| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Full Paper |
Department of Clinical Physics & Bioengineering, Health Physics Division, Lower Ground Floor, Divisional Offices (west), Western Infirmary, Glasgow G11 6NT, UK
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The purpose of this study was to determine the magnitude and distribution of doses to the legs of radiologists when carrying out interventional procedures, to compare these with doses to the hands and evaluate the effectiveness of different types of shielding in reducing these doses. The results have been used to assess when doses to the legs might approach a dose limit. The doses across the lower limbs and to the hands were measured using themoluminescent dosimeters (TLDs). Magnitudes of leg doses have been related to the dosearea products (DAPs) which quantify the amount of radiation used in each examination. In addition, doses to the legs of other personnel within the room have been investigated. Measurements of direct scatter air kerma were also made during simulated patient examinations. These results have been used in conjunction with clinical data in the interpretation of results.
| Methodology |
|---|
|
|
|---|
|
The distribution of scatter air kerma around the couch of a Philips Integris V3000 (Philips Medical Systems, Andover, MA) undercouch C-arm interventional suite (A in Table 1
) was measured using a RANDO phantom to simulate a patient. The phantom was used in positions routinely encountered in clinical practice and the experimental conditions, such as phantom position on table, table height and sourceintensifier distance (SID), were set as close as possible to those employed during clinical procedures.
The scatter air kerma was measured using a Radcal 9010 radiation monitor, with a 180 cc chamber. Measurements were made in both the vertical and horizontal planes with the aid of grids of 100 mm squares marked on 2 m x 2 m polythene sheets to assist in positioning. The sheets were held in position in either plane using clamp stands located at either end side of the table with sections cut away to take the X-ray unit and phantom. The RANDO phantom was then irradiated under fluoroscopy using a 100 mm x 100 mm field size in the midline from the level of T12 to L4 vertebrae, using an image intensifier field size of 38 cm and a general abdominal setting. Under automatic exposure rate control, this exposed the phantom to a beam of quality 70 kV and 3 mA with an entrance surface dose rate of 25 mGy min-1 and a DAP rate of 2.9 Gy cm2 min-1. The scatter air kerma was measured with the 180 cc chamber clamped at the grid intersections at 100 mm intervals in areas of higher dose rate, and at 200 mm intervals at positions further from the X-ray beam. Over 200 individual measurements were made around the table for each plane. The data were used to make isodose contour plots, to aid interpretation of the TLD results.
| Results |
|---|
|
|
|---|
|
|
During transjugular intrahepatic portosystemic shunts (TIPS) the doses to the legs were two to three times greater than those to the hands. This was despite an integral lead screen for protection of the legs being used in hospital E, but not in hospital A. The mean dose to the radiologists' legs in hospital A was 2.61±0.50 mSv per procedure, whilst that in hospital E was 0.50±0.36 mSv per procedure.
During stenting, embolisation and angioplasty procedures the mean dose to the legs ranged from 0.03 mSv to 0.97 mSv, with the highest doses being for embolisations. In hospital A, where no lead protection for the lower limb was available, the leg dose was in most cases greater than that to the hand, and for some types of procedure the lower limb dose was as much as three times the dose to the hand. In hospital B, where lead protection which was integral to the table was used, the leg dose was significantly lower. This was also seen in hospital C and F where a mobile screen was used consistently during biliary and stenting procedures, respectively. In hospital C however, during the embolisation procedure the mobile lead screen was not used, resulting in the lower limb receiving a far higher dose than expected.
The relationship between the DAP and dose to the foot of the most exposed limb was also investigated. There was a clear linear relationship between the dose to the most exposed foot and DAP reading per procedure when no lead protection was used (Figure 2
) (r=0.96), but there was little correlation between DAP and screening time (r=0.39). Procedures that resulted in a DAP of approximately 100 Gy cm2 gave a dose of 0.9 mSv to the leg nearest to the X-ray field (usually left) and 0.7 mSv to the other leg.
|
Figures 3 and 4![]()
show plots of the scatter air kerma around the RANDO phantom irradiated in the posteroanterior (PA) projection in the vertical plane at right angles. A maximum air kerma rate between 300 and 600 µGy min-1 was measured along the side of the phantom, and immediately below the level of the table adjacent to the X-ray field. The typical positions adopted by staff around the couch are illustrated in Figure 5
. In general the radiologist stands to the right of the C-arm. For the particular X-ray unit, phantom and field used for the measurements, the radiologists' legs would be exposed to a scatter air kerma rate of 30 µGy min-1 to 150 µGy min-1. In the area where scrubbed nursing staff or secondary radiologists may stand at around 1 m away from the X-ray tube, the air kerma rate falls to between 15 µGy min-1 and 30 µGy min-1. At the bottom of the table where radiography staff were usually positioned, the scatter air kerma rate dropped to 10 µGy min-1 and below, and in areas where other staff might be present, it was below 10 µGy min-1.
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
During biliary procedures the doses to the legs are in most cases lower than those to the hands, irrespective of whether lead protection is employed. There is greater potential for the radiologists' hands to receive higher doses in biliary procedures, because they need to be close to the patient's side when screening in order to manipulate the catheter effectively, and therefore in an area of relatively high scatter dose rate (Figure 3
). However, the procedures do not generally have high DAPs and as a result the doses to the legs are lower.
The mean leg dose during TIPS procedures varied markedly between the two hospitals which performed this procedure, with those for hospital A being the highest of any procedures monitored. The doses for TIPS procedures are likely to be high, as they are commonly the longest procedures undertaken and require the use of several different projections. The differences between the two centres could be attributed to several factors, the most important of which was the dose performance of the X-ray units, but also included the experience of the radiologists in performing TIPS procedures and the techniques used. The doses to the legs in both hospitals were higher than those to the hands. TIPS procedures involve the radiologist inserting a catheter down the internal jugular vein. Once the catheter is inserted the radiologist will screen as the catheter advances through the superior vena cava, inferior vena cava and down to the right hepatic vein. It is from here that an artificial channel will be created from the hepatic vein to the portal vein, thus shunting blood away from liver sinusoids and reducing portal venous pressure. The hands do not need to be close to the X-ray field, as most of the screening will be performed over the hepatic area, while the hands remain up at the neck manipulating the catheter end. Thus the legs are only slightly further from the X-ray beam than the hands. The reason that doses to the legs are higher than those to the hands, is that the legs are exposed to radiation scattered directly from the lower surface of the couch, whereas scattered radiation to which the hands are exposed is attenuated by passage through the body.
It is perhaps surprising that the leg doses for TIPS were higher than those to the hands in hospital E, where an integral type lead screen had been provided. Measurements had shown that for other procedures the introduction of a lead screen had a significant effect on dose to the lower limb, in many cases reducing the leg dose to the minimum detectable level for the TLDs used. The difference with TIPS was that the lead screen, which was fixed to the side of the table, did not provide protection to the top of the table, which is the position where the radiologist stands to insert the catheter in the neck (X in Figure 5
).
The measured doses to the lower limbs during stenting, embolisation and angioplasty procedures ranged from 0.02 mSv to 0.97 mSv. The corresponding doses to the hands were between 0.04 mSv and 0.59 mSv. In general (excluding biliary procedures) where no lead protection was available (hospital A) the leg doses were higher than those to the hands by a factor of 23. If lead protection was available (hospitals B, C and F) the dose to the legs was significantly reduced. In the case of embolisation procedures at hospital C however, the legs received a higher dose than that to the hands because the mobile shield was not used.
The results of this study have shown that lead screens provide an effective method of protecting the lower limbs during fluoroscopy in most cases. There are two basic types of screen, ones that are integral to the table and screens that are freely mobile. Both have inherent advantages and disadvantages. The advantage of the screen that is integral to the table is that it is always in place and no conscious decision is needed to use it. As it is integral to the table, the risk of collision with other parts of the unit is minimal, as the screen will rise and fall with the table as it is moved. This is of particular relevance as newer interventional units also permit cranial and caudal tilting. The main disadvantage of this system in most cases is that it has a restricted range of horizontal movement. The results for TIPS procedures highlight this difficulty. Usually the lead screen is attached to a limited area of the table via studs or is permanently attached. This means that radiologists standing side-on to the patient will be adequately protected. However, when the radiologist stands at the top of the table, as in the case of TIPS procedures, the lead screen provides no protection at all. Therefore the legs, which remain relatively static, receive a high dose.
This particular problem is alleviated by use of a mobile lead screen. Such a screen can be placed in the most appropriate position to protect the radiologist's legs wherever he/she stands. There are, however, problems associated with this type of screen. Firstly there needs to be a conscious decision to use it. The lead screen has to be put into place before the procedure begins, as all procedures require a sterile environment. There is also a greater risk of collision with other pieces of equipment. Of particular concern is collision with the screen when tilting the table, or moving the table up or down. Thus in general for protection where the radiologist stands at the side of the unit, e.g. biliary, stents, embolisations and angioplasties, the integral screen provides the best option, but consideration should be given to the types of procedure being performed when a unit is being purchased and appropriate protection specified.
There was a clear linear relationship between the DAP reading and the dose to the feet of the radiologists (Figure 2
). From the data in Table 1
, a "rule of thumb" was established to provide guidance on the approximate magnitude of doses to the legs. This can be used to determine whether protection might be required. The rule is that a DAP of 100 Gy cm2 will give a dose of approximately 1 mSv to the legs, if no shielding is present. If lead protection is available, this dose would drop to approximately 0.02 mSv.
Doses for nursing and radiography staff assisting in interventional procedures were significantly lower than those received by radiologists and should never approach any dose limit for the extremities. However, the education of staff about the location of high scatter areas within the room allows the staff to optimise their dose personally in accordance with the ALARP principle.
| Conclusions |
|---|
|
|
|---|
The study has demonstrated that the inclusion of a lead screen to protect the lower limbs is an effective method of dose reduction when performing interventional procedures. It has highlighted the importance not only of using lead screens, but also of good screen design, in order to alleviate the problems associated with protecting the radiologist when performing procedures that require him/her to stand at different positions in relation to the X-ray unit and the couch. The inclusion of lead screens when purchasing new interventional suites is recommended as it provides an excellent and cost effective method in the optimization of radiologist extremity doses. In procedures such as TIPS, where the radiologist stands at the head of the couch, a mobile screen, which is placed in position prior to commencement of the procedure, may provide the most effective method for restricting doses to the legs. However, for most other procedures, a lead screen integral to the table is likely to be the better option.
| Acknowledgments |
|---|
Received for publication April 24, 2002. Revision received November 13, 2002. Accepted for publication February 6, 2003.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. J. Martin A REVIEW OF RADIOLOGY STAFF DOSES AND DOSE MONITORING REQUIREMENTS Radiat Prot Dosimetry, September 16, 2009; (2009) ncp168v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
V TSAPAKI, P N MANIATIS, A MAGGINAS, V VOUDRIS, S PATSILINAKOS, T VRANZTA, E VANO, and D S COKKINOS What are the clinical and technical factors that influence the kerma-area product in percutaneous coronary intervention? Br. J. Radiol., December 1, 2008; 81(972): 940 - 945. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Vanhavere, E. Carinou, L. Donadille, M. Ginjaume, J. Jankowski, A. Rimpler, and M. Sans Merce AN OVERVIEW ON EXTREMITY DOSIMETRY IN MEDICAL APPLICATIONS Radiat Prot Dosimetry, April 29, 2008; (2008) ncn149v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Tsapaki, S. Patsilinakos, V. Voudris, A. Magginas, S. Pavlidis, T. Maounis, G. Theodorakis, M. Koutelou, T. Vrantza, M. Nearchou, et al. Level of patient and operator dose in the largest cardiac centre in Greece Radiat Prot Dosimetry, March 1, 2008; 129(1-3): 71 - 73. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Schueler, T. J. Vrieze, H. Bjarnason, and A. W. Stanson An Investigation of Operator Exposure in Interventional Radiology RadioGraphics, September 1, 2006; 26(5): 1533 - 1541. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Vano, L. Gonzalez, J. M. Fernandez, C. Prieto, and E. Guibelalde Influence of patient thickness and operation modes on occupational and patient radiation doses in interventional cardiology Radiat Prot Dosimetry, June 1, 2006; 118(3): 325 - 330. [Abstract] [Full Text] [PDF] |
||||
![]() |
N Hidajat, P Wust, M Kreuschner, R Felix, and R-J Schroder Radiation risks for the radiologist performing transjugular intrahepatic portosystemic shunt (TIPS). Br. J. Radiol., June 1, 2006; 79(942): 483 - 486. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Goni, D. Papadopoulou, Em. Yakoumakis, N. Stratigis, J. Benos, V. Siriopoulou, Tr. Makri, and Ev. Georgiou Investigation of occupational radiation exposure during interventional cardiac catheterisations performed via radial artery Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 107 - 110. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Whitby and C J Martin A study of the distribution of dose across the hands of interventional radiologists and cardiologists Br. J. Radiol., March 1, 2005; 78(927): 219 - 229. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |