The British Journal of Radiology, Vol 73, Issue 875 1192-1199, Copyright © 2000 by British Institute of Radiology


ARTICLES

The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria

A Almen, A Tingberg, S Mattsson, J Besjakov, S Kheddache, B Lanhede, LG Mansson and M Zankl
Department of Radiation Physics, Malmo University Hospital, SE-205 02 Malmo, Sweden.

In this study we have investigated the image quality of lumbar spine radiographs taken after recording technical and physical parameters. Two technical parameters were altered, tube voltage (70 kV and 90 kV for the anteroposterior (AP) projection and 77 kV and 95 kV for the lateral projection) and sensitivity of the film-screen system (sensitivity class 400 and 600). In total, 85 images were included in the study. Entrance surface dose (ESD) was measured using thermoluminescent dosemeters. The mean value of ESD for the different technique groups varied between 1.9 mGy (90 kV, sensitivity class 400) and 4.6 mGy (70 kV, sensitivity class 400) for the AP projection, and between 6.4 mGy (95 kV, sensitivity class 600) and 20.4 mGy (70 kV, sensitivity class 400) for the lateral projection. Image criteria given in the "European Guidelines on Quality Criteria for Radiographic Images" were used to assess image quality. Two evaluation methods have been employed. A straightforward scoring of fulfilled image criteria, and visual grading analysis using the structures defined in the image criteria. The latter method provided a sharper distinction between groups of images taken using different radiographic techniques. The average number of fulfilled image criteria for the AP projections varied between 0.74 (90 kV, sensitivity class 400) and 0.87 (70 kV, sensitivity class 400). For the lateral projection this number varied between 0.79 (95 kV, sensitivity class 600) and 0.84 (77 kV, sensitivity class 600). This study shows that image criteria are useful tools in clinical studies of image quality.


This article has been cited by other articles:


Home page
Br. J. Radiol.Home page
G McVey
The effect of phantom type, beam quality, field size and field position on X-ray scattering simulated using Monte Carlo techniques
Br. J. Radiol., February 1, 2006; 79(938): 130 - 141.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
S. Borjesson, M. Hakansson, M. Bath, S. Kheddache, S. Svensson, A. Tingberg, A. Grahn, M. Ruschin, B. Hemdal, S. Mattsson, et al.
A software tool for increased efficiency in observer performance studies in radiology
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 45 - 52.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
A. Tingberg, M. Bath, M. Hakansson, J. Medin, J. Besjakov, M. Sandborg, G. Alm-Carlsson, S. Mattsson, and L. G. Mansson
Evaluation of image quality of lumbar spine images: a comparison between FFE and VGA
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 53 - 61.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
A. Tingberg, F. Eriksson, J. Medin, J. Besjakov, M. Bath, M. Hakansson, M. Sandborg, A. Almen, B. Lanhede, G. Alm-Carlsson, et al.
Inter-observer variation in masked and unmasked images for quality evaluation of clinical radiographs
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 62 - 68.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
H. Geijer and J. Persliden
Varied tube potential with constant effective dose at lumbar spine radiography using a flat-panel digital detector
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 240 - 245.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
A. Grahn, B. Hemdal, I. Andersson, M. Ruschin, A. Thilander-Klang, S. Borjesson, A. Tingberg, S. Mattsson, M. Hakansson, M. Bath, et al.
Clinical evaluation of a new set of image quality criteria for mammography
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 389 - 394.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
A Tingberg, C Herrmann, B Lanhede, A Almen, M Sandborg, G McVey, S Mattsson, W Panzer, J Besjakov, L G Mansson, et al.
Influence of the characteristic curve on the clinical image quality of lumbar spine and chest radiographs
Br. J. Radiol., March 1, 2004; 77(915): 204 - 215.
[Abstract] [Full Text] [PDF]


Home page
Dentomaxillofac RadiolHome page
B Molander, H-G Grondahl, and A Ekestubbe
Quality of film-based and digital panoramic radiography
Dentomaxillofac. Radiol., January 1, 2004; 33(1): 32 - 36.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
A C Offiah and C M Hall
Evaluation of the Commission of the European Communities quality criteria for the paediatric lateral spine
Br. J. Radiol., December 1, 2003; 76(912): 885 - 890.
[Abstract] [Full Text] [PDF]


Home page
Dentomaxillofac RadiolHome page
A Ekestubbe, H-G Grondahl, and B Molander
Quality of digital pre-implant tomography: comparison of film-screen images with storage phosphor images at normal and low dose
Dentomaxillofac. Radiol., September 1, 2003; 32(5): 322 - 326.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
G McVey, M Sandborg, D R Dance, and G Alm Carlsson
A study and optimization of lumbar spine X-ray imaging systems
Br. J. Radiol., March 1, 2003; 76(903): 177 - 188.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
B Lanhede, M Bath, S Kheddache, P Sund, L Bjorneld, M Widell, A Almen, J Besjakov, S Mattsson, A Tingberg, et al.
The influence of different technique factors on image quality of chest radiographs as evaluated by modified CEC image quality criteria
Br. J. Radiol., January 1, 2002; 75(889): 38 - 49.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
M Sandborg, A Tingberg, D R Dance, B Lanhede, A Almen, G McVey, P Sund, S Kheddache, J Besjakov, S Mattsson, et al.
Demonstration of correlations between clinical and physical image quality measures in chest and lumbar spine screen-film radiography
Br. J. Radiol., January 1, 2001; 74(882): 520 - 528.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS 
Copyright © 2000 by the British Institute of Radiology.