1 Department of Radiology, University Hospital Maastricht, P. Debijelaan 25, 6229 HX Maastricht and 2 Interfaculty Reactor Institute, Delft University of Technology, Mekelweg 15, 2629 JB Delft, The Netherlands
Patient radiation dose in angiography of the renal arteries was assessed and optimized after installing new radiological equipment. In three separate studies (n=50, 25 and 20) patient exposure was monitored in detail. For the first study default factory settings were used, for the second the number of digital subtraction angiography (DSA) images was halved and the X-ray beam filtering during fluoroscopy was increased, and for the third study filtering during DSA was increased as well. Standard projections were derived and used in Monte Carlo simulations to derive dose conversion coefficients to calculate effective dose from the dosearea product (DAP). Dose conversion coefficients were also calculated for CT angiography (CTA). Using default factory settings on the new angiography system, DAP, number of images and effective dose were much higher than on the replaced unit. For the studies given above, DAP was reduced from 144 Gy cm2 to 65 Gy cm2 to 32 Gy cm2, and effective dose from 22 mSv to 11 mSv to 9.1 mSv, respectively. Effective dose due to CTA was 5.2 mSv. It is concluded that modern angiography systems, resulting in high customer satisfaction, may readily cause much higher patient exposure than older systems. These doses may also be much higher than necessary. Optimization before putting such systems into use is absolutely essential. Internationally accepted recommendations for image quality and technique factors in angiography would be of great help.
This article has been cited by other articles:
![]() |
F. A. Mettler Jr, W. Huda, T. T. Yoshizumi, and M. Mahesh Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog Radiology, July 1, 2008; 248(1): 254 - 263. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.H. Backes and R.J. Nijenhuis Advances in Spinal Cord MR Angiography AJNR Am. J. Neuroradiol., April 1, 2008; 29(4): 619 - 631. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. W. Schultz and J. Zoetelief Dose conversion coefficients for interventional procedures Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 225 - 230. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |