Department of Radio-Oncology, University Hospital Zurich, Raemistrasse 100, 8091 Zurich, Switzerland * Department of Radiation Physics, St Bartholomew's Hospital, West Smithfield, London EC1A 7BE, UK
This study investigates the output from blocked treatment fields with 6 MV and 18 MV photons. Square fields, defined collimators, were blocked down to smaller symmetrical square, rectangular and asymmetrical rectangular fields using blocks on tray beneath the collimators. Up to 7.5% difference was found between the output measured for the blocked field and the output calculated using a collimator output factor for the surface field size. A method is described by which output factors for the blocked fields can be measured and used in practice. Three methods of approximating these output factors are also presented.
Key Words: Output factors Blocks Scatter
Received for publication June 17, 1991. Revision received October 24, 1991. Accepted for publication December 6, 1991.