| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Short communication |
Department of Health Physics, Azienda Ospedaliera San Giovanni Battista di Torino, c.so Bramante 88, 10126 Torino, Italy
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Since orthopantomography is often performed in paediatric cases and involves a large population, a reference dose appears to be worthwhile. Recommended image quality and reference dose criteria are provided in document EUR 16260 [1] for several standard radiographic procedures. In the case of examinations that include projections not described in document EUR 16260, such as orthopantomography, there is no consensus on methodology for dosimetry.
Several methods have been proposed for assessing patient dose in panoramic dental radiography [25] and different values of DRL have also been proposed [3, 6].
As part of the National Radiological Protection Board's (NRPB) dental X-ray protection services, Napier [2] reported dosewidth product (DWP) as a useful tool for panoramic dental radiographic reference dose determination. In fact, DWP is well correlated with the dosearea product (DAP), and DAP can be converted to effective dose using the conversion factor estimated by Williams and Montgomery [4]. The NRPB assessment of panoramic X-ray sets measures the absorbed dose (to air) at the front side of the secondary collimator, integrated over a standard adult exposure cycle. Again, this quantity is used for its ease of measurement and does not require a patient to be present during a measurement. The dose per exposure cycle is then multiplied by the horizontal width of the beam at the receiving slot. The NRPB recommends the adoption of a reference DWP of 65 mGy mm for a standard adult panoramic radiograph. This reference value can be used as a guide to accepted clinical practice.
The proposed methods that use film and/or thermoluminescent dosemeters (TLDs) to assess DWP suffer from several disadvantages, such as set-up complexity and delay from dose measurement to the reading of TLDs or film, while DRL should be assessed by an easily measurable quantity [7].
The aims of this study were to develop an easy method for assessing DWP using a pencil ionization chamber, to compare the proposed method with that using TLDs and to assess the accuracy and reproducibility of the method.
| Methods and materials |
|---|
|
|
|---|
The pencil chamber is usually used to measure the CT dose index (CTDI), which is designed to represent the dose per slice received by a patient in a multislice study. If the calibration factor is given in terms of mGy C-1, CTDI is defined as
|
|
where T is the nominal slice thickness, Da(z) is the dose profile, M is the instrument reading, F is the calibration factor to convert the instrument reading to absorbed dose, L is the chamber length (as specified by the manufacturer), and ftp is the factor to correct for any difference in temperature and pressure at the time of measurement from those prevailing when the instrument was calibrated. The limits of integration are, theoretically, from -
to +
, but in practice the limits can be taken to be zi±
zi, where zi is the centre point of the dose profile and
zi is a cut-off beyond which the contribution can be considered to be negligible. The value
|
|
Measurements were performed on five panoramic dental X-ray units, listed in Table 1
, using a medium adult setting. Measurement of dose at the secondary collimator on dental X-ray units was made using a pencil ionization chamber (chamber PC-4P; Capintec, Pittsburgh, PA and dosemeter; UNIDOS, Freiburg, Germany). In two cases the chamber was mounted in an air-equivalent jig (Figure 1
) and in three cases the chamber was directly mounted at the secondary collimator. The ionization chamber was perpendicular to the secondary collimator. Before each measurement, test rotations were made to ensure that there was sufficient space between the jig and the headrest whilst the secondary collimator was rotated. Measurements were repeated on a few sets to test the reproducibility of results. The ionization chamber was calibrated by a calibration centre of the Italian National Calibration system.
|
|
|
The film was exposed at the secondary collimator, digitized (ARCUS II; AGFA, Livingstone, UK) and the FWHM of the dose profile calculated using commercial software (FotoLook 95 V2.08; AGFA, Livingstone, UK). The digitization was performed with a resolution of 118 pixels per centimetre and 212 grey levels; for the dose profiles obtained (Figure 3
) the value of FWHM was calculated by a simple macro with an uncertainty of less than 2%.
|
| Results |
|---|
|
|
|---|
Measurements were repeated six times on unit number 5 in order to test reproducibility. The repeated results agreed within 5.7%.
Uncertainties in the measurement of P arise from the pencil chamber calibration. A standard error of less than 5% at the 95% confidence level can be considered as typical.
| Discussion |
|---|
|
|
|---|
A pencil chamber can be used to make a simple measurement of DWP because the electrometer reads the charge product over the entire length of the chamber; the length of the chamber is greater than the width of the beam, so the result of the measurement corresponds to the integrated dose that can be considered equivalent to the DWP for a square beam profile.
The results obtained with the pencil chamber method agree within 8.6% with those obtained by the conventional measurement of DWP; reproducibility was 5.7%, comparable with 7.2% for the conventional method (series of six measurements).
The major advantage of using a pencil chamber in order to obtain DWP is its immediacy; the method does not require use of film and/or TLDs and the chamber can be easily positioned for the measurement.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
Received for publication April 8, 2002. Revision received July 31, 2002. Accepted for publication October 21, 2002.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. J. Martin RADIATION DOSIMETRY FOR DIAGNOSTIC MEDICAL EXPOSURES Radiat Prot Dosimetry, March 28, 2008; (2008) ncm495v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Lofthag-Hansen, A Thilander-Klang, A Ekestubbe, E Helmrot, and K Grondahl Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD Dentomaxillofac. Radiol., February 1, 2008; 37(2): 72 - 79. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Doyle, C J Martin, and J Robertson Techniques for measurement of dose width product in panoramic dental radiography Br. J. Radiol., February 1, 2006; 79(938): 142 - 147. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Helmrot and G. Alm Carlsson Measurement of radiation dose in dental radiology Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 168 - 171. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |