In North Trent, UK, an entrance dose survey of lateral lumbar spine and antero-posterior (AP) abdomen examinations has been carried out in 17 radiology departments. The survey comprised 294 lateral lumbar spine and 322 AP abdomen entrance dose measurements. The mean entrance doses were found to be approximately half of the relevant national reference entrance dose levels of 30 mGy and 10 mGy, respectively. The effect of generator waveform on entrance dose was studied by separating the generators into two types: "pulsating potential" (PP) generators and "constant potential" (CP) generators for each examination. PP generators comprised 23% of the total number of examinations. The mean entrance dose and radiographic exposure factors from CP generators were found to be significantly lower than those from PP generators. The use of CP type generators, together with low attenuation components can significantly reduce patient
entrance doses for these examinations.
Received for publication August 12, 1994. Revision received November 21, 1994. Accepted for publication December 9, 1994.