The British Journal of Radiology, Vol 70, Issue 830 201-206, Copyright © 1997 by British Institute of Radiology
ARTICLES |
J Meyer and JA Mills
Coventry University, School of Natural and Environmental Sciences, UK.
The field uniformity of a superficial X-ray machine operating at 90 kV, 10 mA and filtered with a 1.1 mm aluminium beam hardening filter was investigated at a depth of 1 cm below the surface. Uniformity measurements were carried out using films and a densitometer to detect the relative absorbed dose across the field. Film dosimetry was assessed by comparison with ionization chamber dosimetry in a water tank. The original flat hardening aluminium filter was replaced by a combined, profiled filter to improve the uniformity across field sizes 20 cm, 5 cm and 2 cm diameter as well as hardening the beam. Flatness of the beam profile was improved for the 20 cm field size from +/-7.5% to +/-1.3% across the anode/cathode direction and from +/-7.9% to +/-4.7% in the anode/cathode direction. For the 5 cm field size the improvement was from +/-4% to +/-3% and from +/-5.3% to +/-3.6% and for 2 cm field size from +/-3.4% to +/-2.8% and from 10.5% to +/-9.7% in the same directions, respectively. Beam quality measurements were made and the original half-value-layer was reduced from 2.21 +/- 0.09 mm to 2.07 +/- 0.09 mm. The project demonstrated that it was possible to build a filter capable of flattening the beam profile for different sized applicators without significantly changing the penetrating ability of the beam.
This article has been cited by other articles:
![]() |
J. Meyer, K. J. Burnham, O. C.L. Haas, J. A. Mills, and E. M. Parvin Application of a least-squares parameter estimation approach for 2-D spatial modelling of compensators for intensity-modulated radiotherapy Transactions of the Institute of Measurement and Control, December 1, 2002; 24(5): 369 - 386. [Abstract] [PDF] |
||||
![]() |
P A Evans, A J Moloney, and P J Mountford Performance assessment of the Gulmay D3300 kilovoltage X-ray therapy unit Br. J. Radiol., June 1, 2001; 74(882): 537 - 547. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |