Bristol General Hospital, Bristol
Two-dimensional images obtained using ultrasound have been digitized from videotape recordings and stored within a maximum of 240 digital memory planes to form a three-dimensional data set using a commercially available image processing unit. This data set has been manipulated to produce images in planes perpendicular to the original scan set. The reformatted images represent not only the scans that could have been obtained by rotating the scan head but also demonstrate planes that are not accessible by conventional scanning. The system has been evaluated with a tissue-equivalent phantom to determine the geometric accuracy of the reformatting process. Clinical material has also been used to evaluate the practical value of such a technique and to highlight difficulties that may be encountered in its routine use.