British Journal of Radiology 75 (2002),435-443 © 2002 The British Institute of Radiology
Patient and staff exposure during endoscopic retrograde cholangiopancreatography
N Buls, MSc1,
J Pages, PhD1,
F Mana, MD2 and
M Osteaux, MD, PhD1
Departments of 1 Radiology and Medical Imaging and 2 Gastro-Enterology, Free University Hospital Brussels (AZ-VUB), Laarbeeklaan 101, B-1090 Brussels, Belgium
 |
Abstract
|
|---|
Despite a number of efforts being put into the radiological protection of both patient and staff during interventional radiological (IR) procedures during recent years, information about radiation exposure during endoscopic retrograde cholangiopancreatography (ERCP) procedures remains scarce. The purpose of this study was to estimate both patient and staff radiation doses during therapeutic ERCP procedures by direct measurement and to compare these results with data from other IR procedures. For 54 patients, effective dose and skin dose were estimated by measuring the dosearea product. For staff, entrance surface doses to the lens of the eye, thyroid and hands were estimated by thermoluminescent dosemeters. A median effective dose of 7.3 mSv and a median entrance surface dose of 271 mGy per procedure were estimated for patients. The gastroenterologist received a median dose of 0.34 mGy to the lens of the eye, 0.30 mGy to the skin at the level of the thyroid and 0.44 mGy to the skin of the hands, per procedure. When comparing the dosimetric quantities presented in this study with data from other IR procedures, it is clear that patient skin doses and doses to staff are high owing to the use of inappropriate X-ray equipment. ERCP requires the same radiation protection practice as all IR procedures. It should be consistently included in future multicentre IR patient and staff dose survey studies at national or international level.
 |
Introduction
|
|---|
Endoscopic retrograde cholangiopancreatography (ERCP) was first introduced in 1969 [1]. It is a common interventional procedure that examines the duodenum, papilla of Vater, bile ducts, gallbladder and pancreatic duct. Although ERCP was introduced as a purely diagnostic procedure, technical advances in instrumentation during the last 2 decades have added a therapeutic dimension.
During the performance of ERCP, a high amount of X-ray fluoroscopy and digital radiographs are performed, making it an interventional radiological (IR) procedure [2]. It is well known that some of the highest doses to patients from diagnostic medical X-rays other than CT arise from IR procedures [3]. One of the key issues emphasised during the recent International Conference on Radiological Protection of Patients in Diagnostic and Interventional Radiology, organized by the International Atomic Energy Agency, was that there is a need for radiation risk evaluation and reference doses in IR to minimize somatic risks and to avoid deterministic injuries [4]. Despite the fact that, during recent years, a number of efforts have been put into the protection of both patients and staff, information about radiation exposure during ERCP remains scarce. Published work on radiation protection in IR is mainly focussed on various angiographic and cardiological procedures [516]. Few international papers have been published that evaluate radiation exposure to patients [7, 1720] and staff [2123] during ERCP.
The purpose of this study is to estimate patient dose, with respect to Article 8 of the Euratom 97/43 directive [24], and to estimate the dose to the lens of the eye, thyroid and hands of staff. For the patient, the effective dose E, as an index of the risk for stochastic effects [25], and the entrance surface dose (ESD) are estimated by measuring the dosearea product (DAP). Doses to critical organs located in the area of exposure are also estimated. The ESD is estimated in order to assess the possibility of skin dose exceeding the threshold for deterministic effects. Concerning staff, the ESD to the eyes, neck and hands are estimated by using thermoluminescent dosemeters (TLDs).
It should be noted that, as in many other IR procedures, ERCP is not performed by a radiologist but by a gastroenterologist (GE). Although it is a requirement of the Euratom 97/43 directive [24] that staff performing practical aspects of a medical exposure should have received adequate training in radiation protection, a GE has not had in-depth training in radiation management using diverse forms of fluoroscopic equipment, nor in the potential harmful effects to patients and staff.
 |
Materials and methods
|
|---|
During a 4-month period (MaySeptember 2000), patient and staff doses were evaluated from 54 therapeutic ERCP examinations; 23 male and 31 female patients. The age of the male group ranged between 43 years and 77 years, with a mean age of 63 years. The age of the female group ranged between 40 years and 85 years, with a mean age of 70 years.
Radiographic unit
The ERCP procedures were performed on a General Electric Prestilix 1600 DRS radiographic unit (General Electric Medical Systems, Milwaukee, WI), equipped with an overcouch tube (model RSN722; General Electric Medical Systems, Milwaukee, WI) (total filtration=3.2 mm Al equivalent) and a high frequency generator. Input air kerma rates for each image intensifier field size selection, in standard and high detail mode, are listed in Table 1
. The input air kerma per image for the digital radiographs were 2.34 µGy (field size 15 cm), 1.59 µGy (23 cm) and 0.81 µGy (38 cm). Measurements were performed on top of the image intensifier with a 150 cm3 ionization chamber (Keithley, Cleveland, OH) and 2 mm additional copper filtration.
ERCP procedure, positioning of the staff and the patient
All the ERCP procedures considered in this study were therapeutic (sphincterotomy by using electrocautery with extraction of biliary stones or placement of a prosthesis in the biliary tract).
Almost every ERCP is performed under local sedation. Full anaesthesia is used in only a few cases. No such cases are included in this study. Endoscopy is always performed with the patient in the recumbent left lateral position. During the procedure, fluoroscopic screening and digital radiographs of the hepaticbiliary area are taken with a varying tube potential between 100 kVp and 125 kVp, depending on patient size. All exposures are performed in the automatic brightness control (ABC) mode. The procedure has a fixed set-up, the patient stays in the same position and the same anatomical area is exposed during the entire examination.
An overview of patient and staff positions is shown in Figure 1
. During the procedure, the GE, who is assisted by a nurse (N1), stands with his left side next to the patient and manipulates the duodenoscope. A second nurse (N2) holds the patient's head in order to position the duodenoscope. With respect to dosimetry, it is important to note that both nurses do not remain in the same position all the time. They often leave their position for short durations to operate monitoring equipment or to reach other medical equipment. All members of staff are classified as occupationally exposed workers and are, in consequence, already monitored by a regulatory dosemeter worn at chest level. They all wear wrap around aprons during procedures, providing a protection of 0.50 mm lead equivalence at the front of the body and 0.25 mm lead equivalence at both sides and the back of the body.

View larger version (35K):
[in this window]
[in a new window]
|
Figure 1. Staff positions relative to the patient and X-ray tube during a typical endoscopic retrograde cholangiopancreatography procedure. Distance variations are marked in centimetres. G.E., gastroenterologist; N.1., assisting nurse; N.2., second nurse.
|
|
A radiographer controls the X-ray unit at the console behind a protective shield. He performs fluoroscopy and digital spots according to the indications of the GE.
Estimation of patient dose
In the field of interventional radiology, where both fluoroscopy and digital radiography are used at various kilovoltages and various X-ray projections are assessed, DAP as a measure of radiation dose to the patient is widely applied [3]. DAP is measured by a transmission ionization chamber (model TV57523-1; PTW, Freiburg, Germany) mounted on the collimator box of the X-ray tube and connected to a Diamentor M2 (PTW, Freiburg, Germany) read-out unit. The DAP equipment was initially calibrated by the Physikalisch-Technische Bundesanstalt (Braunschweig, Germany) and is verified every 6 months in our department.
During examination, DAP values were determined for each fraction of fluoroscopic exposure and digital radiograph as a function of the selected kilovoltage and the anatomical projection. Diasoft software (PTW, Freiburg, Germany) was used to register DAP values. Effective dose and organ doses were estimated separately for each patient by using recorded DAP values in combination with effective dose conversion factors derived from Monte Carlo computer calculations for male and female mathematical anthropomorphic phantoms [26]. These factors are expressed as a function of beam quality, applied kilovoltage and radiographic projection. Since the radiographic projection that was used for a typical ERCP is not specifically defined by Hart et al [26], it was approximated by a combination of the stomach left lateral and the duodenum left posterior oblique projections. An overview of a typical ERCP procedure in terms of irradiation site, radiographic projection (as used for the dose calculation), type of exposure (fluoroscopy or digital images), image intensifier field size selection, tube potential, air kerma mode for fluoroscopy and number of digital radiographs is shown in Table 2
. For a typical ERCP procedure, four DAP values were determined; one for each kilovoltage at each projection. Fractionation of the DAP data was carried out during the procedure with the Diasoft software (PTW, Freiburg, Germany). Dose was calculated for each patient from each contributing DAP fraction and the summation per organ was made, giving the total effective dose and organ doses from that procedure.
ESD was estimated by DAP value and a conversion factor that was experimentally determined by phantom measurements. An Alderson-rando phantom (Phantom Laboratory, Salem, NY) was used to simulate the patient. It has been shown that this phantom can be used to simulate a patient for diagnostic radiographs with energies higher than 90 kVp [27]. Measurements were performed with the same radiographic equipment and in the same geometric conditions as during a typical ERCP; the phantom was placed in the recumbent left lateral position and a radiographer, who frequently performs ERCP studies, determined the radiographic projection. For each field size selected, ESD and DAP were repeatedly measured for various levels of tube load. For each measurement, ESD was calculated as the average reading of five TLDs positioned in a cross pattern in the centre of the exposed area on the body of the phantom. The focusskin distance (FSD) was 64 cm. The dosemeters (TLD-100 LiF: Mg, Ti; 3.2 mm x 3.2 mm x 0.9 mm; HarshawBicron, Wermelskirchen, Germany) were individually calibrated in air at 110 kVp (this equals the average energy used during ERCP) on a diagnostic X-ray unit (Orbix; Siemens, Erlangen, Germany) equipped with an overcouch tube (Bi 125/20/40R-100L; Siemens, Erlangen, Germany; total filtration, 2.9 mm Al equivalent) and a 12 pulse generator (Garantix 800; Siemens, Erlangen, Germany). The exposure measurements were conducted with a 30 cm3 cylindrical ionization chamber (model PM-30; Capintec, Pittsburg, KS) connected to an electrometer (model 192; Capintec, Pittsburg, KS). The TLDs were processed with a Harshaw-Bicron (Wermelskirchen, Germany) 5500 TLD reader in an atmosphere of inert nitrogen and annealed in a PTW-TLDO oven (PTW, Freiburg, Germany).
Estimation of staff dose
As a classified worker, each member of staff is already monitored by a regulatory dosemeter worn at chest level under the apron. Since we were interested in the ESDs per procedure on parts of the body that are not protected by the apron, additional measurements were performed. ESDs at three different positions on the body were estimated by using a batch of 50 TLDs TLD-100H (LiF: Mg, Cu, P; 3.2 mm x 3.2 mm x 0.5 mm; Harshaw-Bicron, Wermelskirchen, Germany). For each position, three TLD chips were inserted in a thin polyethylene plastic bag and attached to the skin by adhesive tape. Three bags with a total of nine TLD chips were used for each staff member per ERCP procedure. One bag was attached to the forehead, one to the skin of the neck at the level of the thyroid and one to the dorsal surface of the hand closest to the patient (left hand for the GE and N1 and right hand for the N2). TLDs were calibrated and processed as previously described.
 |
Results
|
|---|
Estimation of patient dose
Phantom measurements show that for the range of ERCP exposure settings (image intensifier field size 38 cm at approximately 90 kVp, field size 23 cm at approximately 105 kVp and field size 15 cm at approximately 125 kVp), the ESD and DAP exhibit an almost perfect linear relationship (r=0.99, p<0.001). Data are shown in Figure 2
(error bars represent 95% confidence intervals). Uncertainties are estimated to be 6% for the DAP reading (including DAP meter calibration and DAP meter response) and 8% for the ESD reading (including TLD calibration, TLD response and TLD positioning). A DAP (Gycm2) to ESD (Gy) conversion coefficient of 6.96 x 10-3 cm-2 is determined for the estimation of the ESD to patients. As shown in Figure 2
, the coefficient is independent of image intensifier field size selection. This is owing to the ABC system of the radiographic equipment, which adjusts exposure factors in order to deliver a constant image intensifier output signal. A smaller image intensifier field size selection automatically implies a smaller field of view, which is compensated by the ABC system increasing exposure factors. As a result, the DAP value stays constant for the three image intensifier field size selections. This makes the established DAP to ESD coefficient independent of the image intensifier field size selection on condition that no additional collimation is used and the ABC system works properly (which is the case for this study). However, caution is necessary when using this conversion coefficient. It was established using a fixed distance between tube and phantom and, despite the fact that phantom and patients were exposed in the same geometrical condition, the FSD will vary slightly depending on the thickness of the patient. A variation in FSD will strongly affect the air kerma at the skin surface, owing to the inverse square law, and thus the ESD. Because of this uncertainty, patients with body weights below 60 kg and above 80 kg were excluded from this study and a 10% variation in FSD was accepted, resulting in a 20% variation of the air kerma at the patient's skin surface (see Accuracy of the estimations).
An overview of the patient dose data per ERCP procedure in terms of DAP, fluoroscopic screening time, ESD and effective dose is presented in Table 3
. Since the data present a skewed distribution, the 1st quartile, median, 3rd quartile and maximum values are shown as well as the mean. The contribution of the fluoroscopic fraction to the total DAP value, averaged over the 54 procedures, is 92%. The DAP and effective dose show a very good linear correlation (r=0.98, p<0.001). A conversion factor of 0.19 mSv Gy-1 cm-2 is calculated based on the median values of the two variables.
View this table:
[in this window]
[in a new window]
|
Table 3. Patient exposure data per endoscopic retrograde cholangiopancreatography procedure, evaluated over 54 examinations
|
|
The following median organ doses were estimated per ERCP procedure: liver, 40.5 mGy; small intestines, 29.1 mGy; colon, 28.4 mGy; stomach, 13.9 mGy; uterus, 7.7 mGy; and gonads, male 0.1 mGy, female 5.9 mGy.
Estimation of staff dose
An overview of the ESD data to the eyes, neck and hand of the GE, N1 and N2 per ERCP procedure is listed in Table 4
.
View this table:
[in this window]
[in a new window]
|
Table 4. Data on entrance surface dose (ESD) to the medical staff per endoscopic retrograde cholangiopancreatography (ERCP) procedure
|
|
Accuracy of estimations
For the patient effective dose, uncertainty is estimated to be 16%, including DAP meter calibration (3%), DAP meter response (5%) and DAP to E conversion factor (15%). As in other published work on patient doses, standard International Commission on Radiological Protection weighting factors [25] have been applied. There will be an additional error in applying these weighting factors to a skewed, elderly population, but this is probably unquantifiable at the present time and has not been included in the above estimates. For patient ESD, the uncertainty is estimated to be 22%, including TLD calibration (3%), TLD response (5%), TLD positioning (5%), DAP meter calibration (3%), DAP meter response (5%) and variation in air kerma at skin level owing to variation in FSD between patient and phantom (20%). For ESD to staff, uncertainty is estimated to be 8%, including TLD calibration (3%), TLD response (5%) and TLD positioning (5%).
 |
Discussion
|
|---|
Patient dose
From the data in Table 3
, it can be seen that exposure to the patient shows large variation. This can easily be understood considering the influence of patient size on tube loading and the variation in the assessed screening time according to the conduct of the procedure and the patient's condition. A median effective dose of 7.3 mSv and a median skin dose of 271 mGy were estimated for the group of 54 patients.
In all cases, ESD remained well below the threshold single fraction dose of 2 Gy for transient erythema [28]. The 3rd quartile value of the group was 420 mGy and the maximum measured value was 1180 mGy. Nevertheless, care should be taken since some "high dose" patients can undergo two or more ERCP procedures consecutively. In normal circumstances, the time period between consecutive ERCPs of a patient is 48 h.
Table 5
compares the effective dose and ESD estimated in this study with data from other common IR procedures previously published in the literature, with special interest to the biliaryhepatic region. Almost all data are retrieved from McParland [5, 6] and Ruiz-Cruces et al [7]. All values presented in Table 5
are median values unless stated differently. A comparison of effective dose from different procedures is made since this property is an index of the risk of stochastic effects. DAP on its own is useful when comparing separate procedures of a similar type e.g. ERCP in this case, but since this value is independent of the radiosensitivity of the screened anatomical region, effective dose is a better index when comparing different types of radiological procedures [5].
View this table:
[in this window]
[in a new window]
|
Table 5. Comparison of patient effective dose and entrance surface dose (ESD) from endoscopic retrograde cholangiopancreatography (ERCP) to those from other interventional radiological procedures published in the literature
|
|
For ERCP patient doses, a comparison is made with data from Ruiz-Cruces et al [7] Heyd et al [17] and Larkin et al [18]. Heyd et al [17] estimated a similar ESD by measuring the dose on the image intensifier and applying a conversion factor determined by phantom measurements. As in our study, Larkin et al [18] measured DAP for 12 ERCP procedures and used the PCXMC Monte Carlo software (Finnish Centre for Radiation and Nuclear Safety, Helsinki, Finland) to estimate E. The procedures they studied required longer fluoroscopic screening times (mean 10.5 min), resulting in higher patient doses; mean DAP, 66.8 Gycm2 and estimated mean E, 12 mSv. The DAP to E coefficient was 0.18 mSv Gy-1 cm-2 and the fluoroscopic exposure component contributed 90% of the total dose, which is in good agreement with our results. Ruiz-Cruces et al [7] determined the DAP and ESD for two types of ERCP procedure and used the EFF-DOSE software (National Board of Health, National Institute of Radiation Hygiene, Bronshoj, Denmark) to estimate E. For biliary stone extraction (nine patients) they measured a median DAP of 72.4 Gycm2 (estimated E, 10.8 mSv) and for biliary stent prosthesis placement (nine patients) they measured a median DAP of 83.6 Gycm2 (estimated E, 12.5 mSv). This yields a lower DAP to E coefficient of 0.15 mSv Gy-1 cm-2, which can be explained by the fact that they used another radiographic projection for E estimation, since the two procedures are limited to the biliary tract, whereas the data in our study and in that of Larkin et al [18] cover several types of ERCP procedure and thus a wider anatomical area. Marshall et al [19] published a median DAP value of 8.6 Gycm2 (mean, 14.5 Gycm2), which was obtained from a large group of 1736 patients in seven X-ray rooms as a part of a large dose monitoring programme in the North of England. Broadhead et al [20] proposed a reference DAP value of 19.4 Gycm2, obtained from 3rd quartile values of a large group of 3297 patients (median DAP value, 9.0 Gycm2), also as part of a large dosimetry survey in the North of England. Although neither Marshall et al nor Broadhead et al stated explicitly which types of ERCP procedures were considered, the lower DAP values suggest that these studies were primarily diagnostic ERCP.
The effective dose to the patient from an ERCP procedure is of the same order of magnitude as the dose from other IR procedures listed in Table 5
. However, the ESD is relatively high owing to the clinical technique used and the limitations of the radiographic equipment (fixed overcouch tube). The area of irradiation remains fixed during the entire ERCP procedure, whereas during the other procedures listed in Table 5
the exposure can be distributed over a larger skin area, which reduces the maximum skin dose value to a specific point.
Staff dose
The data presented in Table 4
illustrate that the GE receives the highest dose to the eyes, thyroid and hands (all cases p<0.02, two-tailed Spearman rank correlation test) in relation to the nursing staff owing to proximity to the area of exposure. Doses to the two members of nursing staff are comparable. Distance to the patient of the N2, handling the patient's head, is larger than that of the N1. The latter, however, is partially shielded by the GE. Although dose to the hands is higher, dose to the lens of the eye will be the critical organ relative to regulatory dose limits. Considering the annual dose limit to the lens of the eye of 150 mSv [25] for a classified worker, and the 3rd quartile values presented in Table 4
, a GE could annually perform approximately 225 ERCP procedures before reaching the dose limit. 3rd quartile values are used from a pragmatic radiation protection point of view. For N1 this value is approximately 440, and for N2 305. As the annual number of ERCP procedures performed in our hospital by one GE is approximately 120, it is very unlikely that this limit will be reached. Nevertheless, doses to staff can and should be decreased to a reasonably achievable level [24, 25].
Table 6
compares the doses to ERCP staff with those from other IR procedures published in the literature. All values are median values unless stated differently. Germanaud et al [21] also used a conventional overcouch tube to perform ERCP and assessed direct measurements of staff dose by thermoluminescent dosimetry. The median dose to the lens of the eye was found to be 0.13 mGy per procedure and the ESD at thyroid level was measured at 0.10 mGy. Although dose to staff was found to be lower in their study, the mean fluoroscopic screening time (6.8 min) was comparable with the mean screening time in our study. The difference in staff doses can be attributed to several factors: they used a lower kilovoltage (between 75 kV and 95 kV), which affects the amount and energy of scatter radiation and therefore dose; location of the GE; and the difference in dose rate during fluoroscopy and statistical variation. Cohen et al [22] estimated a much higher dose when performing indirect phantom measurements and assuming a fluoroscopic screening time of 20 min. Unfortunately, they did not specify the type of X-ray equipment used.
View this table:
[in this window]
[in a new window]
|
Table 6. Comparison of endoscopic retrograde cholangiopancreatography staff doses with other interventional radiological (IR) procedures published in literature
|
|
Dose to staff per ERCP procedure is high compared with data from other IR procedures. This is clearly owing to the type of X-ray equipment used. Since the procedure is performed on a general X-ray unit equipped with an overcouch tube, doses to the upper body parts of staff are high given the high amount of scatter radiation. This is well illustrated by data from Bowsher et al [12] who evaluated the dose to staff during renal surgery with an overcouch and undercouch X-ray tube. The other IR procedures listed in Table 6
were performed with a C-arm or an undercouch tube.
General discussion
This study demonstrates that DAP monitoring is an easy and useful tool for estimating patient effective dose and ESD in ERCP procedures. From the dosimetric quantities presented in this study, it is obvious that an ERCP procedure has the potential to cause high patient and staff doses and consequently requires attention regarding radiation protection. Whilst it is well known that an overcouch tube X-ray unit is not adequate for performing IR procedures [10], all of the studies consulted regarding ERCP [1723] involved the use of this type of equipment (or did not specify which equipment type was used). This may indicate that ERCP procedures are often performed without attention to equipment and radiation protection in comparison with more common IR procedures such as angiographic and cardiological interventions. The reason for this lack of ERCP dedicated X-ray equipment is probably cost-benefit related.
It is clear that a correctly operated C-arm unit with the availability of pulsed fluoroscopy will dramatically reduce dose to both patients and staff. In a second stage, staff protection can be complemented by a thyroid collar shield, protective glasses for the GE and protective gloves for the N2. Secondary radiation shielding can also be very effective in reducing staff exposure. Chen et al [23] showed, in a 20 patient ERCP study, that a 0.5 mm lead equivalent acrylic shield reduced staff exposure by a factor of 11. The shield did not interfere with performance of the examination.
Another important factor affecting patient and staff exposure is communication between the radiographer and the GE. The radiographer should only perform fluoroscopic screening as required by the GE, and the latter should indicate clearly when no more screening is needed. From our experience it is clear that a relatively large amount of unnecessary exposure is delivered when no strict arrangements are made between staff members.
The authors advise that ERCP should be consistently included in future clinical multicentre IR patient and staff dose survey studies at national and international level.
 |
Acknowledgments
|
|---|
The authors wish to thank W Vroonen, RT, M Freson, MD, D Urbain, MD and the nursing staff of the gastroenterology department for their cooperation in obtaining measurements, as well as W Wijnen for his help in editing the manuscript.
Received for publication May 24, 2001.
Accepted for publication November 29, 2001.
 |
References
|
|---|
-
Evans AF. Endoscopic retrograde cholangiopancreatography (ERCP). In: Whitehouse GH, Worthington BS, editors. Techniques in diagnostic radiology. Oxford: Blackwell Scientific Publications 1983:8790.
-
Adam A. The definition of interventional radiology (or, "When is a barium enema an interventional procedure?"). Eur Radiol 1998;8:10145.[Medline]
-
National Radiological Protection Board. Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures. Documents of the NRPB 10(1). London: HMSO, 1999:216.
-
Faulkner K, Vaño E, Padovani R, Zoetelief J. Radiation risk evaluation and reference doses in interventional radiology. In: The International Conference on Radiological Protection of Patients in Diagnostic and Interventional Radiology, Nuclear Medicine and Radiotherapy; 2001 March 2630; Malaga. Vienna: International Atomic Energy Agency, 2001.
-
McParland BJ. A study of patient radiation doses in interventional radiological procedures. Br J Radiol 1998;71:17585.[Abstract]
-
McParland BJ. Entrance skin dose estimates derived from dose-area product measurements in interventional radiological procedures. Br J Radiol 1998;71:128895.[Abstract]
-
Ruiz-Cruces R, Pérez-Martínez M, Martín-Palanca A, Flores A, Christófol J, Martínez-Morillo M, et al. Patient dose in radiologically guided interventional vascular procedures: conventional versus digital systems. Radiology 1997;205:38593.[Abstract/Free Full Text]
-
Mooney RB. Skin dose to patients from interventional radiology and cardiology procedures with potentially long fluoroscopy times. Radiat Prot Dosim 2000;90:1236.[Abstract]
-
Williams JR. The interdependence of staff and patient doses in interventional radiology. Br J Radiol 1997;70:498503.[Abstract]
-
Faulkner K, Vañó A, Ortiz P, Ruiz R. Practical aspects of radiation protection in interventional radiology. In: The 10th International Congress of the International Radiation Protection Association; 2000 May 1419; Hiroshima. The International Radiation Protection Association, 2000.
-
Lima FRA, Khoury HJ, Hazin A, Luz LP. Doses to the operating staff during interventional cardiology procedures. In: The 10th International Congress of the International Radiation Protection Association; 2000 May 1419; Hiroshima. The International Radiation Protection Association, 2000.
-
Bowsher WG, Blott P, Whitfield HN. Radiation protection in percutaneous renal surgery. Br J Urol 1992;69:2313.[Medline]
-
Felmlee JP, McGough PF, Morin RL, Classic KL. Hand dose measurements in interventional radiology. Health Phys 1991;60:1657.
-
Janssen RJ, Hadders RH, Henkelman MS, Bos AJ. Exposure to operating staff during cardiac catheterisation measured by thermoluminescence dosimetry. Radiat Prot Dosim 1992;43:1757.[Abstract]
-
Vehmans T. Radiation exposure during standard and complex interventional procedures. Br J Radiol 1997;70:2968.[Abstract]
-
Ishiguchi T, Nakamura H, Okazaki M, Sawada S, Takayasu Y, Hashimoto S, et al. Radiation exposure to patient and radiologist during transcatheter arterial embolization therapy for hepatocellular carcinoma. In: The 10th International Congress of the International Radiation Protection Association; 2000 May 1419; Hiroshima. The International Radiation Protection Association, 2000.
-
Heyd RL, Kopecky KK, Sherman S, Lehman GA, Stockberger SM. Radiation exposure to patients and personnel during interventional ERCP at a teaching institution. Gastrointest Endosc 1996;44:28792.[Medline]
-
Larkin CJ, Workman A, Wright RE, Tham TC. Radiation doses to patients during ERCP. Gastrointest Endosc 2001;53:1614.[Medline]
-
Marshall NW, Chapple CL, Kotre CL. Diagnostic reference levels in interventional radiology. Phys Med Biol 2000;45:383346.[Medline]
-
Broadhead DA, Chapple CL, Faulkner K. Reference doses during fluoroscopic procedures. Radiat Prot Dosim 1998;80:1434.[Abstract]
-
Germanaud J, Legoux J, Sabattier R, Causse X, Trinh DH. Radiation protection of operators during endoscopic retrograde cholangiopancreatography. Gastroenterol Clin Biol 1993;17:25963.[Medline]
-
Cohen RV, Aldred MA, Paes WS, Fausto AM, Nucci JR, Yoshimura EM, et al. How safe is ERCP to the endoscopist? Surg Endosc 1997;11:6157.[Medline]
-
Chen MY, Van Swaeringen FL, Mittchell R, Ott DJ. Radiation exposure during ERCP: effect of a protective shield. Gastrointest Endosc 1996;43:15.[Medline]
-
Council directive 97/43/Euratom on Health protection of individuals against the dangers of ionising radiation in relation to medical exposure. Official Journal of the European Community 1997;L180:22.
-
International Commission on Radiological Protection. Recommendations of the International Commisson on Radiological Protection, ICRP publication 60. Oxford: Pergamon, 1991.
-
Hart D, Jones DG, Wall BF. Estimation of effective dose in diagnostic radiology from entrance surface dose and dose-area product measurements. Report NRPB-R262. London: HMSO, 1994.
-
Shrimpton PC, Wall BF, Fisher ES. The tissue-equivalence of the Alderson Rando anthropomorphic phantom for X-rays of diagnostic qualities. Phys Med Biol 1981;26:1339.[Medline]
-
Wagner LK, Eifel PJ, Geise RA. Potential biological effects following high x-ray dose interventional procedures. J Vasc Interv Radiol 1994;5:7184.[Medline]
This article has been cited by other articles:

|
 |

|
 |
 
J. A CHAFFINS
Radiation Protection and Procedures in the OR
Radiol. Technol.,
May 1, 2008;
79(5):
415 - 428.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. K. Looe, F. Eenboom, N. Chofor, A. Pfaffenberger, M. Sering, A. Ruhmann, A. Poplawski, K. Willborn, and B. Poppe
Dose-area product measurements and determination of conversion coefficients for the estimation of effective dose in dental lateral cephalometric radiology
Radiat Prot Dosimetry,
April 1, 2007;
124(2):
181 - 186.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R S Livingstone, C G Koshy, and D V Raj
Evaluation of work practices and radiation dose during adult micturating cystourethrography examinations performed using a digital imaging system
Br. J. Radiol.,
November 1, 2004;
77(923):
927 - 930.
[Abstract]
[Full Text]
[PDF]
|
 |
|