1 Clinical Physics Group, St Bartholomew's Hospital, Queen Mary University, University of London, London EC1A 7BE and 2 Clinical Physics Group, St Bartholomew's Hospital, Barts and the London NHS Trust, London EC1A 7BE, UK
Patient radiation doses during interventional radiology procedures may reach the thresholds for radiation-induced skin and eye lens injuries. This study investigates the irradiated areas and doses received by patients undergoing cerebral embolisation, which is regarded as a high dose interventional radiology procedure. For each procedure the fluoroscopic and digital dosearea product (DAP), the fluoroscopic time, the total number of acquired images and entrance-skin dose (ESD) calculated by the angiographic unit were recorded. The ESD was measured by means of thermoluminescent dosimeters. In this study, the skin, eye and thyroid gland doses and the irradiated area for 30 patients were recorded. The average ESD was found to be 0.77 Gy for the posteroanterior plane and 0.78 Gy for the lateral plane. The average DAP was 48 Gy cm2 for the posteroanterior plane and 58 Gy cm2 for the lateral plane. The patient's average right eye dose was 60 mGy and the dose to the thyroid gland was 24 mGy. Seven patients received a dose above 1 Gy, one patient exceeded the threshold for transient erythema and one exceeded the threshold for temporary epilation. A good correlation between the DAP and the ESD for both planes has been found. The doctor's eye dose has also been measured for 17 procedures and the average dose per procedure was 0.13 mGy.
This article has been cited by other articles:
![]() |
T. Moritake, Y. Matsumaru, T. Takigawa, K. Nishizawa, A. Matsumura, and K. Tsuboi Dose Measurement on Both Patients and Operators during Neurointerventional Procedures Using Photoluminescence Glass Dosimeters AJNR Am. J. Neuroradiol., November 1, 2008; 29(10): 1910 - 1917. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Suzuki, S. Furui, Y. Matsumaru, S. Nobuyuki, M. Ebara, T. Abe, and D. Itoh Patient Skin Dose during Neuroembolization by Multiple-Point Measurement Using a Radiosensitive Indicator AJNR Am. J. Neuroradiol., June 1, 2008; 29(6): 1076 - 1081. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. Martin RADIATION DOSIMETRY FOR DIAGNOSTIC MEDICAL EXPOSURES Radiat Prot Dosimetry, March 28, 2008; (2008) ncm495v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Rampado and R. Ropolo Entrances skin dose distribution maps for interventional neuroradiological procedures: a preliminary study Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 256 - 259. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Suzuki, S. Furui, I. Kobayashi, T. Yamauchi, H. Kohtake, K. Takeshita, K. Takada, and M. Yamagishi Radiation Dose to Patients and Radiologists During Transcatheter Arterial Embolization: Comparison of a Digital Flat-Panel System and Conventional Unit Am. J. Roentgenol., October 1, 2005; 185(4): 855 - 859. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Struelens, F. Vanhavere, H. Bosmans, R. Van Loon, and H. Mol Skin dose measurements on patients for diagnostic and interventional neuroradiology: a multicentre study Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 143 - 146. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Persliden Patient and staff doses in interventional X-ray procedures in Sweden Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 150 - 157. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |