BJR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

British Journal of Radiology (2003) 76, 824-831
© 2003 British Institute of Radiology
doi: 10.1259/bjr/25639755

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Amin, M N
Right arrow Articles by Harding, P F
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Amin, M N
Right arrow Articles by Harding, P F

Full Paper

A comparison of polyacrylamide gels and radiochromic film for source measurements in intravascular brachytherapy

M N Amin, MSc, MPhil 1 M A Horsfield, BSc, PhD 1 D E Bonnett, MSc, PhD 2 M J Dunn, MSc 3 M Poulton, MSc 3 and P F Harding, ONC 3

1 Division of Medical Physics, University of Leicester, Leicester LE1 5WW 2 Department of Medical Physics, Kent Oncology Centre, Maidstone and Tunbridge Wells NHS Trust, Kent ME16 9QQ and 3 Department of Medical Physics, University Hospitals of Leicester, Leicester LE1 5WW, UK

Correspondence: D E Bonnett

For intravascular brachytherapy with catheter-based systems, AAPM Task Group 60 has recommended measurements that should be made to characterize the sources. Beta emitters, including 90Sr/90Y are ideal for intravascular brachytherapy, but problems arise in measuring dose distributions in the high dose gradient region at short distances from the source. In this paper, measurements of radial and orthogonal dose distributions and dose profiles for a 90Sr/90Y source train using polyacrylamide gel (PAG) dosimetry and a high-field 4.7 Tesla MRI scanner are presented and compared with measurements made with two types of radiochromic film, MD-55 and HD-810. For the PAG system, the dose distributions were determined with in-plane resolutions of 0.4 mm and 0.2 mm. The measurements of absorbed dose distributions both orthogonal and parallel to the source axis show good agreement between the PAG and radiochromic film. The absolute dose at a radial distance of 2 mm in the central 32 mm of a line parallel to the axis was measured. For the PAG the measured absorbed dose was 1.25% lower, for MD-55 4% higher and for the HD-810 1.6% higher when compared with the value given by the source calibration. These results confirm that both absorbed dose and dose distributions for high gradient vascular brachytherapy sources can be measured using PAG but the disadvantages of gel manufacture and the need for access to a high resolution scanner suggests that the use of radiochromic film is the method of choice.




This article has been cited by other articles:


Home page
Radiat Prot DosimetryHome page
J. Kalef-Ezra, G. Bozios, P. Tsekeris, and L. Michalis
Patient and personnel dosimetry in endovascular radiotherapy with 90Sr/90Y sources
Radiat Prot Dosimetry, June 28, 2005; 114(4): 514 - 523.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS 
Copyright © 2003 by the British Institute of Radiology.