BJR
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

British Journal of Radiology (2003) 76, 541-545
© 2003 British Institute of Radiology
doi: 10.1259/bjr/60897046

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hatziioannou, K
Right arrow Articles by Bousbouras, P
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hatziioannou, K
Right arrow Articles by Bousbouras, P

Full Paper

A contribution to the establishment of diagnostic reference levels in CT

K Hatziioannou, MSc, PhD1, E Papanastassiou, MSc2, M Delichas, MSc, PhD2 and P Bousbouras, MSc3

1 Medical Physics Department, Papageorgiou Hospital, 56403 Thessaloniki, 2 Medical Physics Laboratory, Medical School, Aristotle University of Thessaloniki, 54006 Thessaloniki and 3 Social Security Institution, 19 Aristotelous st., 54624 Thessaloniki, Greece

CT has become the major source of population exposure to diagnostic X-rays. CT dose index (CTDI) and dose–length product (DLP) have been proposed as the appropriate dose quantities for the establishment of diagnostic reference levels for optimizing patient exposure. Dose measurements on 27 CT scanners in Northern Greece involving six routine CT examinations have been performed in order to compare their performance with the currently proposed European reference dose values and to produce a preliminary set of data for the establishment of local diagnostic reference levels. All measurements were performed using a pencil shaped ionization chamber introduced into polymethyl methacrylate cylindrical head and body phantoms. The results revealed significant discrepancies in dose values among the CT scanners, which can be mainly attributed to variations in the examination protocols and the different kinds of scanners. Significant overdosing compared with the European reference levels has not been observed, with the exception of the routine head examination, where 47% of the scanners exceeded the corresponding CTDIw value. CT scans in the trunk region result in the higher effective doses, which can reach estimated maximal values of the order of 15 mSv.




This article has been cited by other articles:


Home page
RadiologyHome page
V. Tsapaki, J. E. Aldrich, R. Sharma, M. A. Staniszewska, A. Krisanachinda, M. Rehani, A. Hufton, C. Triantopoulou, P. N. Maniatis, J. Papailiou, et al.
Dose Reduction in CT while Maintaining Diagnostic Confidence: Diagnostic Reference Levels at Routine Head, Chest, and Abdominal CT--IAEA-coordinated Research Project
Radiology, September 1, 2006; 240(3): 828 - 834.