| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Full paper |
1 Medical Physics Group, Radiology Department, Complutense University, 28040 Madrid
2 3, Ricardo de la Vega St., 28901 Getafe, Spain
Correspondence: Prof. L González, Catedra de Fisica Medica, Facultad de Medicina, Universidad Complutense, 28040 Madrid, Spain
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
In particular, the installation and use of X-ray equipment for medical diagnosis with regard to technical specifications in the design of a facility and the provision of radiation protection resources are governed by decrees 1891/1991 and 1976/1999. The latter provides quality criteria for the compulsory evaluation of X-ray images, on a yearly basis, as well as their adaptation to European quality standards, as the basis of a quality assurance programme that must have been implemented by June 2000. Compulsory initial acceptance tests on new equipment and reference performances for old equipment establish the baseline performance. These must satisfy the requirements of acceptability criteria in the decree, which agree with the European Commission document on criteria for acceptability of radiology (including radiotherapy) and nuclear medicine installations [7].
This regulation follows the recommended image quality and reference dose criteria of document EUR 16260 [8] for chest, skull, lumbar spine, pelvis and urinary tract examinations. For examinations using other projections, image quality can be evaluated by means of anatomical criteria developed for this purpose by the specialist responsible. This may set the initial average dose level as the provisional reference value for the examination, until such time as representative reference levels are determined.
Dose evaluations must be based on a minimum of 10 estimates for simple examinations (i.e. less than four projections, without fluoroscopy), which is the case in dentistry. The image reject rate should also be evaluated and recorded. Exceeding the reference doses requires investigation of the causes, may lead to a correction of the protocols and/or suitable repair or reconditioning of the equipment, followed by a subsequent check on the efficacy of the changes made.
As in the case of all examinations that include projections other than the ones described in document EUR 16260 [8], there are no anatomical criteria developed to assess dental radiological images, although a reference dose value of 7 mGy is adopted for intraoral dental radiology, as proposed by the International Atomic Energy Agency (IAEA) "Basic safety standards" [9]. In previous work, a set of specific quality criteria has been proposed for intraoral [10], orthopantomography and cranial teleradiography images [11] so that image quality can be compared.
There is no dose value for orthopantomography. Previous literature provides doses measured in phantoms [1216] or following hybrid procedures using patients and phantoms [17, 18]. Owing to different measurement points, parameters used, equipment and perhaps also different sensitivity of the films available when measurements were performed, results from the various authors are not comparable, and it is possible to find average values from below 100 µGy to almost 3 mGy. In particular, measurements performed on patients are scarce. Moreover, entrance surface dose measurement is not the only solution to determine reference levels. For example, in the UK, rather than measuring the actual patient dose at given skin locations, it is accepted practice to measure the dosewidth product at the film for the whole panoramic exposure and average adult settings, using a dosemeter and film. The reference value is 75 mGy mm [19]. As the width of the beam should be less than 10 mm, the dose measured at the cassette should be less than 7.5 mGy for the whole exposure. This method provides information that is independent of geometry, but it is difficult to correlate this with skin doses owing to beam movement. In Finland during 19951999, regulatory control of panoramic units has been performed by measuring the dosearea product [20], with values in the range 34254 mGy cm2 and a mean of 94 mGy cm2.
According to European Union article 4(2) of the Medical Exposure Directive [1], member states shall promote the establishment and use of diagnostic reference levels (DRLs). DRLs are defined as dose levels in groups of standard sized patients or standard phantoms, for typical examinations and for broadly defined types of equipment. These levels should not be exceeded for standard procedures when good and normal practice is applied with regard to diagnostic and technical performance. According to Radiation Protection Document 109 from the European Commission [21], DRLs play an important role in clinical auditing to guarantee the performance of diagnostic services and as a support to improve procedures. Point 14 of the document states the applicability of DRLs for standard procedures in all areas of diagnostic radiology, although it identifies areas of particular utility where a relatively large reduction in risk can be achieved, as in the case of examinations with more radiosensitive patients such as children.
Since orthopantomography is often performed in paediatric cases, this study describes an experiment undertaken to obtain a local reference dose value based on patient measurements. Also, a critical evaluation of the adopted reference dose in intraoral X-ray diagnostics is undertaken using data gathered from over 300 intraoral X-ray facilities.
| Method |
|---|
|
|
|---|
Doses in orthopantomography were monitored with lithium fluoride thermoluminescent dosemeters (TLDs) of type TLD-100 (Harshaw TLD/Bicron/NE-Technology, Solon, OH). These TLDs were calibrated individually. A sample of at least 10 patients per X-ray unit was used. TLD chips inside sachets were used in pairs to reduce uncertainties in measurements. The sachets were located using an elastic ribbon in the occipital region at the level of the posterior cerebral fossa at the occlusal plane in each patient and, in a number of cases, on the temporal region, to compare results.
Exposures were classed according to the tube potential (kV) and tube current (mA) settings usually employed in each facility (up to three different parameter sets were found to be used in given facilities, although operation with a constant technique is not uncommon), using a sachet in each typical setting for between 5 and 21 patients. Data from three University facilities, a large hospital, a private centre devoted exclusively to dental radiology and five private dental clinics, were collected. 12 X-ray units were studied. A total of 190 measurements on adult and child patients were performed. Scanning times in all machines ranged from 1215 s.
An overall statistical uncertainty of 7% was estimated in the TLD read-out process. Reliability in doses owing to uncertainty in the placement of sachets was tested in a phantom, simulating possible shifts due to changes in focus-to-chip distances that could occur as a result of the presence of more or less hair, and was estimated to be better than 10%. Uncertainties due to patient size have been simulated by measuring anteroposterior distances in a sample of patients, concluding shifts of about 20% related to the mean value.
For intraoral exposures, dose estimates have been gathered based on quality control tests performed on 307 X-ray odontology installations. Data on X-ray tubes together with usual maximum and minimum exposure times used in the different projections were obtained. A mean exposure time was used (not weighted by the number of exposures). A mean entrance surface dose was calculated at the appropriate focus-to-skin distance, using a backscatter factor of 1.1 and a mass attenuation coefficient ratio from human tissue to air of 1.06.
| Results |
|---|
|
|
|---|
|
|
| Discussion |
|---|
|
|
|---|
The third quartile value, 670 µGy, is not owing to old equipment, as units with scanning times over 15 s (the oldest ones) have not been surveyed. Accordingly, a provisional local DRL value could be set at 0.7 mGy, with the aim of starting the optimization process.
Regarding dose estimates in intraoral radiology, the mean value differs from the value proposed by IAEA and adopted as the Spanish reference dose. The standard deviation suggests that very few X-ray units irradiate the patient above 7 mGy (less than 5% based on this work). Thus this value is not indicative of good practice and 3.5 mGy (approximately corresponding to the third quartile value) is a more appropriate DRL. This would also agree with the guidance value of 3 mGy considered by IAEA in initial draft documents.
With regard to the suitability of this value as a reference dose, it is worth noting that the average exposure rate produces a maximum dose value, using the average value of maximum exposure time, of 3.5 mGy; thus, the DRL proposed is not over restrictive, although the exposure time used in its estimation may be judged not to represent the most usual exposure condition. The recent appearance of an F-type film, which ensures E-type behaviour in manual processing, should lead to a further reduction in this value. Digital dental radiology would need to use specific DRLs and patient dose measurements, thus such units have not been included in the present study.
In the UK, the reference dose for bite wing exposure is 2.5 mGy at 70 kVp using E-speed film, and 5.0 mGy at 50 kVp. These values are higher when using D-speed film [19]. As the film currently used in most centres is the E-type, the reference doses proposed here agree with the UK approach. In addition, the mean dose value reported by the Finnish Radiation and Nuclear Safety Authority from molar X-ray dose measurements made on 909 dental units in 1999 is 3.5 mGy [20], equal to the value assessed in this work with the average maximum exposure times, and within the range of 0.816.4 mGy, close to the values presented here.
| Acknowledgments |
|---|
Received for publication May 30, 2000. Revision received July 28, 2000. Accepted for publication September 25, 2000.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. K. Looe, F. Eenboom, N. Chofor, A. Pfaffenberger, M. Steinhoff, A. Ruhmann, A. Poplawski, K. Willborn, and B. Poppe Conversion coefficients for the estimation of effective doses in intraoral and panoramic dental radiology from dose-area product values Radiat Prot Dosimetry, September 1, 2008; 131(3): 365 - 373. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. Mettler Jr, W. Huda, T. T. Yoshizumi, and M. Mahesh Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog Radiology, July 1, 2008; 248(1): 254 - 263. [Abstract] [Full Text] [PDF] |
||||
![]() |
L Gonzalez and J Moro Patient radiation dose management in dental facilities according to the X-ray focal distance and the image receptor type Dentomaxillofac. Radiol., July 1, 2007; 36(5): 282 - 284. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Poppe, H. K. Looe, A. Pfaffenberger, F. Eenboom, N. Chofor, M. Sering, A. Ruhmann, A. Poplawski, and K. Willborn Radiation exposure and dose evaluation in intraoral dental radiology Radiat Prot Dosimetry, February 1, 2007; 123(2): 262 - 267. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Poppe, H. K. Looe, A. Pfaffenberger, N. Chofor, F. Eenboom, M. Sering, A. Ruhmann, A. Poplawski, and K. Willborn Dose-area product measurements in panoramic dental radiology Radiat Prot Dosimetry, January 1, 2007; 123(1): 131 - 134. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Alcaraz, C Navarro, V Vicente, and M Canteras Dose reduction of intraoral dental radiography in Spain. Dentomaxillofac. Radiol., July 1, 2006; 35(4): 295 - 298. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Hatziioannou, E Psarouli, E Papanastassiou, P Bousbouras, H Kodona, O Kimoundri, and M Delichas Quality control and diagnostic reference levels in intraoral dental radiographic facilities Dentomaxillofac. Radiol., September 1, 2005; 34(5): 304 - 307. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. E. Tierris, E. N. Yakoumakis, G. N. Bramis, and E. Georgiou Dose area product reference levels in dental panoramic radiology Radiat Prot Dosimetry, October 11, 2004; 111(3): 283 - 287. [Abstract] [Full Text] [PDF] |
||||
![]() |
L Gonzalez, R Fernandez, V Ziraldo, E Vano, and R Ortega Reference level for patient dose in dental skull lateral teleradiography Br. J. Radiol., September 1, 2004; 77(921): 735 - 739. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Aroua, I Buchillier-Decka, K Dula, Y Nedjadi, M Perrier, J-P Vader, and J-F Valley Radiation exposure in dental radiology: a 1998 nationwide survey in Switzerland Dentomaxillofac. Radiol., July 1, 2004; 33(4): 211 - 219. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Isoardi and R Ropolo Measurement of dose-width product in panoramic dental radiology Br. J. Radiol., February 1, 2003; 76(902): 129 - 131. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |