| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Short communication |
Applied Physics Department, Konstantin Preslavsky University, 9712 Shumen, Bulgaria
| Abstract |
|---|
|
|
|---|
Ca and scatter fraction Ps and two subjectively estimated, a low contrast visualization index Plow and a high contrast visualization index Phigh. To demonstrate the potential of this phantom method it was applied to an X-ray unit to find the optical film density that ensures optimal visualization in different anatomical areas. It was found for the X-ray system under investigation that the automatic exposure control could be set to produce an optical density of about 1.8 in the lung field. The reported method is easily implemented in any clinical situation where optimization of chest radiography is needed. | Introduction |
|---|
|
|
|---|
Image quality can be investigated in two ways, one based on physical measurements and the other on psychophysical assessments [2]. Many authors show that objective physical measurements alone are not sufficient to demonstrate the clinical advantage of one imaging technique over another [2, 3]. Subjective clinical evaluation of patients' radiographs, although being a powerful method, needs considerable resources and is often unacceptable for radiation protection reasons. The method of choice therefore involves a complex investigation of images of appropriate phantoms.
Various phantoms of different complexity are in use to simulate the human thorax. Anthropomorphic phantoms simulating shape, size and tissue composition are not available in each X-ray department because of the relative complexity of manufacture and high cost. Simpler geometrical phantoms are also designed for imaging and dosimetric purposes [46].
One of the most frequently used geometrical phantoms is the LucAl phantom (Standard Dosimetric/Calibration Phantom; Center for Devices and Radiological Health, Carson, CA), described by Conway et al [4]. The LucAl phantom consists of 250 mm x 250 mm polymethyl-methacrylate (PMMA) plates and 1100 aluminum (Al) alloy with specified thicknesses (Figure 1a
). The overall thickness of the phantom is 267 mm, with 4.1 mm Al, 73 mm PMMA and a 190 mm air gap. The thicknesses and relative positions of the different components have been designed to accurately simulate primary and scatter transmission through the lung-field regions of a patient-equivalent anthropomorphic chest phantom (Humanoid Systems, Carson, CA) for up to 150 kVp X-rays typically used in chest radiography. Good spectral equivalence between LucAl and Alderson-Rando male and female anthropomorphic phantoms has also been found [6]. Clinical tests have shown that the LucAl phantom reliably approximates a 22.5 cm thick patient for a posteroanterior (PA) chest projection controlled by the lateral automatic exposure control (AEC) chambers [4]. The phantom was originally designed for dose measurements during chest radiography with AEC [4]. Subsequently, it was also used for image quality evaluation by inserting different test objects into the phantom [7, 8]. However, the image of this homogeneous phantom does not provide information regarding image quality under the clinically important mediastinal region of the thorax. At the same time it was shown that approximately 26% of the lung volume and 43% of the lung area are obscured by organs [9].
|
| Anthropomorphic LucAl phantom |
|---|
|
|
|---|
For image quality evaluation, different test objects were arranged on a 10 mm thick PMMA plate. This image quality test (IQT) plate was inserted into the phantom, substituting for one PMMA layer with the same thickness (Figure 1
). The position of the IQT plate can be varied according to the imaging task. The different test objects are arranged on the IQT plate to determine image quality parameters in three important anatomical areas; lung, cardiac, and subdiaphragmal regions. The objects' positioning is shown in Figure 1c
. The objects are set in two groups, according to the method of estimation of quality parameters. The test objects for subjective analysis are situated on the left side of the IQT plate and those for OD measurements are situated on the right side.
Objects for subjective estimation of image quality
The objects to be simulated, their size and structure were selected following clinical requirements for chest radiography [1, 3]. Phantom materials were chosen with respect to their tissue equivalence and accessibility [10]. Low contrast objects, such as subtle lung nodules or small tumours with diffuse edges, are simulated by three groups of eight paraffin wax hemispheres (N1, N2, N3) with diameters from 3 mm up to 10 mm. Four polyethylene cylinders (M) with diameters from 7 mm to 20 mm with different heights are used to simulate lesions with more defined edges. Pulmonary infiltrates are represented by large flakes of paraffin wax 12 mm thick (I1, I2). Small vessels are simulated by nylon cords with a diameter of 1.5 mm and 2.0 mm (V1, V2). Groups of these objects were positioned in the lung, cardiac and subdiaphragmal areas of the phantom. Two groups of grain, one of 0.5 mm diameter copper (G1, G2) and one of 1 mm diameter polytetrafluoroethylene (F1, F2), simulating fibrosis, are situated in the lung and the retrocardiac areas. Two polyvinylchloride cylinders with drilled holes of 2.5 mm and 5 mm in diameter and 2 mm high (L1, L2), positioned in the lung region, are used to determine low contrast visualization. To assess high contrast resolution, two groups of wire meshes are positioned over the "lung" and "heart", respectively (W1, W2). Each group consists of four meshes with 9 wires, 13 wires, 15 wires and 30 wires per mm, respectively, the first two with 0.2 mm diameter wire, the third with 0.15 mm diameter wire and the fourth with 0.1 mm diameter wire. Two groups of copper wires of 0.3 mm, 0.2 mm and 0.14 mm in diameter (P1, P2) are used to simulate pneumothorax. Additionally, a resolution test object consisting of 11 line groups with spatial frequency from 0.6 mm-1 up to 3.3 mm-1 (H) is positioned in the lung field to check the system's high contrast resolution capability. The positions of all these objects can be seen in Figure 1c
.
Objects for physical measurements
20 areas for OD measurement are arranged on the IQT plate to enable evaluation of objective image quality indices. To this end, polyethylene cylinders 20 mm in diameter are arranged in five groups of two cylinders, one 4 mm and one 8 mm high in each group. These groups are placed in the lung, cardiac, subdiaphragmal and upper mediastinal areas and over the "rib" (Figure 2
). An additional cylinder is placed on the crossing of two ribs. To assess radiographic contrast, round reference areas are marked near the test objects. Lead stoppers 7 mm thick and 10 mm x 10 mm in size, to measure the scatter fraction, are arranged in three areas; lung, cardiac and subdiaphragmal regions. Using the electronic catalogue of spectra [11] it was found that this lead thickness corresponds to approximately 25 half value layers of X-rays generated at 120 kV, and practically fully stops the primary beam.
|
| Method for phantom image evaluation |
|---|
|
|
|---|
Objective quality indices
Two objective indices are proposed; the areal contrast index
Ca and a scatter fraction Ps. Optical densities ODi (i=120) are measured in the 20 fixed spots described above (Figures 1c and 2![]()
). Radiographic contrast for each object is calculated as the difference between ODi under the test object i and ODref in the relevant reference area. The areal contrast index
Ca (a=lung, heart, subdiaphragm, upper mediastinum, ribs) is calculated by summing the density differences for the objects located in each of the five anatomical areas.
The scatter fraction Ps, representing the contribution of scattered radiation to the image formation, is calculated using the formula:
|
|
where Ks and (Ks+Kp) are values of air kerma behind the lead stopper and close to it, respectively. These values are derived from measuring OD at these points and then transforming them into air kerma values using the SFC characteristic curve.
Subjective quality indices
Images are subjectively evaluated at optimal ambient light using a clinical viewing box with homogeneous brightness. Visualization of each object is scored using a four point scale; 3 if the object is clearly defined, 2 if it is visible, 1 if it is hardly discernible and 0 if it is fully missing. Two quality indices are then formed for each anatomical area by summing the scores of the low contrast objects N, M, I, V, F, G and L for the low contrast visualization index Plow and those of the high contrast objects P, W and H for the high contrast visualization index Phigh.
| Application of the anthropomorphic LucAl phantom |
|---|
|
|
|---|
le;1 and ±1.5% at OD>1.
Standard PA chest radiography was simulated on the chest stand with the anthropomorphic LucAl phantom. Eight films were exposed with a standard soft-beam technique of 70 kVp and 1 mm Al added filter. The only varied parameter was mAs (between 3.2 mAs and 16 mAs). The same radiographic cassette was used for each exposure. The quality indices
Ca, Plow and Phigh were evaluated. The incident air kerma on the phantom entrance surface was measured directly during exposure with the dosemeter WD 10 (Wellhofer, Schwarzenbruck, Germany) and a diode detector, with an error of less than 5%.
Results are presented in Table 1
and Figure 3
. Table 1
shows the change of optical density ODref in lung, cardiac and subdiaphragmal areas and in the ribs as a function of incident air kerma on the phantom surface. Areal contrast index
Ca for four anatomical areas (lung, cardiac and subdiaphragmal regions, and in the ribs) as a function of ODlung is presented in Figure 3a
. Subjectively, derived low contrast visualization index Plow for lung, cardiac and subdiaphragmal areas are given in Figure 3b
as a function of ODlung. On the same graph this dependence is shown for high contrast visualization index Phigh in the lung field.
|
|
Ca in the cardiac and subdiaphragmal areas increases but
Ca in the lung field first increases until OD=1.7, and then drops. In the ribs,
Ca is maximum at ODlung >2.1 when the ODribs reaches approximately 1.8. In Figure 3bThe contrast in cardiac and subdiaphragmal areas increases quickly with increasing ODlung. This is the reason that subjective image quality indices in these areas increase. Above ODlung=2.0, however, there is a rapid deterioration in visualization of important subtle round details in the lung with a diameter of 3 mm and low contrast linear details with a diameter of 12 mm. This is owing to contrast reduction in the lung as well as decreased visual perception at higher ODs, which also depends on viewing box brightness.
Taking into account changes of all quality indices, we could decide that in our particular case, the AEC could be set to produce ODlung in the lung field of approximately 1.8. The optimal OD has to be found for each particular combination of SFC, viewing box and viewing conditions.
| Conclusion |
|---|
|
|
|---|
By modifying the standard LucAl chest phantom by adding an anthropomorphic insert and image quality test plate, we can evaluate the performance of different components of the imaging chain and their influence on the final result; image quality and patient dose. This method is easily applicable to any clinical situation in which optimization of the procedure is needed.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received for publication January 29, 2002. Revision received June 25, 2002. Accepted for publication July 5, 2002.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C S MOORE, A W BEAVIS, and J R SAUNDERSON Investigation of optimum X-ray beam tube voltage and filtration for chest radiography with a computed radiography system Br. J. Radiol., October 1, 2008; 81(970): 771 - 777. [Abstract] [Full Text] [PDF] |
||||
![]() |
C S Moore, G P Liney, A W Beavis, and J R Saunderson A method to optimize the processing algorithm of a computed radiography system for chest radiography Br. J. Radiol., September 1, 2007; 80(957): 724 - 730. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Vassileva A phantom approach to find the optimal technical parameters for plain chest radiography Br. J. Radiol., August 1, 2004; 77(920): 648 - 653. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |