The British Journal of Radiology, Vol 73, Issue 874 1056-1067, Copyright © 2000 by British Institute of Radiology


ARTICLES

Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study

DR Dance, AK Thilander, M Sandborg, CL Skinner, IA Castellano and GA Carlsson
Joint Department of Physics, The Royal Marsden NHS Trust and Institute of Cancer Research, Fulham Road, London SW3 6JJ, UK.

The comparative performance of mammographic X-ray systems that use different anode/filter combinations has been assessed for screen-film and digital imaging. Monte Carlo techniques have been used to calculate average glandular dose as well as contrast and signal-to-noise ratio for imaging two test details. Five anode/filter combinations have been studied to establish the potential for dose saving or image quality improvement. For screen-film mammography, it was found that little benefit is gained by changing from a standard 28 kV molybdenum/molybdenum spectrum for breasts up to 6 cm thick. For thicker breasts, where the tube potential for the standard technique might be increased, 20% improvement in contrast can be achieved without dose penalty using molybdenum/rhodium or rhodium/rhodium spectra, whereas dose savings of more than 50% can be attained whilst maintaining contrast using tungsten/rhodium or rhodium/aluminium spectra. In digital mammography, a molybdenum/molybdenum spectrum delivers the lowest dose for a 2 cm breast, but gives the highest dose for thicker breasts. Tungsten/rhodium or rhodium/aluminium spectra provide the lowest doses at greater thicknesses. It is concluded that for screen-film mammography, molybdenum/molybdenum is the spectrum of choice for all but the thickest or most glandular breasts. In digital mammography, an alternative spectrum is preferable for breasts thicker than 2 cm.


This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
R. L. Smathers, J. M. Boone, L. J. Lee, E. A. Berns, R. A. Miller, and A. M. Wright
Radiation Dose Reduction for Augmentation Mammography
Am. J. Roentgenol., May 1, 2007; 188(5): 1414 - 1421.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
E. Samei, R. S. Saunders Jr, J. A. Baker, and D. M. Delong
Digital Mammography: Effects of Reduced Radiation Dose on Diagnostic Performance
Radiology, May 1, 2007; 243(2): 396 - 404.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
K C Young, J M Oduko, H Bosmans, K Nijs, and L Martinez
Optimal beam quality selection in digital mammography
Br. J. Radiol., December 1, 2006; 79(948): 981 - 990.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
H. Bosmans, A.-K. Carton, F. Rogge, F. Zanca, J. Jacobs, C. Van Ongeval, K. Nijs, A. Van Steen, and G. Marchal
Image quality measurements and metrics in full field digital mammography: an overview
Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 120 - 130.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
T. K. Nishino, X. Wu, and R. F. Johnson Jr.
Thickness of Molybdenum Filter and Squared Contrast-to-Noise Ratio per Dose for Digital Mammography
Am. J. Roentgenol., October 1, 2005; 185(4): 960 - 963.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
D. R. Dance, R. A. Hunt, P. R. Bakic, A. D. A. Maidment, M. Sandborg, G. Ullman, and G. Alm Carlsson
Breast dosimetry using high-resolution voxel phantoms
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 359 - 363.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
R. Klausz and N. Shramchenko
Dose to population as a metric in the design of optimised exposure control in digital mammography
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 369 - 374.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
R. A. Hunt, D. R. Dance, P. R. Bakic, A. D. A. Maidment, M. Sandborg, G. Ullman, and G. Alm Carlsson
Calculation of the properties of digital mammograms using a computer simulation
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 395 - 398.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
R. A. Hunt, D. R. Dance, M. Pachoud, G. Alm Carlsson, M. Sandborg, G. Ullman, and F. R. Verdun
Monte Carlo simulation of a mammographic test phantom
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 432 - 435.
[Abstract] [Full Text] [PDF]


Home page
Radiat Prot DosimetryHome page
M. C. Aznar, J. Medin, B. Hemdal, A. Thilander Klang, L. Botter-Jensen, and S. Mattsson
A Monte Carlo study of the energy dependence of Al2O3:C crystals for real-time in vivo dosimetry in mammography
Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 444 - 449.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
S Obenauer, K-P Hermann, and E Grabbe
Dose reduction in full-field digital mammography: an anthropomorphic breast phantom study
Br. J. Radiol., July 1, 2003; 76(907): 478 - 482.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
BJR DMFR IMAGING  ALL BIR JOURNALS 
Copyright © 2000 by the British Institute of Radiology.