British Journal of Radiology (2009) 82, 62-68
© 2009 British Institute of Radiology
doi: 10.1259/bjr/56572915
Optimizing the tube potential for lumbar spine radiography with a flat-panel digital detector
H GEIJER, MD, PhD, BA
E NORRMAN, PhD
and
J PERSLIDEN, PhD
1 Department of Radiology, Örebro University Hospital, SE-701 85, 2 Department of Clinical Medicine, Örebro University, SE-701 82, 3 Department of Medical Physics, Örebro University Hospital, SE-701 85 and 4 Department of Natural Sciences, Örebro University, SE-701 82, Örebro, Sweden

View larger version (18K):
[in this window]
[in a new window]
|
Figure 1. Overlay contour plot for the anteroposterior projection with 20 cm polymethylmethacrylate and a filtration of 4.5 mm Al. The default clinical settings — a tube potential of 77 kV and system speed 400 — can be found along the x-axis. The white area indicates possible settings that do not lead to a lower image quality or higher effective dose. Lower image quality figure (IQF) values indicate better image quality.
|
|

View larger version (17K):
[in this window]
[in a new window]
|
Figure 2. Overlay contour plot for the anteroposterior projection with 25 cm polymethylmethacrylate and a filtration of 4.5 mm Al, simulating a heavier patient. IQF, image quality figure.
|
|

View larger version (17K):
[in this window]
[in a new window]
|
Figure 3. Overlay contour plot for the lateral projection with 24 cm polymethylmethacrylate and a filtration of 4.5 mm Al. IQF, image quality figure.
|
|

View larger version (16K):
[in this window]
[in a new window]
|
Figure 4. Overlay contour plot for the lateral projection with 29 cm polymethylmethacrylate and a filtration of 4.5 mm Al, simulating a heavier patient. IQF, image quality figure.
|
|
Copyright © 2009 by the British Institute of Radiology.