| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
The British Journal of Radiology, Vol 69, Issue 827 1032-1037, Copyright © 1996 by British Institute of Radiology
ARTICLES |
JR Williams
Department of Medical Physics and Medical Engineering, Western General Hospital, Edinburgh, UK.
Scatter dose as a function of dose-area product (DAP) has been studied over a range of tube potentials and scattering angles. Data are presented to permit the calculation of scatter dose from DAP at distances of 1 m or more from the patient. This method has been used to calculate absorbed dose to the walls of six existing X-ray rooms which are used for interventional radiology, angiography or barium contrast studies. These calculations have been compared with measurements using thermoluminescent dosemeters. The median ratio of measured to calculated dose was 0.50 and the maximum value was 1.26. It was concluded that the calculation of scatter dose from DAP using the factors derived in this report can be used as the basis for the specification of radiation protection barriers. The busiest room in this study in terms of DAP had a DAP rate of 224 Gy cm2 per day. From the scatter factors presented here, it was shown that the room walls did not need to have more than 1.2 mm of lead equivalence.
This article has been cited by other articles:
![]() |
I. A. Tsalafoutas, V. Tsapaki, A. Kaliakmanis, S. Pneumaticos, F. Tsoronis, E. D. Koulentianos, and G. Papachristou Estimation of radiation doses to patients and surgeons from various fluoroscopically guided orthopaedic surgeries Radiat Prot Dosimetry, January 1, 2008; 128(1): 112 - 119. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Schueler, T. J. Vrieze, H. Bjarnason, and A. W. Stanson An Investigation of Operator Exposure in Interventional Radiology RadioGraphics, September 1, 2006; 26(5): 1533 - 1541. [Abstract] [Full Text] [PDF] |
||||
![]() |
G McVey The effect of phantom type, beam quality, field size and field position on X-ray scattering simulated using Monte Carlo techniques Br. J. Radiol., February 1, 2006; 79(938): 130 - 141. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Kuon, J. B. Dahm, K. Empen, D. M. Robinson, G. Reuter, and M. Wucherer Identification of less-irradiating tube angulations in invasive cardiology J. Am. Coll. Cardiol., October 6, 2004; 44(7): 1420 - 1428. [Abstract] [Full Text] [PDF] |
||||
![]() |
G McVey and H Weatherburn A study of scatter in diagnostic X-ray rooms Br. J. Radiol., January 1, 2004; 77(913): 28 - 38. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Kuon, J Birkel, M Schmitt, and J B Dahm Radiation exposure benefit of a lead cap in invasive cardiology Heart, October 1, 2003; 89(10): 1205 - 1210. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. N. King, A. M. Champlin, C. A. Kelsey, and D. A. Tripp Using a Sterile Disposable Protective Surgical Drape for Reduction of Radiation Exposure to Interventionalists Am. J. Roentgenol., January 1, 2002; 178(1): 153 - 157. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |