British Journal of Radiology (2004) 77, 745-750
© 2004 British Institute of Radiology
doi: 10.1259/bjr/92529963

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Ramírez-Jiménez, F J
Right arrow Articles by Pacheco-Sotelo, J O
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Ramírez-Jiménez, F J
Right arrow Articles by Pacheco-Sotelo, J O

Full Paper

Considerations on the measurement of practical peak voltage in diagnostic radiology

F J Ramírez-Jiménez, ENG R López-Callejas, PhD J S Benítez-Read, PhD and J O Pacheco-Sotelo, PhD

Instituto Nacional de Investigaciones Nucleares, Direcciones de Servicios Tecnológicos e Investigación Científica, Apartado Postal 18-1027, Colonia Escandón, C.P. 11801, México D.F., México and Instituto Tecnológico de Toluca, Departamento de Electrónica, Apartado Postal 890, Toluca, México

An evaluation of the non-invasive measurement of the practical peak voltage (PPV) in the quality control of X-ray units used in diagnostic radiology was carried out. Two instruments were employed: the PTW Diavolt Universal Tester with readings in PPV and the Waveform Tester for X-rays (WATEX) prototype proposed here, which uses a PIN structure (P-type diffusion, Intrinsic region, N-type diffusion) photodiode as a sensor. The reference for the measurements was the voltage signal obtained in an oscilloscope from an invasive high voltage divider in order to verify the accuracy and precision of the measurements. The readings of the PPV in the Diavolt show a systematic error between 1% and 8%, always being less than the real value. An explanation for this difference is proposed, based on the relation between the effect of the X-rays in the film and the response of the sensor to the product of the applied voltage to the X-ray tube (peak voltage kVp), and the anode current. This explanation was confirmed using the WATEX waveform tester.




This article has been cited by other articles:


Home page
Br. J. Radiol.Home page
H-M KRAMER and H-J SELBACH
Extension of the range of definition of the practical peak voltage up to 300 kV
Br. J. Radiol., September 1, 2008; 81(969): 693 - 698.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS 
Copyright © 2004 by the British Institute of Radiology.