Figure 5. Plate reader artefact. The horizontal white line (arrow) shown on this upright chest radiograph was caused by dirt on the light guide in the plate reader. The light guide collects light emitted from the imaging plate when it is scanned by the laser. Artefact remedy: the light guide of the photomultiplier tube was cleaned by service personnel.
|