BJR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

British Journal of Radiology (2004) 77, 735-739
© 2004 British Institute of Radiology
doi: 10.1259/bjr/72698808

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gonzalez, L
Right arrow Articles by Ortega, R
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gonzalez, L
Right arrow Articles by Ortega, R

Full Paper

Reference level for patient dose in dental skull lateral teleradiography

L Gonzalez, PhD 1 R Fernandez, MD 2 V Ziraldo, MD 3 E Vano, PhD 2 and R Ortega, MD 4

1 Medical Physics Group, Radiology Department, Complutense University, 28040 Madrid, 2 Private dental specialist, Ricardo de la Vega St., 3, 28901 Getafe, 3 Private dental specialist, Mostoles St., 56 bis, 28943 Fuenlabrada and 4 Radiodiagnostic Service, Dentistry School, Complutense University, 28040 Madrid, Spain

The present work describes an experiment to obtain a local reference dose value in lateral skull teleradiography for dental applications. The reference value was based on patient measurements in seven dental X-ray units, using thermoluminescent chips, and measurements on a rubber globe filled with water at another 78 installations. Dosemeters were located initially in the head and neck of a human phantom to select the most suitable locations, and on the cephalostat of the X-ray unit at two appropriate locations, which did not interfere with the patient or with the usual imaging routine. The skin projection of the Porion point was considered the best position. 523 patients were monitored in the seven units; then patient doses and dose values from measurements on rubber balloons were compared to normalize and combine the data. The provisional reference value proposed is 400 µGy.




This article has been cited by other articles:


Home page
RadiologyHome page
F. A. Mettler Jr, W. Huda, T. T. Yoshizumi, and M. Mahesh
Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog
Radiology, July 1, 2008; 248(1): 254 - 263.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS 
Copyright © 2004 by the British Institute of Radiology.