by Saunders, J E
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gallacher, D J
Right arrow Articles by Saunders, J E


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS