Regional Medical Physics Department, Newcastle General Hospital, Westgate Road, Newcastle-Upon-Tyne NE4 6BE, UK
The measurement and prediction of scattered radiation dose to staff in diagnostic radiology is particularly important, owing to the increased use and complexity of interventional radiology. The air kerma-area product and scattered radiation dose in the vicinity of the patient couch, for both overcouch and undercouch X-ray tube geometries, were simultaneously monitored. The scattered radiation distribution at the couchside was deduced at a range of tube potentials for both overcouch and undercouch X-ray tube geometries. The variation of scattered radiation with field size on both geometries was investigated, as well as the variation with focus–table distance on an overcouch tube geometry. It was discovered that the scattered radiation dose at a point correlated with the air kerma–area product and the result may be used for radiation protection purposes. A method of predicting the scattered radiation dose at a given position is described.
Key Words: Radiology Radiation protection
Received for publication April 19, 1991. Revision received July 9, 1991. Accepted for publication July 15, 1991.
This article has been cited by other articles:
![]() |
O. O. Lie, G. U. Paulsen, and T. Wohni ASSESSMENT OF EFFECTIVE DOSE AND DOSE TO THE LENS OF THE EYE FOR THE INTERVENTIONAL CARDIOLOGIST Radiat Prot Dosimetry, December 4, 2008; (2008) ncn296v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
G McVey The effect of phantom type, beam quality, field size and field position on X-ray scattering simulated using Monte Carlo techniques Br. J. Radiol., February 1, 2006; 79(938): 130 - 141. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Faulkner Personnel and patient doses: are there ethical consequences to the use of X-rays? Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 30 - 33. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Kuon, J. B. Dahm, K. Empen, D. M. Robinson, G. Reuter, and M. Wucherer Identification of less-irradiating tube angulations in invasive cardiology J. Am. Coll. Cardiol., October 6, 2004; 44(7): 1420 - 1428. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Kuon, J Birkel, M Schmitt, and J B Dahm Radiation exposure benefit of a lead cap in invasive cardiology Heart, October 1, 2003; 89(10): 1205 - 1210. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Theocharopoulos, K. Perisinakis, J. Damilakis, G. Papadokostakis, A. Hadjipavlou, and N. Gourtsoyiannis Occupational Exposure from Common Fluoroscopic Projections Used in Orthopaedic Surgery J. Bone Joint Surg. Am., September 1, 2003; 85(9): 1698 - 1703. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Kovoor, M. Ricciardello, L. Collins, J. B. Uther, and D. L. Ross Risk to Patients From Radiation Associated With Radiofrequency Ablation for Supraventricular Tachycardia Circulation, October 13, 1998; 98(15): 1534 - 1540. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| BJR | DMFR | IMAGING | ALL BIR JOURNALS |