The National Radiological Protection Board has recommended the introduction of dosewidth product (DWP) for the measurement of patient dose in panoramic dental radiology. The present work describes a method for measuring DWP using a pencil ionization chamber. The technique was tested on five panoramic dental units; the reproducibility of the method was 5.7%. In order to test the method, DWP was also assessed using thermoluminescent dosemeters and film. The results obtained agreed within 8.6% with those obtained using the pencil ionization chamber method. The proposed method appears to be simple and precise.