BJR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

British Journal of Radiology (2006) 79, 562-571
© 2006 British Institute of Radiology
doi: 10.1259/bjr/51844219

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lee, J M
Right arrow Articles by Choi, B I
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lee, J M
Right arrow Articles by Choi, B I

Full paper

Radiofrequency ablation in pig lungs: in vivo comparison of internally cooled, perfusion and multitined expandable electrodes

J M Lee, MD1,2, J K Han, MD1,2, J M Chang, MD1,2, S Y Chung, MD1,2, S H Kim, MD1,2, J Y Lee, MD1,2 and B I Choi, MD1,2

1 Department of Radiology, and Institute of Radiation Medicine, Seoul National University College of Medicine, 28 Yongon-dong, Chongno-gu, Seoul 110-744, 2 Clinical Research Institute, Seoul National University Hospital, Seoul, Korea

The purpose of this study was to compare the amounts of in vivo coagulation obtained by radiofrequency (RF) ablation in porcine lung, using three types of electrodes. 15 in vivo ablation procedures were performed in the lungs of five pigs using three kinds of currently available RF devices under CT guidance. After placing an electrode in the lung, three ablation zones were created at each of three different regimens: Group A: RF ablation with an internally cooled electrode; Group B: RF ablation with a perfusion electrode, with instillation of 0.9% NaCl solution at a rate of 1.5 ml min–1; Group C: RF ablation with a multitined expandable electrode. According to the manufacturer's recommendations, RF application times were 12 min in group A and 20 min in group B. In group C, RF energy was delivered for 7 min after a mean temperature of 110°C was reached at 5 cm deployment. 36 min after the procedures, contrast-enhanced CT scans were obtained to evaluate the volume of zone of coagulation, and lungs were harvested for gross measurements. After macroscopic and histopathological analyses of 5 mm-thick lung sections, diameters, volumes and variation coefficients of regions of central coagulation were assessed. During RF ablation, the perfusion electrode allowed a larger energy delivery than the internally cooled or the multitined expandable electrodes, i.e. 33.6±4.7 kJ in group A, 40.0±8.2 kJ in group B and 23.5±6.1 kJ in group C (p<0.05). On gross observation, the cut surface of the gross specimen containing RF-induced coagulation showed that the ablated tissue appeared to be a central, firm, dark-brown area surrounded by an irregular outer margin (approximately 3–10 mm thick) of bright red tissue. In vivo studies showed that RF ablation using the perfusion electrode achieved larger coagulation volume than RF ablation using the other electrodes (p<0.05): 7.2±4.1 cm3 in group A; 16.9±5.5 cm3 in group B; 7.5±3.3 cm3 in group C. The corresponding variation coefficients were 0.55, 0.31, and 0.45, respectively. Our study shows that RF ablation using a perfusion electrode achieves a larger coagulation volume with an irregular margin than RF ablation using internally cooled or multitined expandable electrodes in the porcine lung.




This article has been cited by other articles:


Home page
Br. J. Radiol.Home page
BJR review of the year - 2006
Br. J. Radiol., March 1, 2007; 80(951): 147 - 151.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS 
Copyright © 2006 by the British Institute of Radiology.